  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Union City",
      "address_1": "30031 Ahern Ave",
      "reason_for_recall": "Possible instability of Orthopedic Table Top.  If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.",
      "address_2": "",
      "product_quantity": "869 devices",
      "code_info": "Table model 5855, sub assembly 5855-901 Orthopedic Table Spar, manufactured between 1996 and 2007.",
      "center_classification_date": "20130213",
      "distribution_pattern": "USA Distribution -- Worldwide Distribution  including the states of  AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and WY., and the countries of Australia, Canada, Egypt, Finland, France, Germany, China, Israel, Italy, Japan, Netherlands, New Zealand, P.R. China, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Taiwan R.O.C., UAE and UK.",
      "state": "CA",
      "product_description": "Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.",
      "report_date": "20130220",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Mizuho OSI",
      "recall_number": "Z-0810-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63991",
      "termination_date": "20160415",
      "more_code_info": "",
      "recall_initiation_date": "20121217",
      "postal_code": "94587-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}