      "country": "United States",
      "city": "Sylmar",
      "address_1": "15900 Valley View Ct",
      "reason_for_recall": "Devices inappropriately displayed a low battery indicator determined to be due to memory corruption.",
      "address_2": "",
      "product_quantity": "2,906 impacted ICMs",
      "code_info": "Merlin PCS Programmer Software, V24.0.1 rev 1, UDI 05414734509725.    Confirm Rx ICM, UDI 05415067027320, US serial numbers:    3161783  3165630  3169553  3172363  3175335  3175336  3175337  3175349  3175356  3175377  3175378  3175385  3175393  3175394  3175395  3175399  3175402  3178901  3178907  3178909  3178916  3178919  3178921  3178930  3178937  3178943  3178945  3178946  3178951  3178954  3178962  3178969  3178974  3178981  3178995  3179005  3179024  3179103  3179123  3179149  3179177  3179199  3179260  3179443  3179452  3179520  3179533  3179539  3179542  3179543  3179570  3179583  3179589  3179595  3179603  3179609  3179850  3179851  3179921  3179929  3179953  3179965  3179972  3179988  3179991  3179997  3180019  3180030  3180038  3180309  3180326  3180333  3180411  3180420  3180421  3180424  3180578  3180604  3180607  3180608  3180617  3180661  3180675  3180679  3180692  3180695  3180704  3180969  3181038  3181066  3181067  3181168  3181171  3181174  3181175  3181199  3181203  3181217  3181226  3181233  3181246  3181248  3181267  3181494  3181655  3181683  3181688  3181742  3181744  3181745  3181776  3182193  3182217  3182228  3182229  3182230  3182248  3182262  3182300  3182523  3182585  3182594  3182617  3182632  3182858  3182863  3182885  3182890  3182939  3182950  3182965  3182972  3182989  3183002  3183003  3183005  3183043  3183105  3183112  3183270  3183281  3183294  3183307  3183321  3183332  3183638  3183715  3183739  3183777  3183782  3183821  3183842  3183847  3183915  3183961  3183979  3183987  3183988  3184008  3184076  3184142  3184160  3184192  3184225  3184238  3184239  3184251  3184259  3184297  3184307  3184321  3184389  3184407  3184415  3184431  3184441  3184442  3184443  3184453  3184469  3184471  3184476  3184487  3184490  3184495  3184510  3184529  3184531  3184547  3184552  3184555  3184565  3184582  3184599  3184656  3184667  3184690  3184700  3184717  3184726  3184727  3184813  3184862  3184887  3184892  3184946  3184953  3184968  3184970  3184972  3184987  3184994  3185013  3185026  3185032  3185078  3185362  3185456  3185821  3186402  3186416  3186422  3186425  3186456  3186463  3186472  3186502  3186505  3186512  3186518  3186527  3186585  3186804  3186807  3186816  3186830  3186835  3186842  3186907  3186923  3186951  3186977  3187010  3187013  3187014  3187016  3187049  3187066  3187070  3187093  3187126  3187130  3187137  3187276  3187298  3187378  3187379  3187384  3187386  3187392  3187394  3187545  3187556  3187560  3187574  3187596  3187614  3187623  3187625  3187637  3187645  3187661  3187664  3187666  3187854  3187865  3187889  3187891  3187896  3187911  3187917  3187932  3187944  3187955  3187994  3188003  3188006  3188014  3188015  3188018  3188031  3188035  3188052  3188064  3188071  3188074  3188078  3188087  3188120  3188140  3188320  3188326  3188334  3188359  3188366  3188404  3188423  3188584  3188586  3188589  3188605  3188632  3188654  3188664  3188671  3188685  3188694  3188697  3188702  3188704  3188708  3188709  3188712  3188718  3188721  3188727  3188841  3188845  3188856  3188862  3188865  3188869  3188872  3188873  3188874  3188880  3188893  3188899  3188902  3188920  3188924  3188938  3188947  3188967  3189159  3189161  3189196  3189212  3189215  3189216  3189249  3189253  3189267  3189271  3189283  3189489  3189503  3189508  3189522  3189547  3189554  3189778  3189785  3189886  3189893  3189914  3189916  3189920  3189923  3189930  3189938  3190096  3190120  3190129  3190171  3190247  3190295  3190312  3190316  3190333  3190336  3190346  3190347  3190350  3190366  3190368  3190379  3190382  3190385  3190386  3190388  3190486  3190535  3190566  3190579  3190624  3190754  3190763  3190783  3190799  3190806  3190809  3190812  3190819  3190826  3190837  3190842  3190996  3191005  3191022  3191042  3191065  3191086  3191101  3191105  3191110  3191329  3191364  3191368  3191395  3191405  3191429  3191465  3191582  3191616  3191644  3191650  3191662  3191667  3191685  3191699  3192031  3192050  3192051  3192064  3192101  3192105  3192111  3192121  3192125  3192130  3192136  3192147  3192291  3192312  3192316  3192368  3192475  3192492  3192556  3192561  3192573  3192576  3192582  3192586  3192655  3192658  3192659  3192735  3192736  3192751  3192758  3192771  3192858  3192866  3192884  3192891  3192899  3192917  3192939  3192949  3192965  3192981  3192990  3192999  3193000  3193022  3193052  3193169  3193225  3193232  3193248  3193252  3193262  3193273  3193275  3193286  3193294  3193297  3193307  3193333  3193366  3193372  3193376  3193453  3193817  3193818  3193846  3193860  3193863  3193881  3193882  3193889  3194026  3194031  3194033  3194035  3194075  3194092  3194103  3194122  3194125  3194416  3194421  3194432  3194440  3194455  3194466  3194479  3194480  3194511  3194539  3194555  3194557  3194577  3195575  3195587  3195688  3195781  3195788  3195797  3195806  3195830  3195841  3195853  3195857  3195859  3196176  3196180  3196189  3196218  3196242  3196296  3196341  3196353  3196356  3196775  3196777  3196786  3196796  3196798  3196805  3196807  3196845  3196855  3196857  3196882  3196908  3196979  3197112  3197133  3197153  3197173  3197180  3197189  3197199  3197222  3197236  3197243  3197432  3197450  3197457  3197470  3197490  3197492  3197494  3197495  3197500  3197521  3197524  3197563  3197570  3197596  3197602  3197606  3197656  3197661  3197690  3197729  3197751  3197755  3197762  3197774  3197915  3197916  3197937  3197971  3197992  3198017  3198025  3198044  3198196  3198218  3198318  3198440  3198444  3198448  3198459  3198480  3198488  3198491  3198498  3198499  3198521  3198544  3198561  3198584  3199050  3199061  3199084  3199092  3199115  3199126  3199140  3199162  3199172  3199192  3199195  3199204  3199205  3199208  3199217  3199219  3199232  3199275  3199286  3199287  3199314  3199319  3199320  3199325  3199328  3199341  3199350  3199355  3199358  3199374  3199377  3199383  3199387  3199393  3199395  3199398  3199399  3199407  3199415  3199426  3199427  3199432  3199434  3199470  3199481  3199503  3199504  3199523  3199526  3199551  3199590  3199961  3199982  3200014  3200073  3200086  3200149  3200168  3200180  3200198  3200203  3200286  3200308  3200320  3200321  3200322  3200324  3200325  3200338  3200342  3200348  3200357  3200358  3200372  3200387  3200412  3200435  3200451  3200627  3200711  3200724  3200733  3200741  3200797  3200806  3200826  3200827  3200847  3200865  3200870  3200883  3200893  3200899  3200917  3200945  3200952  3200953  3200965  3201027  3201033  3201043  3201048  3201051  3201059  3201066  3201093  3201097  3201101  3201103  3201120  3201121  3201127  3201135  3201138  3201140  3201145  3201151  3201157  3201162  3201173  3201178  3201180  3201184  3201186  3201187  3201192  3201193  3201366  3201380  3201390  3201395  3201399  3201401  3201434  3201461  3201469  3201482  3201484  3201493  3201505  3201506  3201510  3201511  3201923  3201926  3202007  3202018  3202021  3202032  3202047  3202075  3202082  3202084  3202094  3202099  3202519  3202533  3202545  3202554  3202561  3202594  3202633  3202734  3202736  3202739  3202742  3202754  3202758  3202809  3202810  3202830  3202846  3202892  3202895  3202898  3202904  3202908  3202914  3202933  3202941  3202952  3202954  3202964  3202969  3202972  3202978  3202985  3202995  3203028  3203031  3203036  3203045  3203047  3203077  3203078  3203086  3203105  3203109  3203114  3203119  3203214  3203222  3203284  3203291  3203296  3203300  3203308  3203439  3203442  3203458  3203464  3203484  3203489  3203506  3203507  3203528  3203564  3203584  3203609  3203610  3203616  3203630  3203647  3203649  3203995  3204019  3204027  3204028  3204044  3204076  3204083  3204100  3204123  3204130  3204147  3204149  3204230  3204237  3204255  3204261  3204262  3204274  3204279  3204287  3204308  3204376  3204386  3204397  3204421  3204424  3204428  3204432  3204474  3204506  3204514  3204522  3204525  3204568  3204569  3204602  3204683  3204714  3204840  3204860  3204870  3204886  3204889  3204891  3204908  3204911  3204925  3204933  3204949  3204951  3204972  3204978  3204979  3204982  3205003  3205011  3205054  3205088  3205094  3205102  3205106  3205115  3205125  3205169  3205245  3205262  3205265  3205271  3205284  3205285  3205289  3205302  3205344  3205364  3205647  3205656  3205679  3205701  3205722  3205770  3205773  3205775  3205779  3205911  3205940  3205951  3205986  3205992  3205994  3206032  3206037  3206041  3206045  3206061  3206228  3206280  3206331  3206346  3206795  3211563  3212438  3214778  3218353  3218402  3218628  3223638  3232608  9269482  9283582  9291621  9313773  and 9402889.    Confirm Rx ICM, UDI 05415067027320, OUS serial numbers:    3154089  3154100  3154112  3154130  3154134  3154146  3154183  3154189  3154198  3154209  3154211  3154236  3154237  3154250  3154255  3154265  3154297  3154298  3154304  3154314  3154319  3154399  3154405  3154406  3154418  3154419  3154429  3154433  3154435  3154460  3154936  3154984  3155011  3155020  3155023  3155027  3155031  3155427  3155428  3155434  3155452  3155455  3155512  3155514  3155518  3155519  3155536  3155551  3155557  3155563  3155564  3155566  3155571  3155576  3155589  3159287  3159292  3159295  3159303  3159306  3159307  3159308  3159311  3159312  3159314  3159315  3159316  3159317  3159326  3159638  3159642  3159646  3159658  3159662  3159665  3159677  3159679  3159741  3159748  3159750  3159751  3159753  3159755  3160364  3160365  3160368  3160375  3164133  3164135  3164137  3164139  3164141  3164142  3164143  3164147  3164153  3164160  3164175  3164177  3164179  3164183  3164184  3164185  3164188  3164189  3164199  3164206  3164208  3164209  3164213  3164214  3164216  3164217  3164219  3164220  3164227  3164250  3164251  3164255  3164256  3164262  3164263  3164264  3164265  3164270  3164274  3164281  3164289  3164295  3164299  3164305  3164306  3164309  3164310  3164317  3164322  3164323  3164330  3164332  3164335  3164337  3164342  3164353  3164357  3164365  3164368  3164369  3164371  3164372  3164376  3164378  3164385  3164388  3164389  3164393  3164397  3164408  3164417  3164418  3164423  3164427  3164428  3164443  3164444  3164445  3164450  3164452  3164455  3164465  3164471  3164481  3164483  3164484  3164488  3164489  3164491  3164492  3164498  3164504  3164509  3164511  3164518  3164520  3164525  3164533  3164539  3164540  3164546  3164553  3164556  3164563  3164565  3164570  3164571  3164572  3164577  3164581  3164583  3164584  3164585  3164589  3164590  3164593  3164599  3164601  3164603  3164605  3164610  3164612  3164613  3164614  3164618  3164619  3164622  3164623  3164625  3164633  3164636  3164638  3164639  3164640  3164642  3164648  3164649  3164652  3164654  3164657  3165466  3165468  3165473  3165474  3165475  3165484  3165489  3165490  3165493  3165502  3165506  3165507  3165508  3165512  3165515  3165519  3165521  3165523  3165524  3165526  3165528  3165531  3165535  3165538  3165544  3165546  3165549  3165550  3165552  3165560  3165561  3165563  3165566  3165567  3165570  3165571  3165573  3165575  3165576  3165579  3165586  3165591  3165593  3165594  3165600  3165604  3165605  3165607  3165608  3165610  3165613  3165615  3165616  3165617  3165623  3165624  3165629  3165634  3165635  3165637  3165638  3165641  3165644  3165647  3165648  3165649  3165652  3165654  3165655  3165659  3165662  3165665  3165667  3165668  3165669  3165673  3165674  3165679  3165680  3165681  3165682  3165685  3165688  3165689  3165693  3165697  3165700  3165703  3165706  3165713  3165717  3165721  3165725  3165727  3165729  3165733  3165736  3165743  3165744  3165746  3165750  3165752  3165754  3165756  3165761  3165768  3165769  3165770  3165771  3165773  3165781  3165784  3165786  3165789  3165794  3165796  3165798  3165811  3165813  3165815  3165816  3165820  3165821  3165823  3165825  3165827  3165828  3165839  3165841  3165846  3165847  3165849  3165850  3165852  3165856  3165857  3165860  3165863  3165864  3165867  3165870  3165871  3165874  3165876  3165884  3165888  3165892  3165894  3165896  3165897  3165898  3165900  3165904  3165905  3165909  3165913  3165917  3165918  3165920  3165924  3165925  3165930  3165933  3165938  3165942  3165943  3165947  3165952  3165958  3165963  3165965  3165966  3165974  3165976  3165979  3165981  3165984  3165989  3165990  3165991  3165992  3165993  3165996  3165997  3165998  3166000  3166002  3166003  3166004  3166005  3166007  3166013  3166015  3166018  3166019  3166020  3166034  3166037  3166039  3166040  3166043  3166044  3166050  3166054  3166055  3166058  3166059  3166063  3166064  3166065  3166069  3166077  3166080  3166082  3166084  3166085  3166086  3166087  3166088  3166095  3166097  3166101  3166119  3166120  3166124  3166125  3166128  3166132  3166136  3166139  3166143  3166151  3166509  3166510  3166512  3166519  3166522  3166525  3166526  3166531  3166539  3166541  3166542  3166543  3166544  3166547  3166548  3166549  3166552  3166554  3166556  3166557  3166559  3166560  3166561  3166568  3166572  3166573  3166576  3166577  3166581  3166583  3166585  3166588  3166589  3166594  3166600  3166601  3166605  3166606  3166607  3166615  3166618  3166619  3166620  3166624  3166626  3166628  3166637  3166638  3166639  3166640  3166648  3166649  3166651  3166652  3166653  3166657  3166663  3166665  3166667  3166668  3166678  3166679  3166680  3166685  3166686  3166691  3166696  3166697  3166704  3166710  3166976  3166977  3166978  3166980  3166987  3166990  3166991  3166996  3167004  3167012  3167015  3167017  3167028  3167030  3167035  3167038  3167039  3167040  3167041  3167043  3167053  3167055  3167064  3167066  3167067  3167223  3167233  3167235  3167238  3167240  3167249  3167254  3167257  3167258  3167270  3167272  3167284  3167286  3167288  3167292  3167296  3167304  3167310  3167312  3167580  3167586  3167589  3167591  3167594  3167611  3167627  3167641  3167643  3167651  3167655  3167668  3167670  3167671  3167796  3167798  3167810  3167813  3167821  3167837  3167843  3167847  3167849  3167854  3167859  3167877  3167881  3167884  3167890  3168029  3168030  3168031  3168033  3168035  3168036  3168037  3168049  3168052  3168058  3168064  3168067  3168072  3168073  3168075  3168077  3168081  3168082  3168087  3168090  3168097  3168098  3168115  3168122  3168125  3168378  3168383  3168384  3168387  3168390  3168395  3168398  3168410  3168411  3168414  3168415  3168416  3168417  3168418  3168420  3168421  3168424  3168426  3168428  3168436  3168440  3168441  3168444  3168457  3168459  3168464  3168465  3168468  3168469  3168679  3168684  3168687  3168690  3168692  3168693  3168703  3168712  3168717  3168734  3168740  3168742  3168746  3168754  3168756  3168761  3168766  3168771  3168775  3168881  3168883  3168890  3168891  3168893  3168897  3168902  3168903  3168904  3168905  3168915  3168917  3168922  3168923  3168924  3168926  3168932  3168937  3168940  3168953  3168962  3168967  3168972  3168973  3168975  3169113  3169115  3169119  3169134  3169136  3169144  3169157  3169161  3169166  3169168  3169176  3169181  3169186  3169195  3169201  3169318  3169320  3169322  3169324  3169341  3169342  3169345  3169346  3169347  3169348  3169349  3169354  3169357  3169358  3169359  3169360  3169361  3169367  3169374  3169380  3169382  3169394  3169401  3169403  3169497  3169503  3169505  3169507  3169508  3169514  3169517  3169521  3169523  3169536  3169544  3169546  3169549  3169552  3169554  3169555  3169559  3169562  3169568  3169591  3169595  3169707  3169710  3169711  3169720  3169722  3169724  3169744  3169756  3169761  3169764  3169766  3169770  3169782  3169787  3169858  3169859  3169861  3169862  3169863  3169864  3169866  3169868  3169873  3169876  3169879  3169880  3169881  3169884  3169885  3169891  3169902  3169907  3169908  3169910  3169912  3169914  3169918  3169925  3169941  3169955  3169956  3170048  3170050  3170051  3170060  3170073  3170083  3170085  3170088  3170095  3170100  3170104  3170107  3170108  3170109  3170110  3170113  3170118  3170120  3170121  3170123  3170124  3170131  3170132  3170137  3170396  3170397  3170412  3170417  3170422  3170429  3170431  3170432  3170445  3170450  3170451  3170452  3170453  3170454  3170556  3170557  3170561  3170565  3170567  3170575  3170579  3170584  3170586  3170588  3170590  3170591  3170594  3170667  3170678  3170680  3170683  3170684  3170685  3170694  3170695  3170697  3170701  3170705  3170706  3170707  3170712  3170721  3170722  3170827  3170834  3170835  3170839  3170925  3170926  3170932  3170936  3170938  3170942  3170949  3170951  3170954  3170956  3170961  3170962  3170963  3170964  3170966  3170969  3170973  3170974  3170976  3170978  3170981  3170983  3170984  3171011  3171012  3171014  3171015  3171018  3171019  3171024  3171025  3171026  3171027  3171036  3171038  3171051  3171053  3171061  3171064  3171065  3171066  3171068  3171082  3171083  3171091  3171101  3171102  3171105  3171115  3171122  3171125  3171129  3171134  3171136  3171437  3171438  3171441  3171443  3171447  3171448  3171461  3171462  3171478  3171486  3171489  3171494  3171497  3171498  3171504  3171511  3171514  3171517  3171524  3171525  3171559  3171561  3171562  3171565  3171569  3171571  3171583  3171590  3171592  3171593  3171595  3171597  3171601  3171602  3171624  3171631  3171639  3171641  3171643  3171646  3171648  3171649  3171651  3171653  3171951  3171958  3171962  3171963  3171976  3171978  3171980  3171981  3171982  3171986  3171987  3171988  3171992  3171999  3172019  3172030  3172033  3172100  3172104  3172107  3172109  3172111  3172117  3172127  3172135  3172139  3172141  3172142  3172143  3172146  3172147  3172156  3172160  3172161  3172163  3172166  3172167  3172169  3172171  3172174  3172175  3172177  3172178  3172180  3172183  3172184  3172185  3172187  3172191  3172192  3172195  3172198  3172350  3172351  3172353  3172354  3172355  3172360  3172364  3172368  3172370  3172372  3172377  3172379  3172382  3172383  3172392  3172394  3172399  3172402  3172403  3172404  3172408  3172411  3172416  3172421  3172425  3172427  3172432  3172433  3172434  3172436  3172440  3172520  3172524  3172528  3172540  3172541  3172542  3172543  3172559  3172561  3172563  3172564  3172574  3172576  3172582  3172583  3172585  3172587  3172588  3172593  3172595  3172597  3172599  3172601  3172605  3172608  3172614  3172616  3172621  3172630  3172633  3172637  3172749  3172753  3172754  3172768  3172770  3172774  3172780  3172787  3172793  3172795  3172796  3172797  3172802  3172803  3172804  3172805  3172807  3172809  3172816  3172818  3172820  3172830  3172831  3172832  3172833  3172836  3172838  3172839  3172842  3172844  3172846  3173070  3173072  3173074  3173077  3173079  3173088  3173090  3173093  3173094  3173099  3173102  3173103  3173104  3173105  3173111  3173114  3173116  3173127  3173130  3173136  3173230  3173231  3173241  3173315  3173318  3173320  3173327  3173328  3173335  3173337  3173340  3173344  3173348  3173350  3173352  3173353  3173354  3173355  3173356  3173358  3173360  3173361  3173363  3173539  3173543  3173551  3173552  3173561  3173568  3173569  3173571  3173572  3173575  3173579  3173580  3173581  3173590  3173593  3173599  3173606  3173613  3173617  3173619  3173621  3173624  3173632  3173635  3173637  3173640  3173644  3173647  3173648  3173652  3173655  3173658  3173878  3173884  3173887  3173896  3173948  3173955  3173959  3173963  3173971  3173972  3173978  3173979  3173983  3173986  3173988  3173991  3173999  3174001  3174009  3174014  3174016  3174017  3174019  3174021  3174023  3174026  3174032  3174034  3174037  3174039  3174042  3174047  3174112  3174113  3174136  3174138  3174143  3174150  3174152  3174156  3174160  3174165  3174170  3174171  3174173  3174175  3174178  3174179  3174181  3174182  3174183  3174184  3174186  3174189  3174194  3174195  3174201  3174206  3174208  3174439  3174443  3174446  3174453  3174459  3174468  3174470  3174481  3174482  3174491  3174492  3174495  3174496  3174501  3174507  3174516  3174517  3174520  3174521  3174527  3174737  3174744  3174749  3174755  3174761  3174769  3174773  3174774  3174777  3174779  3174783  3174792  3174793  3174794  3174797  3174804  3174814  3174816  3174817  3174819  3174820  3174821  3174822  3174823  3174824  3174825  3174826  3174827  3174886  3174893  3174895  3174898  3174901  3174902  3174908  3174909  3174910  3174920  3174927  3174929  3174932  3174936  3174939  3174944  3174945  3174950  3174953  3174957  3174959  3174963  3174966  3174967  3174970  3174971  3174975  3174977  3174978  3174980  3175161  3175165  3175167  3175171  3175186  3175188  3175189  3175193  3175197  3175198  3175203  3175207  3175208  3175210  3175216  3175218  3175228  3175232  3175241  3175246  3175248  3175252  3175255  3175257  3175258  3175325  3175347  3175368  3175382  3175396  3178923  3178959  3179097  3179133  3179179  3179327  3179407  3179423  3179430  3179493  3179499  3179501  3179506  3179510  3179611  3179865  3179876  3179919  3179949  3179955  3179974  3180314  3180383  3180681  3180689  3180700  3180952  3180964  3180981  3180996  3181004  3181007  3181008  3181012  3181016  3181029  3181031  3181041  3181051  3181057  3181212  3181220  3181504  3181654  3181659  3181754  3182507  3182517  3182584  3182867  3182868  3182881  3182896  3182904  3182928  3182929  3182941  3182962  3182963  3182977  3182993  3183037  3183039  3183051  3183053  3183063  3183075  3183108  3183114  3183115  3183260  3183264  3183266  3183282  3183322  3183324  3183334  3183338  3183349  3183636  3183649  3183650  3183662  3183677  3183678  3183687  3183724  3183725  3183737  3183747  3183752  3183760  3183762  3183778  3183830  3183833  3183835  3183913  3183919  3183920  3183937  3183943  3183954  3183956  3183959  3183960  3183966  3183976  3183980  3183997  3184009  3184010  3184015  3184033  3184035  3184042  3184081  3184082  3184149  3184164  3184175  3184250  3184344  3184354  3184378  3184411  3184428  3184429  3184434  3184437  3184439  3184444  3184451  3184452  3184455  3184456  3184460  3184484  3184494  3184509  3184513  3184515  3184519  3184576  3184593  3184596  3184646  3184710  3184721  3184765  3184773  3184803  3184857  3184889  3184893  3184986  3185003  3185119  3185125  3185127  3185129  3185131  3185134  3185135  3185141  3185146  3185148  3185152  3185154  3185157  3185163  3185166  3185170  3185173  3185175  3185176  3185177  3185179  3185190  3185191  3185192  3185195  3185196  3185199  3185202  3185214  3185225  3185227  3185233  3185236  3185239  3185241  3185243  3185247  3185249  3185251  3185252  3185254  3185258  3185259  3185262  3185268  3185300  3185301  3185303  3185307  3185308  3185309  3185310  3185317  3185320  3185322  3185326  3185332  3185338  3185339  3185343  3185350  3185352  3185353  3185354  3185355  3185356  3185357  3185360  3185366  3185370  3185377  3185378  3185381  3185386  3185389  3185390  3185391  3185393  3185395  3185397  3185400  3185408  3185409  3185410  3185412  3185413  3185414  3185419  3185420  3185422  3185423  3185424  3185427  3185431  3185441  3185454  3185458  3185459  3185460  3185461  3185464  3185471  3185472  3185474  3185477  3185478  3185480  3185484  3185485  3185489  3185492  3185495  3185520  3185522  3185528  3185530  3185562  3185607  3185615  3185631  3185669  3185671  3185672  3185673  3185674  3185678  3185694  3185698  3185700  3185702  3185704  3185711  3185712  3185716  3185728  3185734  3185738  3185751  3185753  3185754  3185900  3185902  3185965  3185983  3185984  3185987  3185988  3185989  3185990  3185991  3186001  3186003  3186005  3186007  3186008  3186041  3186043  3186048  3186049  3186050  3186051  3186054  3186057  3186059  3186060  3186071  3186072  3186075  3186076  3186080  3186081  3186082  3186083  3186088  3186089  3186094  3186095  3186096  3186100  3186103  3186106  3186117  3186129  3186132  3186137  3186140  3186141  3186144  3186150  3186163  3186164  3186167  3186169  3186170  3186171  3186176  3186186  3186187  3186231  3186232  3186237  3186241  3186244  3186245  3186246  3186247  3186250  3186254  3186257  3186263  3186267  3186275  3186279  3186281  3186282  3186288  3186289  3186292  3186297  3186298  3186305  3186309  3186310  3186311  3186314  3186315  3186316  3186317  3186319  3186321  3186324  3186329  3186334  3186335  3186337  3186340  3186341  3186344  3186348  3186350  3186352  3186355  3186358  3186360  3186362  3186364  3186365  3186369  3186370  3186371  3186381  3186383  3186384  3186387  3186390  3186393  3186394  3186398  3186405  3186408  3186409  3186417  3186423  3186424  3186439  3186443  3186447  3186451  3186453  3186471  3186477  3186478  3186480  3186481  3186482  3186487  3186494  3186496  3186498  3186504  3186506  3186509  3186510  3186525  3186528  3186538  3186539  3186546  3186549  3186552  3186554  3186555  3186572  3186578  3186579  3186582  3186594  3186598  3186605  3186609  3186615  3186621  3186623  3187027  3187028  3187116  3194690  3194693  3194696  3194697  3194698  3194699  3194700  3194704  3194706  3194707  3194708  3194709  3194714  3194718  3194726  3194727  3194728  3194735  3194738  3194743  3194744  3194745  3194747  3194750  3194752  3194755  3194756  3194758  3194759  3194760  3194761  3194763  3194764  3194766  3194767  3194775  3194777  3194783  3194784  3194785  3194786  3194788  3194796  3194797  3194798  3194802  3194813  3194814  3194816  3194819  3194820  3194825  3194828  3194934  3194937  3194944  3194946  3194948  3194949  3194954  3194959  3194962  3194965  3194967  3194969  3194970  3194976  3194980  3194983  3194984  3194985  3194988  3194991  3194992  3194995  3195000  3195002  3195003  3195006  3195015  3195025  3195028  3195029  3195032  3195033  3195035  3195036  3195041  3195055  3195056  3195060  3195061  3195063  3195069  3195071  3195079  3195081  3195082  3195444  3195448  3195452  3195453  3195454  3195456  3195458  3195461  3195463  3195465  3195468  3195469  3195473  3195475  3195476  3195480  3195483  3195485  3195490  3195492  3195493  3195495  3195496  3195502  3195507  3216724  3218731  3224523  3224554  3229886  3229978  3232312  3243862  3246173  and 3261583",
      "center_classification_date": "20200113",
      "distribution_pattern": "Worldwide distribution - US Nationwide. distribution. There was also government/military distribution.     Countries of Canada, Aruba, Australia, Belgium, Brazil, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Luxembourg, Martinique, Netherlands, New Caledonia, New Zealand, Norway, Oman, Poland, Portugal, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, and United Kingdom.",
      "state": "CA",
      "product_description": "Merlin PCS Programmer software model 3330, used with Confirm Rx Insertable Cardiac Monitor (ICM), Model DM3500, Sterile.  The firm name on the label of the ICM is St. Jude Medical Cardiac Rhythm Management Division, Sylmar, CA. - Product Usage: The Confirm Rx insertable cardiac monitor is designed to detect arrhythmias and wirelessly transmit data to the Merlin.net Patient Care Network. The Confirm Rx ICM is indicated for the monitoring and diagnostic evaluation of patients who experience unexplained symptoms such as: dizziness, palpitations, chest pain, syncope, and shortness of breath, as well as patients who are at risk for cardiac arrhythmias. It is also indicated for patients who have been previously diagnosed with atrial fibrillation or who are susceptible to developing atrial fibrillation.",
      "report_date": "20200122",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "St Jude Medical, Cardiac Rhythm Management Division",
      "recall_number": "Z-0808-2020",
      "initial_firm_notification": "Other",
      "product_type": "Devices",
      "event_id": "84453",
      "termination_date": "20210210",
      "more_code_info": "",
      "recall_initiation_date": "20180413",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}