      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67085",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED KINGDOM, IRELAND, IRAQ, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MARITIUS, MALDOVA, NEW ZEALAND, NIGERIA, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, QATAR, RUSSIA, SLOVAKIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, TUNISIA, UNITED ARAB EMIRATES, VENEZUELA, CHILE, CANADA, and BERMUDA.",
      "recall_number": "Z-0808-2014",
      "product_description": "GE Healthcare, TRAM MODULE, High-acuity patient monitoring. Product Usage: The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another.   K900540: TRAM modules are intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The intended use of TRAM modules is to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. The TRAM module functions as a transport monitor during transport, and in the bedside monitor during ICU or OR monitoring. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. TRAM modules acquire, process and store information regarding these parameters. TRAM modules incorporate different monitoring capabilities based on their configuration. K011000: The TRAM 2001 Module is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. The TRAM 2001 Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric, and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM 2001 Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM 2001 Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport. K053121: The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system. Physiological parameter data includes ECG, invasive blood pressure, non-invasive blood pressure, non-invasive blood pressure, pulse oximetry, cardiac output, temperature and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctors office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.",
      "product_quantity": "1,948  (1,834 US; 114 OUS)",
      "reason_for_recall": "Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.",
      "recall_initiation_date": "20131205",
      "center_classification_date": "20140122",
      "termination_date": "20160725",
      "report_date": "20140129",
      "code_info": "Serial Number  C7HA1686G  C7HA1700G  C7HA1701G  C7HA1696G  C7HA1692G  C7HA1688G  C7HA1691G  C7HA1689G  C7HA1687G  C7HA1702G  B7HA1657G  C7HA1683G  C7HA1703G  C7HA1707G  C7HA1709G  C7HA1711G  D8HA3141G  D8HA3154G  D8HA3135G  D8HA3136G  D8HA3149G  D8HA3148G  D8HA3128G  D8HA3134G  D8HA3152G  D8HA3140G  D8HA3137G  D8HA3144G  H3S20070G  H3S20072G  J3S20073G  J3S20075G  J3S20078G  J3S20084G  J3S20089G  J3S20090G  J3S20093G  J3S20085G  A4S20482G  A4S20471G  G4S21111G  G4S21108G  G4S21112G  G4S21090G  G4S21106G  F5HB0489G  E5HB0418G  F5HB0493G  F5HB0491G  F5HB0468G  F5HB0497G  F5HB0478G  E5HB0459G  C5HB0346G  B5HB0261G  A6HB0980G  A6HB0978G  D6HB1379G  K7HB4546G  K7HB4542G  K7HB4552G  K7HB4553G  K7HB4547G  K7HB4556G  C8HB5556G  C8HB5743G  C8HB5744G  C8HB5758G  C8HB5747G  C8HB5746G  C8HB5750G  C8HB5756G  C8HB5757G  C8HB5676G  D8HB5922G  G8HB6864G  G8HB6859G  B9HA4387G  B9HA4403G  B9HA4405G  B9HA4402G  F8HB6464G  L4HB0063G  L4HB0068G  L4HB0064G  M4HB0151G  M4HB0143G  A5HA0120G  A5HA0118G  A5HA0125G  A5HA0122G  A5HA0112G  A5HA0117G  A5HA0111G  A5HA0119G  A5HA0109G  A5HA0114G  A5HA0124G  L4HC0003G  H5HA0362G  H5HA0367G  D6HB1475G  D6HB1467G  D6HB1455G  D6HB1458G  D6HB1442G  D6HB1469G  D6HB1430G  D6HB1465G  D6HB1435G  D6HB1476G  D6HB1440G  D6HB1463G  D6HB1439G  D6HB1438G  D6HB1461G  D6HB1460G  M6HD8751F  M6HD8723F  M6HD8732F  M6HD8683F  M6HD8749F  M6HD8725F  M6HD8748F  M6HD8728F  M6HD8737F  M6HD8730F  M6HD8750F  M6HD8726F  M6HD8747F  M6HD8724F  M6HD8736F  M6HD8744F  M6HD8740F  M6HD8731F  M6HD8745F  M6HD8722F  M6HD8739F  M6HD8741F  M6HD8742F  M6HD8738F  M6HD8746F  A7HA1527G  A7HA1564G  A7HA1550G  A7HA1546G  A7HA1562G  A7HA1529G  A7HA1561G  A7HA1563G  A7HA1560G  B7HA1598G  B7HA1595G  B7HA1583G  B7HA1579G  B7HA1591G  B7HB2686G  B7HB2694G  B7HB2676G  F7HB3564G  F7HB3571G  F7HB3561G  F7HB3567G  F7HB3499G  F7HB3498G  F7HB3526G  F7HB3511G  F7HB3516G  F7HB3536G  F7HB3537G  F7HB3538G  F7HB3541G  F7HB3534G  F7HB3533G  F7HB3539G  F7HB3532G  F7HB3535G  F7HB3548G  F7HB3540G  F7HB3589G  F7HB3556G  G7HB3740G  G7HB3773G  G7HB3772G  G7HB3774G  G7HB3747G  G7HB3769G  G7HB3775G  G7HB3748G  G7HB3766G  G7HB3746G  G7HB3735G  G7HB3741G  G7HB3726G  H7HB4010G  C8HB5492G  C8HB5488G  C8HB5481G  C8HB5480G  C8HB5493G  C8HB5476G  C8HB5508G  C8HB5485G  C8HB5482G  C8HB5449G  C8HB5484G  C8HB5475G  C8HB5490G  C8HB5512G  C8HB5491G  C8HB5474G  C8HB5453G  C8HB5459G  C8HB5456G  M8HB8682G  M8HB8677G  A9HB8818G  A9HB8832G  M8HB8756G  A9HB8806G  A9HB8802G  A9HB8824G  A9HB8817G  A9HB8809G  M8HC0420G  M8HC0427G  M8HC0413G  M8HC0422G  M8HC0430G  M8HC0414G  M8HC0419G  M8HC0416G  M8HC0424G  M8HC0417G  M8HC0418G  M8HC0423G  M8HC0425G  M8HC0415G  M8HC0421G  M8HC0428G  M8HC0426G  A9HC0432G  A9HC0433G  A9HC0436G  A9HC0435G  A9HC0446G  A9HC0439G  A9HC0451G  A9HC0437G  A9HC0440G  A9HC0434G  A9HC0447G  A9HC0444G  C9HB9411G  C9HB9415G  C9HB9445G  C9HB9419G  C9HB9298G  C9HB9323G  B9HB9240G  C9HB9549G  C9HB9563G  C9HB9326G  C9HB9322G  C9HB9366G  C9HB9439G  C9HB9452G  C9HB9451G  C9HB9447G  C9HB9450G  C9HB9530G  A8HB5120G  RT405523022GA  SBB06251137GA  D2TW2498G  RTW07332120GA  L2T14637G  A2TW1861G  J2T14014G  SBB06523641GA  D3T42441G  A4T19101G  RT105433210GA  K1T10362G  M1TW1346G  SBT06180805GA  RT106044903GA  RT106044763GA  RT410139269GA  SBB11440867GA  SBB09488690GA  A0HD3119F  E0HD3395F  E0HD3397F  K0HD3752F  K0HD3759F  K0HD3762F  K1HD4284F  K1HD4291F  E8HA3281GR  M8HA4222GR  G9HA5212G  F9HA5071G  F9HA5059G  F9HA5134G  C9HA4702G  F9HA5131G  G9HA5268G  G9HA5265G  G9HA5262G  F9HA5145G  K9HB1855G  K0HA7181G  K0HA7178G  K0HA7182G  K0HA7189G  K0HA7173G  K0HA7167G  J0HA7192G  J0HA7195G  L0HA7794G  L0HA7818G  L0HA7815G  L0HA7804G  H9HB0934GR  L9HB2176GR  D0HB4296G  C0HB4122GR  D9HB9695G  F9HB0324G  F9HB0497G  E0HB4448G  E0HB4502G  E0HB4547G  E0HB4604G  E0HB4616G  E0HB4680G  E0HB4687G  E0HB4819G  E0HB4881G  F0HB5010G  D0HC0713G  J0HB5841G  J0HB5865G  J0HB5975G  M0HB6696G  M0HB6711G  M0HB6760G  M0HB6922G  M0HB6925G  M0HB6950G  M0HB6953G  M0HB6971G  M0HB6973G  M0HB6974G  M0HB6976G  M0HB6977G  M0HB6979G  M0HB6980G  M0HB6982G  M0HB6983G  M0HB6986G  M0HB6989G  M0HB6990G  M0HB6994G  M0HB6995G  M0HB6996G  M0HB6999G  M0HB7000G  M0HB7001G  M0HB7002G  M0HB7006G  M0HB6968G  M0HB6988G  M0HB6991G  M0HB7010G  M0HB7011G  M0HB7012G  M0HB7013G  M0HB7015G  M0HB7016G  M0HB7017G  M0HB7018G  M0HB7019G  M0HB7020G  M0HB7023G  M0HB7024G  M0HB7025G  M0HB7026G  M0HB7028G  M0HB7029G  A1HB7547G  A1HB7598G  A1HB7601G  A1HB7604G  M0HB6970G  C1HB8191G  C1HB8220G  C1HB8224G  H1HB9502G  H1HB9503G  H1HB9504G  H1HB9513G  J9HB1067GR  J9HB1245GR  M9HB2825GR  M9HB2899GR  A0HB3179GR  A0HB3216GR  C0HA6610G  C0HA6649G  C0HA6503G  C0HA6510G  J0HB5995G  K0HB6241G  K0HB6275G  K0HB6301G  B1HB7863G  C1HB8082G  C1HB8088G  C1HB8092G  C1HB8107G  C1HB8108G  C1HB8136G  C1HB8137G  C1HB8147G  C1HB8156G  C1HB8162G  C1HB8163G  C1HB8170G  C1HB8175G  C1HB8176G  C1HB8178G  C1HB8210G  C1HB8219G  C1HB8230G  D1HB8610G  D1HB8632G  H1HD4215F  RT405160722GA  RT407206340GA  RT407206403GA  RT407226441GA  SCS07171073GA  SCS07171078GA  SCS07140998GA  RT407266610GA  SCS07461977GA  RT407487540GA  RT407517574GA  RT407517605GA  RT407517619GA  RT408027651GA  RT408298081GA  RT408298094GA  RT408338100GA  RT408338106GA  RT408338108GA  RT408338115GA  RT408338118GA  RT408338122GA  RT408338129GA  RT408338131GA  RT408338141GA  RT408358146GA  RT408358150GA  RT408358151GA  RT408358153GA  RT408358154GA  RT408418385GA  RT408418423GA  RT408438511GA  RT408438512GA  RT408438542GA  RT408478600GA  RT408478648GA  RT408468610GA  RT408478606GA  SCS09464285GA  SCS09464300GA  RST10331659GA  RST10331672GA  SCS10415184GA  B3T15631G  D3T16359G  D4T19517G  D4T19634G  D4T19637G  D4T19638G  RT105392851GA  RT105392872GA  RT105392880GA  RT106034493GA  RT106034498GA  RT106024368GA  RT106034464GA  RT106034486GA  RT106034488GA  RT106034490GA  RT106034492GA  RT106034496GA  RT106034518GA  RT106044521GA  RT106044537GA  RT106044540GA  RT106044541GA  RT106044545GA  RT106044546GA  RT106044549GA  RT106044551GA  RT106044552GA  RT106044553GA  RT106044555GA  RT106044556GA  RT106044599GA  RT106044600GA  RT106044606GA  RT106044653GA  RT106044656GA  RT106044657GA  RT106044659GA  RT106044660GA  SBB06140254GA  SBB06140258GA  SBB06140262GA  SBB06140263GA  SBB06140265GA  SBB06140266GA  SBB06140270GA  SBB06140290GA  SBB06140293GA  SBB06140279GA  SBB06140280GA  SBB06140287GA  SBB06140346GA  SBB06140348GA  SBB06140349GA  SBB06140353GA  SBB06140354GA  SBB06140356GA  SBB06150393GA  SBB06261196GA  SBB06261199GA  SBB06251122GA  SBB06261164GA  SBB06261168GA  SBB06261174GA  SBB06281253GA  SBB06442791GA  SBB06442801GA  SBB06442808GA  SBB06442814GA  SBB06442822GA  SBB06442841GA  SBB06442842GA  SBB06442847GA  SBB06442849GA  SBB06452929GA  SBB06452937GA  SBB06452947GA  SBB06463042GA  SBB06463048GA  SBB06463052GA  SBB06463055GA  SBB06473062GA  SBB06473063GA  SBB06473066GA  SBB06473079GA  SBB06473092GA  SBB06473095GA  SBB06483125GA  SBB06483129GA  SBB06483135GA  SCS06440389GA  SCS06440393GA  SCS06460454GA  SCS06460459GA  SCS06460461GA  SBB07054052GA  SBB07054093GA  SBB07054094GA  SBB07054096GA  SBB07054116GA  SBB07064134GA  SBB07064139GA  SBB07064172GA  SBB07074249GA  SBB07074255GA  SBB07054016GA  SBB07054022GA  SBB07064127GA  SBB07094279GA  SBB07114328GA  SBB07134417GA  SBB07134422GA  SBB07134423GA  SBB07134426GA  SBB07134429GA  SBB07325311GA  SBB07435898GA  SBB07465993GA  SBB07435891GA  SBB07465996GA  SBB07466021GA  SBB07405607GA  SBB08307124GA  SBB08327193GA  RST10129859GA  RST10129860GA  RST10129866GA  RST10129867GA  RST10129872GA  SBB10129121GA  SBB10129127GA  SBB10129133GA  SBB10129134GA  SBB10129144GA  RST10311218GA  RST10311239GA  RST10311240GA  RST10311274GA  RST10321454GA  RST10422237GA  RST10422254GA  RST10422257GA  RST10422258GA  RST10422269GA  SBB10410099GA  SBB10410109GA  SBB10410132GA  SBB10410137GA  SBB10430157GA  SBB10430179GA  SBB10430182GA  SBB10430187GA  SBB10430202GA  SCS10365077GA  SCS10365083GA  SCS10455305GA  SCS10465331GA  SCS10465332GA  SCS10465333GA  M3HD4847F  RT105100736GA  RT105100749GA  RT105120796GA  RT105130919GA  RT105140929GA  RT105140936GA  RT105140939GA  RT105140941GA  RT105140942GA  RT105140943GA  RT105140946GA  RT105140964GA  RT105140972GA  RT105262035GA  RT105262047GA  RT105262066GA  SBB08256964GA  SBB08256969GA  SBB08256984GA  SBB08256986GA  SBB08256996GA  SBB08257013GA  SBB08257031GA  SBB08267040GA  SBB08267041GA  SBB08267048GA  RT408258009GA  RT408258011GA  RT408268060GA  RT408268069GA  RT408268075GA  RT408268078GA  RT408298087GA  RT408298088GA  RT408298093GA  RT408338102GA  RT408338104GA  RT408338109GA  RT408338120GA  RT408338134GA  RT408338135GA  RT408338136GA  RT408338138GA  RT408338139GA  RT408358143GA  RT408358144GA  RT408358145GA  RT408358147GA  RT408358148GA  RT408358156GA  RT408358158GA  RT408358162GA  RT408358163GA  RT408358164GA  RT408358165GA  RT408358166GA  RT408358167GA  RT408358169GA  RT408358171GA  RT408358172GA  RT408358173GA  RT408358176GA  RT408358177GA  RT408368188GA  RT408368195GA  RT408368196GA  RT408368197GA  RT408368201GA  RT408368202GA  RT408368203GA  RT408368204GA  RT408368206GA  RT408368207GA  RT408368208GA  RT408378224GA  RT408378236GA  RT408378245GA  RT408378257GA  RT408378265GA  RT408378268GA  RT408378273GA  RT408378276GA  RT408388283GA  RT408388284GA  RT408388285GA  RT408388286GA  RT408388287GA  RT408388288GA  RT408388289GA  RT408388290GA  RT408388292GA  RT408388293GA  RT408388294GA  RT408388295GA  RT408388297GA  RT408388298GA  RT408388299GA  RT408388300GA  RT408388301GA  RT408388302GA  RT408388303GA  RT408338127GA  RT408388307GA  RT408408314GA  RT408408320GA  RT408408323GA  RT408408329GA  RT408408333GA  RT408408352GA  RT408408355GA  RT408408357GA  RT408408358GA  RT408408364GA  RT408408366GA  RT408408367GA  RT409318900GA  RT409318902GA  RT409318909GA  RT409398955GA  RST10240507GA  RST10240519GA  RST10240521GA  RST10240530GA  RST10270797GA  RST10270800GA  RST10270801GA  RST10280987GA  RST10280992GA  RST10280994GA  RST10280995GA  RST10280996GA  RST10281004GA  RST10281006GA  RT410289447GA  RT410289449GA  RT410289452GA  RT410289456GA  RT410289458GA  RT410289466GA  RT410289473GA  K2T14065G  K2T14070G  K2T14071G  K2T14150G  K2T14155G  K2T14092G  K2T14101G  K2T14145G  K2T14151G  K2T14156G  K2T14162G  K2T14229G  K2T14231G  J2T14029G  J2T14042G  J2T14056G  J2T14059G  K2T14063G  K2T14067G  K2T14074G  K2T14077G  K2T14080G  K2T14088G  K2T14095G  K2T14099G  K2T14103G  K2T14104G  K2T14106G  K2T14107G  K2T14109G  K2T14119G  K2T14120G  K2T14124G  K2T14134G  K2T14137G  K2T14138G  K2T14143G  K2T14144G  K2T14147G  K2T14152G  K2T14153G  K2T14158G  K2T14167G  K2T14177G  K2T14180G  K2T14186G  K2T14195G  K2T14196G  K2T14227G  K2T14261G  G2T13326G  H2T13621G  H2T13623G  H2T13627G  H2T13636G  G2T13413G  G2T13420G  G2T13443G  G2T13445G  G2T13468G  G2T13512G  E2TW2857G  H2TW3254G  G2T13339G  G2T13345G  G2T13346G  D2T12788G  D2T12830G  D2T12873G  E2T12963G  E2T13009G  E2T13014G  E2T13018G  E2T13086G  F2T13126G  F2T13129G  F2T13136G  F2T13137G  F2T13141G  F2T13156G  F2T13159G  C2T12377G  D2T12574G  D2T12682G  D2T12701G  D2T12800G  D2T12825G  E2T12897G  E2T12969G  E2T12974G  E2T12977G  E2T12980G  E2T13023G  E2T13046G  E2T13052G  E2T13053G  E2T13055G  E2T13056G  E2T13060G  E2T13066G  E2T13067G  E2T13069G  E2T13073G  E2T13075G  E2T13079G  E2T13084G  F2T13112G  F2T13113G  F2T13114G  F2T13116G  F2T13118G  F2T13119G  F2T13128G  F2T13130G  F2T13131G  F2T13132G  F2T13134G  F2T13135G  F2T13148G  F2T13149G  F2T13151G  F2T13152G  F2T13153G  F2T13155G  F2T13158G  F2T13160G  F2T13161G  F2T13164G  F2T13167G  F2T13169G  F2T13176G  F2T13177G  F2T13181G  F2T13187G  F2T13190G  F2T13195G  F2T13202G  F2T13203G  F2T13211G  E2T13039G  E2T13049G  D0HB4427GR  B3TW4319G  B3TW4323G  B3TW4340G  B3TW4342G  B3TW4350G  B3TW4377G  B3TW4380G  B3TW4382G  B3TW4386G  B3TW4390G  B3TW4392G  B3TW4393G  B3TW4398G  B3TW4402G  B3TW4496G  B3TW4502G  B3TW4505G  A4TW5401G  B4TW5519G  B4TW5525G  B4TW5578G  H4T11734G  H4T11750G  H4T11752G  H4T11754G  H4T11756G  H4T11792G  H4T11798G  H4T11799G  M4T13110G  M4T13114G  M4T13115G  M4T13116G  M4T13118G  M4T13120G  M4T13122G  M4T13123G  M4T13125G  M4T13126G  M4T13152G  M4T13154G  M4T13160G  M4T13165G  M4T13167G  M4T13180G  M4T13183G  M4T13194G  M4T13227G  M4T46110G  SUSGNM47115  RT105453509GA  RT105453544GA  SCS06500651GA  SCS07181135GA  SBB07184673GA  SBB07234864GA  SBB07234865GA  SBB07234870GA  SBB07234875GA  SBB07234876GA  SBB07234878GA  SBB07234880GA  SBB07234882GA  SBB07234885GA  SBB07234887GA  SBB07234890GA  SBB07234891GA  SBB07244937GA  SBB07244942GA  SBB07254960GA  SBB07254962GA  SBB07254974GA  SBB07254975GA  SBB07254980GA  SBB07254983GA  SBB07254985GA  SBB07254994GA  SBB07254995GA  SBB07254998GA  SBB07255001GA  SCS07311413GA  SCS07311421GA  SBB07325331GA  SBB06180432GA  SBB07214829GA  SBB07345435GA  SBB07345439GA  SBB07345451GA  SBB07345457GA  SBB07345461GA  SBB07345462GA  SBB07345471GA  SBB07345481GA  SBB07385490GA  SBB07385498GA  SBB07385501GA  SBB07385505GA  SBB07385506GA  SBB07385507GA  SBB07395521GA  SBB07395530GA  SBB07395532GA  SBB07395539GA  SBB07395543GA  SBB07395549GA  SBB07395550GA  SBB07395570GA  SBB07395575GA  SBB07405591GA  SBB07405613GA  SBB07405614GA  SCS08082345GA  SCS09283835GA  B3TW4455G  B3TW4460G  B3TW4466G  D3T16283G  D3T16327G  D3T16347G  D3T16350G  D3T16358G  D3T16363G  D3T16377G  D3T16379G  D3T16383G  D3T16394G  D3T16446G  D3T16449G  D3T16459G  D3T16465G  E3T16494G  E3T16560G  E3T16581G  E3T16588G  E3T16596G  E3T16598G  E3T16609G  C3T16156G  C3T16159G  D3T16259G  D3T16263G  D3T16275G  D3T16307G  D3T16308G  D3T16310G  D3T16316G  D3T16318G  D3T16321G  D3T16340G  D3T16356G  D3T16357G  D3T16375G  E3T16551G  E3T16553G  E3T16554G  E3T16582G  E3T16603G  E3T16612G  M3T18575G  M3T18577G  M3T18578G  M3T18582G  M3T18587G  M3T18595G  B4T19207G  SAS06196716PA  SAS06216794PA  SCS06480552GA  SBB07466014GA  SBB08497860GA  SBB08497867GA  SBB08497870GA  SBB08497871GA  SBB08497875GA  SBB08497876GA  SBB08497877GA  SBB08497878GA  SBB08497897GA  SBB08497899GA  SCS08493382GA  SCS08493412GA  SBB09258365GA  SCS09283834GA  SCS09293853GA  SCS09293854GA  SCS09293857GA  SAS08392774PA  SCS09404163GA  RST10381868GA  RST10381888GA  RST10381890GA  RST10381898GA  RST10381900GA  SBB10309759GA  SBB10319769GA  SBB10319770GA  SBB10319772GA  SBB10319776GA  RST10391959GA  SBB10149402GA  L1TW1076G  L1TW1079G  L1TW1136G  L1TW1147G  L1TW1154G  L1TW1186G  L1TW1192G  L1TW1193G  H1TW0021G  H1TW0069G  H1TW0072G  H1TW0084G  G2TW2998G  G2T13439G  G2T13464G  G2T13494G  G2T13533G  G2T13535G  G2T13536G  G2T13538G  G2T13541G  G2T13543G  G2T13544G  G2T13547G  G2T13552G  A4T18869G  A4T18892G  A4T18907G  A4T19078G  A4T19121G  A4T19123G  A4T19137G  C4T44640G  C4T44641G  C4T44642G  C4T44644G  C4T44645G  C4T44656G  C4T44657G  F4T45119G  F4T45122G  F4T45124G  F4T45125G  F4T45127G  J4T45414G  SUSGNN43693  R3S04530230GA  RT405020252GA  RT405412029GA  RT405412032GA  RT405502636GA  RT405502664GA  RT405502674GA  RT405502681GA  RT405502682GA  RT405502725GA  RT405502726GA  RT405502728GA  RT405512780GA  RT405512790GA  RT405512793GA  RT405512794GA  RT405512810GA  RT406033048GA  RT406053190GA  RT406053194GA  RT406173772GA  RT406334376GA  RT406334377GA  RT406334384GA  RT406334390GA  RT406344393GA  RT406344435GA  RT406354500GA  RT406354502GA  RT406354503GA  RT406404504GA  RT406404506GA  RT406404507GA  RT406404509GA  RT406404510GA  RT406404511GA  RT406404513GA  RT406404514GA  RT406404515GA  RT406404516GA  RT406404525GA  RT407517608GA  RT407517611GA  RT407055584GA  RT407517597GA  RT407517620GA  RT407527633GA  RT407527637GA  RT407527643GA  RT407527645GA  RT407527646GA  RT408027648GA  RT408027650GA  RT408027652GA  RT408047654GA  RT408047656GA  RT408047658GA  RT408047662GA  RT408047666GA  RT408047664GA  RT408358178GA  RT408358183GA  RT408358185GA  RT408378226GA  RT408378230GA  RT408378241GA  RT408378243GA  RT408378244GA  RT408378254GA  RT408378255GA  RT408378256GA  RT408378259GA  RT408378260GA  RT408378261GA  RT408378262GA  RT408378267GA  RT408378269GA  RT408378270GA  RT408378274GA  RT410109169GA  RT410109174GA  RT410109175GA  M2T14867G  M2T14910G  M2T14928G  M2T14958G  M2T14973G  M2T14979G  M2T14983G  B2T11968G  B2T11981G  C2T12345G  C2T12366G  C2T12370G  C2T12373G  C2T12394G  C2T12400G  C2T12454G  RT106054991GA  RT409398972GA  E3T16734G  F3T16786G  F3T16800G  F3T16815G  F3T16817G  F3T16831G  F3T16834G  F3T16870G  F3T16874G  F3T16879G  F3T16919G  F3T16921G  F3T16923G  F3T16925G  F3T16937G  F3T16945G  F3T16965G  F3T16966G  F3T16971G  F3T16972G  F3T16975G  L4T12780G  RT105010221GA  RT105020255GA  RT105020256GA  RT105020260GA  RT405030274GA  RT405040333GA  RT105030338GA  RT105140976GA  RT105251872GA  RT105221588GA  RT405221041GA  RT405221045GA  RT405221049GA  RT405321400GA  RT405331444GA  RT405452250GA  RT405452258GA  RT405452262GA  RT405523010GA  RT406033108GA  RT406033114GA  RT406063266GA  RT406063271GA  RT406073274GA  RT406073281GA  RT406073283GA  RT406073284GA  RT406073285GA  RT406073286GA  RT406073287GA  RT406073288GA  RT406073289GA  RT406073290GA  RT406073291GA  RT406073292GA  RT406073293GA  RT406073294GA  RT406073295GA  RT406073296GA  RT406073298GA  RT406073300GA  RT406073301GA  RT806080111GA  RT806080112GA  RT806080113GA  RT406103490GA  RT406103494GA  RT406103495GA  RT406163642GA  RT406163656GA  RT406163658GA  RT406284074GA  RT406243940GA  RT406263994GA  RT406264003GA  RT406404552GA  RT406404554GA  RT406404556GA  SCS06430351GA  SCS06440369GA  RT406434651GA  RT406484967GA  RT406484972GA  RT406484994GA  RT406485016GA  RT406485034GA  RT406485038GA  RT406485078GA  RT406485087GA  RT406485088GA  RT406485090GA  SCS06480514GA  SCS06480522GA  SCS06480524GA  SCS06480532GA  SCS06480535GA  SCS06480536GA  SCS06480537GA  SCS06480540GA  SCS06480545GA  SCS06480546GA  RT407075636GA  RT407075638GA  RT407075642GA  RT407085680GA  RT407085681GA  RT407125916GA  RT407125964GA  RT407125906GA  RT407125970GA  RT407125997GA  RT407126005GA  RT407126010GA  RT407126013GA  RT407126016GA  RT407136029GA  RT407136042GA  RT407136044GA  RT407136047GA  RT407136048GA  RT407136049GA  RT407136053GA  SCS07080895GA  SCS07080902GA  SCS07090912GA  SCS07171075GA  RT407176231GA  RT407196311GA  RT407236488GA  RT407236490GA  RT407246504GA  SCS07391700GA  SCS07462006GA  RT407367168GA  RT407517571GA  RT407517596GA  RT407517601GA  RT407517604GA  RT407517617GA  SCS07522160GA  SCS07522162GA  SCS07522172GA  SCS07522176GA  SCS07522177GA  SCS07522178GA  RT408247913GR  RT408247916GR  RT408257927GR  RT408257950GR  RT408257956GR  RST10119789GA  RST10119814GA  RT410139259GA  RT410139265GA  RT410139270GA  RST10442485GA  E2T40976G  E2T41076G  e2T41023G  e2t41071g  F2T13252G  G2T13314G  G2T13342G  G2T13350G  A3T15433G  C3T16230G  C3T16236G  D3T16264G  J3T17753G  L3T18188G  A4T18843G  A4T18844G  A4T18846G  A4T18847G  A4T18848G  A4T18850G  A4T18856G  A4T18857G  A4T18858G  A4T18859G  A4T18860G  A4T18861G  A4T18865G  A4T18866G  A4T18867G  A4T18983G  A4T19044G  A4T19063G  A4T19070G  K4T12393G  K4T12429G  K4T12442G  R3S04430008GA  R3S04430010GA  R3S04430011GA  RT105262008GA  RT105433276GA  RT105433286GA  SBB06230815GA  SBB06230817GA  SBB06230820GA  SBB06230822GA  SBB06230823GA  SBB06230826GA  SBB06230829GA  SBB06230830GA  SBB06230831GA  SBB06230832GA  SBB06230833GA  SBB06230835GA  SBB06230839GA  SCE06250079GA  SCE06520221GA  SCE07010222GA  SCE07010226GA  SCE07130228GA  SCE07130229GA  SCE07130230GA  SCE07130231GA  SCE07130232GA  SCE07130233GA  SCE07130234GA  SCE07130235GA  SCE07130236GA  SCE07130237GA  SCE07130238GA  SCE07130239GA  SCE07130240GA  SCE07130241GA  SCE07130242GA  SCE07130243GA  SCE07130244GA  SCE07130245GA  SCE07130246GA  SCE07130248GA  SCE07130249GA  SCE07130251GA  SCE07130252GA  SCS07090910GA  SCS07090914GA  SCS07090925GA  SCS07090927GA  RTW07041930GA  SBB07455977GA  SBB07465979GA  SBB07465981GA  SBB07476040GA  SBB07476042GA  SBB08126433GA  SBB08136494GA  SBB08236792GA  SBB08246851GA  SBB08236813GA  SBB08246854GA  SBB08357288GA  SBB08307131GA  SBB08387435GA  SBB08387474GA  SBB08437655GA  SBB08437671GA  RT408268061GA  SBB08267058GA  SBB08337218GA  SBB08337236GA  SBB08337243GA  SBB08357270GA  SBB08377339GA  SBB08387443GA  SBB08387449GA  SBB08387452GA  SBB08387453GA  SBB08387464GA  SBB08387465GA  SBB08387472GA  SBB08387478GA  SBB08387482GA  SBB08387483GA  SBB08387484GA  SBB08387488GA  SBB08387489GA  SBB08387491GA  SBB08387496GA  SBB08387498GA  SBB08387501GA  SBB08387503GA  SBB08387505GA  SBB08387508GA  SBB08387522GA  SBB08497814GA  SBB07395540GR  SBB08467765GA  SBB08467782GA  SBB08467798GA  SCS08493337GA  SCS08493344GA  SCS08493393GA  SCS08493395GA  SBB08497898GA  SCS08493392GA  SCS08493406GA  SCS08493408GA  SCS08493409GA  SBB08517990GA  SBB09078031GA  SBB09088048GA  SBB09238292GA  SBB09238294GA  SBB09238295GA  SBB09238301GA  SBB09238304GA  SBB09238309GA  SBB09238310GA  SBB09238311GA  SBB09238312GA  SBB09238313GA  SBB09238316GA  SBB09238317GA  SBB09238318GA  SBB09238319GA  SBB09238320GA  SBB09238323GA  SBB09468665GA  SBB09468667GA  RST10109647GA  RST10109649GA  RST10109733GA  SCS10074423GA  SCS10074424GA  SCS10074428GA  SAT10211283GA  SAT10241288GA  RST10059307GA  RT410309515GA  RT410329528GA  RST10442460GA  RST10442465GA  RST10442466GA  RST10442467GA  RST10442468GA  RST10442471GA  RST10442475GA  RST10442477GA  RT410419594GA  RT410429620GA  RT410449790GA  RT410449793GA  RT410449794GA  RT410449798GA  RT410449908GA  RT410449909GA  L1TW1097G  L1TW1103G  L1TW1105G  M1TW1290G  M1TW1307G  M1TW1311G  M1TW1318G  K1TW0992G  L1TW1075G  M1TW1270G  M1TW1289G  M1TW1295G  M1TW1299G  M1TW1300G  M1TW1304G  L1TW1148G  M1TW1276G  M1TW1279G  M1TW1284G  M1TW1288G  M1TW1293G  M1TW1294G  M1TW1296G  M1TW1301G  M1TW1303G  M2T14850G  K2T14296G  K2T14423G  K2T14446G  K2T14453G  F9HB0293G  F9HB0297G  G9HB0649G  G9HB0735G  H9HB0777G  A1HB7183G  B1HB7873G  C1HB7997G  C1HB8001G  C1HB8015G  C1HB8023G  B3T15558G  J3T17751G  K3T17835G  M3T18747G  M3T18770G  M3T18789G  A4T18929G  L3T18234G  C4T19392G  C4T19405G  C4T19406G  C4T19412G  C4T19413G  C4T19414G  RT105463624GA  RT105463627GA  RT105483639GA  RT105483671GA  RT105483672GA  RT105483673GA  RT105483729GA  RT105483730GA  RT105483732GA  RT105483733GA  RT105483736GA  RT105483741GA  RT105483742GA  RT105483745GA  RT105483748GA  SBS06050023GA  SCS06480507GA  SCS06480534GA  SCS06480557GA  SCS06480563GA  SCS06480566GA  SCS07020733GA  SCS07030741GA  SBB07214850GA  SBB07234867GA  SBB07244915GA  SBB07254965GA  SCE07240262GA  SBB07265040GA  SCS07291401GA  SBB07435881GA  SBB07465989GA  SBB07496079GA  SBB07345430GA  SBB07405629GA  SBB07465986GA  SBB08096246GA  SBB08106269GA  SBB08116367GA  SBB08126443GA  SBB08126445GA  SCS09133654GA  SCS09133660GA  SCS09153690GA  SCS09153694GA  SCS09293849GA  J1TW0400G  J1TW0404G  J1TW0432G  J1TW0461G  J1TW0523G  J1TW0525G  J1TW0526G  J1TW0527G  J1TW0530G  J1TW0532G  J1TW0547G  J1TW0657G  J1TW0658G  J1TW0663G  J1TW0664G  J1TW0665G  J1TW0667G  J1TW0669G  J1TW0670G  J1TW0671G  J1TW0672G  J1TW0673G  J1TW0677G  J1TW0678G  J1TW0679G  J1TW0682G  J1TW0684G  J1TW0686G  J1TW0688G  J1TW0691G  J1TW0694G  J1TW0695G  J1TW0696G  J1TW0698G  J1TW0700G  J1TW0702G  J1TW0704G  J1TW0705G  J1TW0706G  J1TW0707G  J1TW0713G  J2T13794G  J2T13799G  J2T13830G  J2T13837G  H2T13594G  H2T13607G  H2T13655G  H2T13693G  H2T13730G  H2T13742G  H2T13744G  H2T13752G  H2T13763G  H2T13767G  H2T13785G  B2T11671G  D2T12575G  D2T12586G  D2T12588G  D2T12605G  D2T12634G  D2T12639G  D2T12651G  D2T12655G  D2T12667G  D2T12669G  D2T12676G  D2T12684G  D2T12689G  D2T12696G  D2T12703G  D2T12710G  D2T12712G  D2T12728G  D2T12729G  D2T12731G  D2T12747G  D2T12758G  D2T12760G  D2T12761G  D2T12762G  D2T12766G  D2T12777G  D2T12786G  D2T12789G  D2T12801G  D2T12804G  D2T12810G  D2T12818G  D2T12826G  D2T12833G  D2T12835G  D2T12840G  D2T12861G  D2T12862G  D2T12863G  D2T12868G  D2T12876G  D2T12877G  D2T12881G  E2T12899G  E2T12908G  E2T12917G  E2T12920G  E2T12925G  E2T12926G  E2T12934G  E2T12956G  K3T18104G  RTW08322703GA  RTW08322705GA  RTW08322715GA  RTW08322719GA  RTW08322726GA  RTW08322727GA  RTW08322729GA  RTW08322731GA  RTW08372827GA  RTW08372828GA  RTW08372834GA  RTW08372837GA  RTW08372838GA  RTW08372839GA  RTW08372840GA  RTW08372841GA  RTW08372843GA  RTW08372844GA  RTW08372845GA  RTW08372846GA  RTW08372847GA  RTW08372849GA  SAS10285494PA  SCS07291390GA  SBB09498770GA  RT410109167GA  RT410109168GA  RT410109172GA  RT410109173GA  RT410109176GA  RT410119191GA  RST10250715GA  RST10250720GA  RST10270859GA  SBB10259622GA  SBB10259628GA  SBB10259633GA  D4T44747G  D4T44755G  D4T44760G  D4T44813G  D4T44817G  D4T44820G  G4T11536G  G4T11544G  G4T11555G  F4T11416G  F4T45044G  G4T45143G  G4T45148G  G4T45149G  G4T45153G  G4T45155G  G4T45159G  G4T45162G  G4T45163G  G4T45164G  G4T45165G  G4T45166G  G4T45170G  G4T45179G  G4T45180G  G4T45181G  G4T45182G  G4T45184G  G4T45186G  G4T45187G  G4T45188G  G4T45189G  G4T45190G  G4T45191G  G4T45193G  G4T45194G  G4T45195G  G4T45196G  G4T45197G  G4T45199G  G4T45204G  G4T45205G  G4T45212G  G4T45220G  G4T45221G  G4T45223G  G4T45225G  G4T45226G  G4T45228G  G4T45230G  G4T45231G  G4T45233G  G4T45234G  G4T45235G  G4T45236G  G4T45237G  G4T45238G  G4T45240G  G4T45242G  G4T45244G  G4T45245G  G4T45261G  G4T45262G  G4TW5830G  M4T46085G  RT404510034GA  RT404530117GA  RTW04510026GA  RT405160751GA  RT408258013GA  RT408268039GA  RT408338125GA  RT408338140GA  RST10321527GA  RST10321532GA  RT410490028GA  RT410490048GA",
      "more_code_info": ""
    }
  ]
}