  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Englewood",
      "address_1": "11039 East Lansing Circle",
      "reason_for_recall": "A circuit in use was occluded by a blue port cap that was inadvertently attached to the circuit as a result of an internal inspection error and will cause occlusion of gas flow.",
      "address_2": "",
      "product_quantity": "1,827,920 units US , 504,760 units OUS",
      "code_info": "PartNumber : 4S78XX00,155512,155518,156505,156545,1576162,1576214,542011,542021,542031,1553196,1572347,1553402,156205,156310,1574701,1574710,1574711,1574712,1574723,1572414,A4JXXXXX,6463,C311123X,C6X1XX1X,207002,C316123X,1560023,201814,1730327,20CF800C,C116123X,1570103,20CT8050,1577112,1552016,A4JXOXXX,CXX61X2X,1552017,1552018,20CT8048,20CT8055,M1171508,M1181639,1552020,M1162025,155200EU,1552017CN,1574792,M1171508VS,M1181639VS,1552026,1552027,1552028,",
      "center_classification_date": "20140123",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including Puerto Rico; and Internationally to: Belgium, Canada, Chile, Mexico, Costa Rica, Kuwait, United Kingdom, Greece, Czech Republic, Latvia, Turkey, China, France, Italy, Brazil, Korea, Germany, Thailand, Egypt, Austria, Dominican Republic, India, Israel, Colombia, Japan, Netherlands, Venezula, Saudi Arabia, Argentina, Malaysia, Sweden, Lebanon, South Korea, Finland, United Arab Emirates, Taiwan, Panama, Uruguay, Singapore, Hong Kong, Ecuador, Peru, Oman, Spain, Lithuania, and South Africa.",
      "state": "CO",
      "product_description": "Vital Signs Anesthesia Circuits    The Ventilator Circuit is intended for use as an interface between a ventilator dependent patient and a ventilator. The Anesthesia Circuit is intended to administer medical gases to a patient during anesthesia. The CPAP Circuit is intended to increase the airway pressure of a patient.",
      "report_date": "20140129",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Vital Signs Colorado Inc.",
      "recall_number": "Z-0812-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67164",
      "termination_date": "20140709",
      "more_code_info": "",
      "recall_initiation_date": "20131226",
      "postal_code": "80112-5909",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}