  "results": [
    {
      "status": "Terminated",
      "city": "Lawrence",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79185",
      "recalling_firm": "NxStage Medical, Inc.",
      "address_1": "350 Merrimack Street",
      "address_2": "N/A",
      "postal_code": "01843-1748",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and to the countries of :  United Kingdom & Spain",
      "recall_number": "Z-0811-2018",
      "product_description": "NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304 UDI-M535SAK3040, 5) REF SAK-306 UDI-M535SAK3060, 6) REF SAK-307 UDI- M535SAK3070, 6) REF SAK-402 UDI-M535SAK4020, 7) REF SAK-405 UDI- M535SAK4050, 8) REF SAK-406 UDI-M535SAK4060 & REF SAK-407 UDI- M535SAK4070    Hemodialysis systems and accessories",
      "product_quantity": "16,039 cases/2 units",
      "reason_for_recall": "May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to prepare  product.",
      "recall_initiation_date": "20180129",
      "center_classification_date": "20180301",
      "termination_date": "20200403",
      "report_date": "20180307",
      "code_info": "SAK-301- Lot # 70879245, Exp Date 7/11/2019  SAK-302- Lot # 71079056, Exp Date 7/30/2019  SAK-303- Lot #'s 70879181 & 70879224,   SAK-304- Lot #'s 70779236, 70879045, 70879105, 70879220 & 70979066, Exp Date 7/8/2019  SAK-306- Lot # 70979110, Exp Date 7/24/2019  SAK-307- Lot #'s  70879187 & 71079144, Exp Date 8/27/2019  SAK-402- Lot # 70879235, Exp Date 7/11/2019  SAK-405- Lot #'s 70779202 & 70879233, Exp Date 7/9/2019  SAK-406- Lot #'s 70979253, Exp Date 7/29/2019  SAK-407- Lot #'s 70879236 , Exp Date 7/11/2019"
    }
  ]
}