  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "67085",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED KINGDOM, IRELAND, IRAQ, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MARITIUS, MALDOVA, NEW ZEALAND, NIGERIA, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, QATAR, RUSSIA, SLOVAKIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, TUNISIA, UNITED ARAB EMIRATES, VENEZUELA, CHILE, CANADA, and BERMUDA.",
      "recall_number": "Z-0807-2014",
      "product_description": "SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient Data Module) Product Usage: The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes.   The Solar 8000i with Patient Data Module/Transport Pro with Patient Data Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional medical facility, such as hospital, clinic, surgical center or doctors office. It can be used in multiple areas such as operating room (OR), post anesthesia care unit (PACU), emergency department (ED), chest pain clinic, general intensive care unit (ICU), critical care unit, surgical intensive care unit (SICU), respiratory intensive care unit, coronary care unit (CCU), medical intensive care unit (MICU), pediatric intensive care unit (PICU), or neonatal intensive care unit (NICU). The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. Physiological parameter data acquired by the PDM includes ECG, invasive pressure, non-invasive blood pressure, pulse oximetry, temperature, cardiac output nad respiration. This device acquires, processes and stores information for all aforementioned parameters and transmits this information to a transport or bedside central processing unit for viewing and alarm surveillance purposes. The Transport Pro patient monitor is intended for use as part of a transport monitoring system for intra-healthcare facility transport. When used with the Patient Data Module (PDM) or the TRAM acquisition module, this device is intended to provide uninterrupted monitoring of physiologic parameter data for adult, pediatric, and neonatal patients during transport from one area of the healthcare facility to another. Physiological parameter data includes ECG, invasive pressure, noninvasive blood pressure, pulse oximetry, temperature and respiration. Both the PDM and TRAM acquisition module acquire, process and store information for all aforementioned parameters. The Solar 8000i patient monitoring system is a multi-parameter physiological patient monitoring system intended for use on adult, pediatric and neonatal patients. It provides uninterrupted monitoring of physiological patient data. The Solar 8000i patient monitoring system is capable of monitoring and analyzing the following parameters for all patient populations: Electrocardiogram, invasive pressure, non-invasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, venous oxygen saturation, transcutaneous pO2 and pCO2, CO2 and respiratory mechanics. The Solar 8000i patient monitoring system is capable of GE Healthcare Page 3 of 6 monitoring the following parameters for adult and pediatric patient populations: anesthetic agent concentrations, O2, impedance cardiography, electroencephalography and bispectral index. The Solar 8000i patient monitoring system interfaces with a variety of third-party peripheral medical devices that support serial and /or analog data outputs. Information from these devices can be displayed, trended and stored in the monitoring system. The solar 8000i patient monitoring system also provides physiological data over the Unity Network.",
      "product_quantity": "2516 (2,321  US, 195 OUS)",
      "reason_for_recall": "Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.",
      "recall_initiation_date": "20131205",
      "center_classification_date": "20140122",
      "termination_date": "20160725",
      "report_date": "20140129",
      "code_info": "Serial Number  SA308044521GR  SA308186360GA  SA310444859GA  SA308186380GA  SA312091241GA  SA308491207GA  SA308186401GA  SA308186385GA  SA308186374GA  SA308155957GA  SA308186391GA  SA308186366GA  SA308186410GA  SA308094871GA  SA308094818GA  SA308186445GA  SA308186381GA  SA312479810GA  SA308094940GA  SA308094803GA  SA308094872GA  SA308084770GA  SA310528992GA  SA310506733GA  SA310506696GA  SA311153223GA  SA311142931GA  SA312080647GA  SA312080894GA  SA308186396GA  SA308166121GA  SA308186397GA  SA308094807GA  SA308186404GA  SA311153169GA  SA308186377GA  SA308398894GA  SA312253398GA  SA308094809GA  SA309263954GA  SA310048616GA  SA309527962GA  SA310038386GA  SA310048604GA  SA310048613GA  SA310028273GA  SA310038462GA  SA308317172GA  SA311184814GA  SA309365237GR  SA311142946GA  SA307332233GA  SA307231341GA  SA312069693GA  SA309456714GA  SA311163858GA  SA312049242GA  SA311317628GA  SA307423146GA  SA310259886GA  SA309142960GA  SA309334557GA  SA307393020GA  SA307453540GA  SA307210827GA  SA312018863GA  SA307241567GA  SA311163574GA  SA312039090GA  SA312080635GA  SA309263938GA  SA309304244GR  SA309233517GA  SA312039205GA  SA312080692GA  SA308347813GA  SA308398654GA  SA307504161GA  SA312059549GA  SA308491112GA  SA312059550GA  SA309021966GA  SA310351543GA  SA309193340GA  SA308186334GA  SA311101421GA  SA310444513GA  SA311518717GA  SA312122048GA  SA312029011GA  SA309203388GA  SA309344857GA  SA312028978GA  SA312253251GA  SA312121695GA  SA311508589GR  SA312152342GA  SA310517769GA  SA310517770GA  SA311297355GA  SA311163926GA  SA312029028GA  SA308419235GA  SA312069661GA  SA311458372GA  SA307382560GA  SA311428153GA  SA308511698GA  SA308511500GA  SA308368183GA  SA309223482GA  SA310361769GA  SA312039183GA  SA312079967GA  SA308491092GA  SA311458333GA  SA310393014GA  SA310507045GA  SA310517560GA  SA310331323GA  SA311163587GA  SA311184495GA  SA307291912GA  SA310393006GA  SA311060085GA  SA309112590GA  SA310361759GA  SA311194946GA  SA309112586GA  SA309112577GA  SA309112584GA  SA309112588GA  SA309112583GA  SA309112582GA  SA309112591GA  SA309112576GA  SA309112569GA  SA308166208GA  SA308084740GA  SA309112585GA  SA309112578GA  SA309112592GA  SA309112572GA  SA309112589GA  SA309112587GA  SA309112580GA  SA309112579GA  SA309122666GA  SA309032010GA  SA309042195GA  SA308459639GA  SA308449596GA  SA311080599GA  SA309092487GA  SA309092489GA  SA308145819GA  SA308521831GA  SA309032015GA  SA309032018GA  SA309032019GA  SA309032020GA  SA308459670GA  SA309082457GA  SA308501419GA  SA308501431GA  SA308501401GA  SA308511444GA  SA308481061GA  SA309062308GA  SA309062315GA  SA309062313GA  SA309062309GA  SA309062310GA  SA309062305GA  SA309062300GA  SA309082454GA  SA309082456GA  SA312469178GA  SA309062269GA  SA309021949GA  SA309062277GA  SA308511441GA  SA308511440GA  SA308511477GA  SA309082443GA  SA309092491GA  SA309092492GA  SA308327318GA  SA308491248GA  SA309062318GA  SA309062296GA  SA309062304GA  SA309042201GA  SA309042207GA  SA308501411GA  SA309062336GA  SA309092493GA  SA309082452GA  SA311174002GA  SA308317141GA  SA309062298GA  SA310518235GR  SA308439427GA  SA308398687GA  SA308429319GA  SA308429314GA  SA308429310GA  SA308439425GA  SA308439442GA  SA308439435GA  SA308388629GA  SA308419275GA  SA308439462GA  SA308429311GA  SA308439431GA  SA308429313GA  SA308429316GA  SA308419276GA  SA308419277GA  SA308439434GA  SA308439426GA  SA308439445GA  SA308439461GA  SA308439444GA  SA308429320GA  SA308439460GA  SA308439433GA  SA308429309GA  SA308429317GA  SA308429364GA  SA308419278GA  SA308439439GA  SA311194970GA  SA311184743GA  SA311195201GA  SA311184673GA  SA309011857GA  SA311287076GA  SA310149349GA  SA308459836GA  SA310118991GA  SA310068803GA  SA308317131GA  SA311428022GA  SA308317113GA  SA308266854GA  SA308409143GA  SA312049299GA  SA308286919GA  SA311488464GA  SA309426162GA  SA310507288GA  SA310403433GA  SA308388490GA  SA307231395GA  SA307453573GA  SA307453582GA  SA307504151GA  SA307483662GA  SA307483670GA  SA307504093GA  SA307504144GA  SA307382578GA  SA307504138GA  SA307251801GA  SA307504094GA  SA307504186GA  SA307504090GA  SA307504092GA  SA307483659GA  SA307504136GA  SA307504152GA  SA307504217GA  SA307504153GA  SA307453611GA  SA307504145GA  SA307483729GA  SA307504219GA  SA307504143GA  SA307382576GA  SA308317143GA  SA311236270GA  SA310434125GA  SA308511624GA  SA311184701GA  SA310310864GA  SA311184767GA  SA309517571GA  SA308358004GA  SA312049462GA  SA308511514GA  SA309365225GR  SA312470025GA  SA312039138GA  SA310517610GR  SA308388529GA  SA310434088GR  SA307251710GR  SA311090936GA  SA311256738GA  SA311195187GA  SA309193309GA  SA311246669GA  SA307231343GA  SA309253845GA  SA308125335GA  SA311184663GA  SA311518704GA  SA307342306GA  SA311428126GA  SA312294770GA  SA309203427GA  SA311205398GA  SA310361897GA  SA311205412GA  SA311195189GA  SA311327842GA  SA309477080GA  SA308459750GA  SA308196503GA  SA307241697GA  SA312080840GA  SA311142745GA  SA312049392GA  SA312152411GA  SA311448177GA  SA311174149GA  SA310518390GR  SA311184749GA  SA308266860GA  SA308286916GA  SA311069715GA  SA308489979GA  SA311184627GA  SA310475501GA  SA311132620GA  SA311184638GA  SA308266876GA  SA311184652GA  SA312069674GA  SA312069679GA  SA308398784GA  SA312080833GA  SA310048683GA  SA307291863GA  SA307261813GA  SA307302000GA  SA307291872GA  SA307301998GA  SA307261839GA  SA312091110GA  SA312091063GA  SA312091091GA  SA312121813GA  SA312091078GA  SA312091036GA  SA312091084GA  SA312091090GA  SA312091072GA  SA312091088GA  SA312091086GA  SA312091111GA  SA312091096GA  SA307261841GA  SA307261811GA  SA307291862GA  SA309021970GA  SA312294497GA  SA311287126GA  SA307301994GA  SA307352370GA  SA307261845GA  SA307382545GA  SA307301993GA  SA307504176GA  SA307251759GA  SA307291900GA  SA307382518GA  SA308357963GA  SA308084720GR  SA307291864GA  SA312080727GA  SA311307419GA  SA312079959GA  SA312029045GA  SA309193295GA  SA311428074GA  SA309334794GA  SA310485811GA  SA309375367GA  SA311428095GA  SA311060325GA  SA308206731GA  SA308388524GA  SA311069845GA  SA308357993GA  SA312468536GA  SA311184807GA  SA309193314GA  SA311195219GA  SA309142932GA  SA311184657GA  SA312091221GA  SA311163934GA  SA310393048GA  SA312070365GA  SA309365229GR  SA307291883GA  SA310361794GA  SA307443325GA  SA308166147GA  SA309112555GA  SA311246649GA  SA310444266GA  SA308095032GA  SA308176282GA  SA307231363GA  SA308196557GA  SA308317092GA  SA312069839GA  SA308135615GR  SA311518709GA  SA311163617GA  SA308491110GA  SA311184580GA  SA311184555GA  SA310139045GA  SA309122626GA  SA311184775GA  SA310219791GA  SA311327933GA  SA311132276GA  SA308368065GA  SA310372059GA  SA309334560GA  SA312069651GA  SA312029022GA  SA311122215GA  SA311307560GA  SA311428082GA  SA309355014GA  SA311060295GR  SA308266845GA  SA311174196GA  SA311090932GA  SA311090941GA  SA308459749GA  SA311307556GA  SA311163718GA  SA312080707GA  SA308388614GA  SA311317638GA  SA312070158GA  SA311337974GA  SA312080624GA  SA312080631GA  SA312080561GA  SA309243678GA  SA309193357GA  SA309334577GA  SA309334566GA  SA309334562GA  SA309324425GA  SA309334650GA  SA309334597GA  SA309243697GA  SA309243693GA  SA309243687GA  SA309243686GA  SA309243685GA  SA309193346GA  SA309193331GA  SA309193329GA  SA309446534GA  SA311317648GA  SA310475634GA  SA312080874GA  SA312080701GA  SA312121711GA  SA312049469GA  SA309233520GA  SA312080824GA  SA311518705GA  SA312355402GA  SA311428161GA  SA312029016GA  SA308286922GA  SA309538046GA  SA309334575GA  SA309334664GA  SA309193343GA  SA308155975GA  SA312152331GA  SA309193362GA  SA309324321GA  SA309324317GA  SA309294071GA  SA309294053GA  SA309243680GA  SA309193336GA  SA309193318GA  SA309193312GA  SA312049463GA  SA312049457GA  SA311060218GA  SA307504201GA  SA312070182GA  SA307140631GA  SA312090928GA  SA312101445GA  SA311174194GA  SA308327455GA  SA309243679GA  SA309243677GA  SA309203393GA  SA309334563GA  SA309324445GA  SA309324437GA  SA309334721GA  SA309334604GA  SA309294028GA  SA309334587GA  SA309243689GA  SA309243684GA  SA309193352GA  SA308095013GA  SA310496469GA  SA312121928GA  SA309324438GA  SA309334598GA  SA311428097GA  SA312069863GA  SA310361762GA  SA311428127GA  SA312111487GA  SA312121926GA  SA309334591GA  SA309477022GR  SA309304234GA  SA312152408GA  SA310219829GA  SA312121902GA  SA311184728GA  SA312091197GA  SA311518716GA  SA307332213GA  SA311090787GA  SA310372063GA  SA311307542GA  SA312091207GA  SA307483808GA  SA310372204GA  SA310424046GA  SA312091201GA  SA307493994GA  SA312080779GA  SA312080769GA  SA312080556GA  SA312080763GA  SA312080690GA  SA312080596GA  SA312080608GA  SA312080797GA  SA309334574GA  SA309324413GA  SA309294090GA  SA309294058GA  SA309294050GA  SA309334596GA  SA309334593GA  SA309334592GA  SA309294029GA  SA309253840GA  SA312091037GA  SA308491186GA  SA312028916GA  SA309324443GA  SA309334701GA  SA309334652GA  SA311174208GA  SA309243682GA  SA308166162GA  SA311256789GA  SA310485804GA  SA310393011GA  SA310259875GA  SA309334621GA  SA312059519GA  SA311184867GA  SA312091335GA  SA308317185GA  SA307140637GA  SA311256730GA  SA312049317GA  SA311184757GA  SA311174007GA  SA311256832GA  SA308145749GA  SA310260055GA  SA311287093GA  SA311256751GA  SA309385605GA  SA311111641GA  SA312069713GA  SA309334564GA  SA309193337GA  SA310361764GA  SA312111513GA  SA308459822GA  SA308409076GA  SA312070304GA  SA312079968GA  SA309243669GA  SA309193356GA  SA309334571GA  SA309334569GA  SA309334561GA  SA309324442GA  SA309294089GA  SA309294067GA  SA309294061GA  SA309294060GA  SA309334665GA  SA309334642GA  SA309294057GA  SA309294051GA  SA309294047GA  SA309334588GA  SA309253750GA  SA309334579GA  SA309193339GA  SA309324394GA  SA309334753GA  SA309334747GA  SA309334713GA  SA309334705GA  SA309294068GA  SA309294065GA  SA309294055GA  SA309263943GA  SA309253909GA  SA309253715GA  SA309334583GA  SA311428007GA  SA312069714GA  SA311307552GA  SA311307549GA  SA311246592GA  SA311327846GA  SA311307493GA  SA311307466GA  SA311317640GA  SA311297246GA  SA311327930GA  SA311307499GA  SA312091286GA  SA312192705GA  SA312253361GA  SA307241684GA  SA307443387GA  SA309132846GA  SA307210819GR  SA312029046GA  SA311337982GA  SA311297337GA  SA312080659GA  SA312080783GA  SA312080708GA  SA312253104GA  SA311287153GA  SA312091196GA  SA311428076GA  SA311297378GA  SA312080901GA  SA311184499GA  SA308501276GA  SA310028197GA  SA311205464GA  SA308084662GA  SA308084646GA  SA310507326GA  SA311060119GA  SA311195041GA  SA310260047GA  SA312080558GA  SA312080542GA  SA312080744GA  SA312080695GA  SA311090765GA  SA311205445GA  SA311060302GA  SA308084652GA  SA308084664GA  SA312090970GA  SA311070537GA  SA312049356GA  SA311287066GA  SA311163416GA  SA311070574GA  SA311070588GA  SA311060141GA  SA311070542GA  SA308155857GA  SA308084619GA  SA308084653GA  SA312018824GA  SA312070164GA  SA312070206GA  SA312070377GA  SA312070366GA  SA312070141GA  SA312070301GA  SA310139188GA  SA311327902GA  SA311327922GA  SA312029043GA  SA310028187GA  SA311060150GA  SA311060158GA  SA311060296GA  SA311069729GA  SA311528738GA  SA311070593GA  SA311060182GA  SA312069653GA  SA310028291GA  SA312029056GA  SA311060285GA  SA308094965GA  SA311090854GA  SA308459790GA  SA311246507GA  SA311090762GA  SA311101184GA  SA311101217GA  SA311060206GA  SA311069714GA  SA311060139GA  SA311060124GA  SA311070523GA  SA311101285GA  SA311060199GA  SA311060278GA  SA311060176GA  SA311060152GA  SA308084658GA  SA308084612GA  SA310455264GA  SA308429334GA  SA311428016GA  SA310219844GA  SA312090934GA  SA311194883GA  SA312080570GA  SA309223484GA  SA311205357GA  SA311142721GA  SA308317134GA  SA308286913GA  SA311317606GA  SA311428010GA  SA311194914GA  SA311122125GA  SA311317729GA  SA310423962GA  SA312039098GA  SA311428049GA  SA308398752GA  SA308398756GA  SA308317189GA  SA311307597GA  SA311059603GA  SA310290660GA  SA311090901GA  SA312070233GA  SA312070380GA  SA312070378GA  SA312090921GA  SA312091011GA  SA312070334GA  SA312059533GA  SA311297340GA  SA312070294GA  SA311297395GA  SA311297383GA  SA311256855GA  SA311256734GA  SA308501408GA  SA308266821GA  SA312070034GA  SA311195239GA  SA311194908GA  SA311194945GA  SA312263785GA  SA310507387GA  SA310455228GA  SA310455227GA  SA312111486GA  SA309011850GA  SA312070367GA  SA312070211GA  SA310382552GA  SA311256815GA  SA310507201GA  SA311297407GA  SA310507379GA  SA310260009GA  SA310118970GA  SA311184745GA  SA311317629GA  SA311184878GA  SA312029002GA  SA311121817GA  SA308491227GA  SA311297301GA  SA310199713GA  SA310199589GA  SA311327926GA  SA311338002GA  SA311297223GA  SA311297370GA  SA310382584GA  SA308074605GA  SA310199704GA  SA310260007GA  SA310199595GA  SA310260021GA  SA310260027GA  SA310260001GA  SA310260082GA  SA310139260GA  SA308378261GA  SA310454928GA  SA311101068GA  SA311317635GA  SA311307543GA  SA311297376GA  SA311297405GA  SA310260006GA  SA310260067GA  SA310259975GA  SA312079934GA  SA309446464GA  SA312090978GA  SA309375473GR  SA310518711GA  SA308398721GA  SA310199650GA  SA310260017GA  SA310260079GA  SA310199664GA  SA310199655GA  SA310260049GA  SA310199752GA  SA312070297GA  SA312070236GA  SA310465312GA  SA310139254GA  SA310260034GA  SA310199651GA  SA310260072GA  SA310260057GA  SA310260078GA  SA310260003GA  SA310260026GA  SA311428055GA  SA312070157GA  SA312070244GA  SA310139075GA  SA310260030GA  SA310199644GA  SA311297387GA  SA310199653GA  SA308501282GA  SA309426069GA  SA311163948GA  SA310199680GA  SA310260018GA  SA310139073GA  SA310139155GA  SA309426099GA  SA309446456GA  SA309446445GA  SA310260028GA  SA312070273GA  SA311184787GA  SA310382581GA  SA310372104GA  SA312070210GA  SA312070225GA  SA311194900GA  SA310517583GA  SA312028980GA  SA311246411GA  SA310260011GA  SA311317608GA  SA310199673GA  SA310199642GA  SA310199703GA  SA310260020GA  SA310139245GA  SA310139264GA  SA309365095GA  SA307483834GA  SA311163838GA  SA311287067GA  SA311256821GA  SA311174032GA  SA311184682GA  SA311297403GA  SA311184500GA  SA312152377GA  SA310260019GA  SA310260023GA  SA310260008GA  SA310139070GA  SA310199646GA  SA311428019GA  SA311195212GA  SA309243700GA  SA311297399GA  SA310139287GA  SA310199587GA  SA310260025GA  SA310260022GA  SA310259988GA  SA310260042GA  SA310259973GA  SA310361876GA  SA311266925GA  SA311184794GA  SA311327917GA  SA311163729GA  SA311297384GA  SA310139269GA  SA310199679GA  SA310139256GA  SA310260024GA  SA310199591GA  SA310199741GA  SA310260053GA  SA310139283GA  SA310139280GA  SA308196533GA  SA311287013GA  SA310507292GA  SA311307574GA  SA310199643GA  SA310199656GA  SA310199753GA  SA310260016GA  SA310259976GA  SA310199668GA  SA311287050GA  SA310382593GA  SA311246593GA  SA308501305GA  SA308327379GA  SA311337987GA  SA312070061GA  SA308347725GA  SA308327370GA  SA311101486GA  SA311101546GA  SA310259874GA  SA310259880GA  SA311101532GA  SA311101537GA  SA310310854GA  SA310310861GA  SA310259857GA  SA310199730GA  SA311195151GA  SA310310925GA  SA311101547GA  SA310259889GA  SA310518666GA  SA310199667GA  SA308166072GA  SA311184720GA  SA309032026GA  SA312059489GA  SA309385567GA  SA310528972GA  SA307483768GA  SA308368099GA  SA312070287GA  SA310392998GA  SA309092502GA  SA310518630GA  SA310361647GA  SA310300774GA  SA312070090GA  SA308491130GA  SA310518655GA  SA308186319GA  SA311508575GA  SA311174053GA  SA311215811GA  SA310455257GA  SA310465298GA  SA311256835GA  SA311163563GA  SA308459796GA  SA311297404GA  SA310118984GA  SA311317613GA  SA308469879GA  SA312090983GA  SA308327362GA  SA307494047GA  SA308459761GA  SA308317079GA  SA309324515GA  SA308378270GA  SA310455033GA  SA310260486GA  SA310260474GA  SA310260469GA  SA307423135GA  SA310260473GA  SA310260470GA  SA307493927GA  SA311184417GA  SA308521776GA  SA307392937GA  SA307423131GA  SA308511587GA  SA307403128GA  SA308125326GA  SA309092527GA  SA307443344GA  SA308084754GA  SA310372097GA  SA308491100GA  SA310507176GA  SA308501271GA  SA308501275GA  SA308501309GA  SA308501261GA  SA309112597GA  SA309122636GA  SA308521798GA  SA309112616GA  SA309112613GA  SA309112618GA  SA311184726GA  SA311184817GA  SA307220880GR  SA310028262GA  SA307231352GR  SA309052228GA  SA311174041GA  SA307241510GR  SA307231205GR  SA310506691GA  SA307231414GR  SA307382453GR  SA308044522GR  SA308166052GR  SA307241513GR  SA307231385GR  SA307382587GR  SA310372302GA  SA309112571GR  SA308409073GA  SA308491212GA  SA309385510GA  SA309385554GA  SA309375468GR  SA308145655GR  SA309375506GR  SA308337671GR  SA309334675GR  SA308419237GR  SA307231475GR  SA309092488GR  SA307231257GR  SA307220892GR  SA312070178GA  SA309253716GA  SA309253726GA  SA308409079GA  SA311060194GA  SA312090995GA  SA308409036GR  SA311184795GA.DEL  SA311184815GA  SA311184820GA  SA308459645GA  SA308206740GA  SA308095029GA  SA310351549GA  SA309385640GA  SA311194902GA  SA311256755GA  SA311327889GA  SA311327895GA  SA311327884GA  SA311327898GA  SA311327893GA  SA311327888GA  SA311215727GA  SA311317707GA  SA310028191GA  SA308095070GA  SA308095043GA  SA311184689GA  SA308491194GA  SA308491192GA  SA308095042GA  SA308095041GA  SA308095020GA  SA308095035GA  SA308095040GA  SA308095033GA  SA308095025GA  SA308491225GA  SA308491228GA  SA308095038GA  SA308095017GA  SA308095019GA  SA308095063GA  SA308095058GA  SA308095069GA  SA308095030GA  SA308095018GA  SA309426042GA  TF0265  SA308145834GA  SA307494033GA  SA307392953GA  SA310341406GA  SA308317191GA  SA308145739GA  SA311184828GA  SA308074538GA  SA311184827GA.DEL  SA311163935GA  TF0310  SA308266844GA  SA311246652GA  SA311184862GA  SA309223466GA  SA308501278GA  SA311266962GA  SA310506989GA  SA310361874GA  SA310382686GA  SA309446517GA  SA307453494GA  SA310260051GA  SA310260081GA  SA310260054GA  SA310260041GA  SA310260062GA  SA307382629GA  SA308459759GA  SA310351534GA  SA308166176GA  SA307453575GA  SA310139074GA  SA311195225GA  SA308429324GA  SA307504172GA  SA308398720GA  SA308388543GA  SA309062284GA  SA309062271GA  SA310169523GA  SA308347781GA  SA309385627GA  SA309385538GA  SA307382659GA  SA309385547GA  SA309062267GA  SA309385579GA  SA309385552GA  SA309062299GA  SA309375492GA  SA307504117GA  SA308095000GA  SA308409190GA  SA308511533GA  SA307504207GA  SA311069837GA  SA309062279GA  SA307443260GA  SA308511519GA  SA307504206GA  SA307291920GA  SA307251805GA  SA308398803GA  SA308491177GA  SA309304218GA  SA309385546GA  SA308409192GA  SA308094942GA  SA308398825GA  SA308084759GA  SA307483811GA  SA308398903GA  SA309385556GA  SA307504196GA  SA307493932GA  SA307504133GA  SA308094934GA  SA308429348GA  SA308398683GA  SA307483774GA  SA307504218GA  SA310372310GA  SA311121876GA  SA308449571GA  SA308176245GA  SA308409193GA  SA308398828GA  SA307241529GA  SA310372158GA  SA310139268GA  SA308398795GA  SA310139270GA  SA308166092GA  SA307443365GA  SA307443370GA  SA310139083GA  SA307514325GA  SA308429343GA  SA308398830GA  SA308398855GA  SA308398810GA  SA309122752GA  SA309122765GA  SA309122764GA  SA309132802GA  SA309122778GA  SA309324378GA  SA308398826GA  SA310139080GA  SA308429318GA  SA310372025GA  SA310372280GA  SA310372304GA  SA310372152GA  SA309122771GA  SA310372144GA  SA307514328GA  SA309263937GA  SA309385608GA  SA308439475GA  SA307504203GA  SA309062285GA  SA308347794GA  SA307504198GA  SA308095074GA  SA308206747GA  SA308398813GA  SA308095015GA  SA307514340GA  SA308347788GA  SA308347786GA  SA307504190GA  SA308084783GA  SA307514331GA  SA307504209GA  SA308084758GA  SA307504128GA  SA308206718GA  SA308206736GA  SA307362382GA  SA307231336GA  SA308145822GA  SA310139068GA  SA308501324GA  SA310528862GA  SA309011864GA  SA309385544GA  SA307453411GA  SA309021923GA  SA307504223GA  SA308094978GA  SA309062302GA  SA309122791GA  SA309385553GA  SA309385607GA  SA307504208GA  SA309062303GA  SA308125325GA  SA310372275GA  SA307504200GA  SA309062273GA  SA308115265GA  SA309122784GA  SA307514333GA  SA307493939GA  SA309062278GA  SA307483786GA  SA309062283GA  SA309385584GA  SA308429332GA  SA308084756GA  SA310382504GA  SA308429358GA  SA309517723GA  SA307504220GA  SA310139267GA  SA307514330GA  SA307392988GA  SA308511505GA  SA307453604GA  SA310260350GA  SA307392949GA  SA307453418GA  SA307443239GA  SA308398802GA  SA308398857GA  SA308398812GA  SA309375507GA  SA307504165GA  SA307453598GA  SA309385628GA  SA308115276GA  SA309324366GA  SA308429336GA  SA307453551GA  SA307443253GA  SA307453606GA  SA307443250GA  SA307382574GA  SA307382531GA  SA307382612GA  SA307453391GA  SA307443225GA  SA307453558GA  SA309062282GA  SA307493938GA  SA307504197GA  SA309324407GA  SA310139284GA  SA310139087GA  SA310139288GA  SA310139265GA  SA310219825GA  SA308449572GA  SA309253766GA  SA308511634GA  SA307453414GA  SA308449516GA  SA310068858GA  SA310068880GA  SA310139082GA  SA307382601GA  SA309324406GA  SA309062281GA  SA308511525GA  SA308491249GA  SA308094949GA  SA308084729GA  SA309324419GA  SA309324364GA  SA307514337GA  SA309324435GA  SA309324420GA  SA309324422GA  SA309324370GA  SA309324381GA  SA309324361GA  SA308327421GA  SA308266846GA  SA308429335GA  SA309324362GA  SA308511502GA  SA308491161GA  SA308368167GA  SA308429344GA  SA309324359GA  SA309324388GA  SA309324367GA  SA309324403GA  SA309324436GA  SA309324363GA  SA309324374GA  SA309324379GA  SA309324397GA  SA309324405GA  SA309324421GA  SA309324355GA  SA309324423GA  SA309324390GA  SA309324358GA  SA309324398GA  SA309324372GA  SA308511517GA  SA308266832GA  SA308439476GA  SA310455248GA  SA308317151GA  SA309203425GA  SA308327334GA  SA310517738GA  SA310219784GA  SA308317144GA  SA308166155GA  SA308317108GA  SA308196495GA  SA309477084GA  SA310260521GA  SA307493950GA  SA310475696GA  SA307392847GA  SA308166045GA  SA308206594GA  SA309122726GA  SA309122724GA  SA309122733GA  SA309122721GA  SA310290714GA  SA310485801GA  SA310393036GA  SA307413104GA  SA310423907GA  SA310259977GA  SA310393012GA  SA310260383GA  SA310485788GA  SA310517774GA.DEL  SA308166054GA  SA309142995GA  SA308166211GA  SA309446451GA  SA308327323GA  SA308176308GA  SA308521773GA  SA310038440GA  SA309497150GA  SA310393158GA  SA308368114GA  SA308357994GA  SA311060306GA.DEL  SA311060276GA.DEL  SA307392994GA  SA308155932GA  SA310038343GA  SA309042170GA  SA307332280GR  SA308449529GA  SA310423822GA  SA309253852GA  SA309042150GA  SA308074537GR  SA308511562GA  SA309042173GA  SA308449481GA  SA308449569GA  SA308459862GA  SA308511527GA  SA308419216GA  SA311060166GA  SA310518681GA  SA308459695GA  SA308459864GA  SA308449504GA  SA309042122GA  SA308511526GA  SA308439459GA  SA308449589GA  SA308459696GA  SA308449530GA  SA308459694GA  SA308459844GA  SA308459839GA  SA308459797GA  SA308459841GA  SA308449583GA  SA308459850GA  SA308459674GA  SA308459692GA  SA308459690GA  SA308459866GA  SA308449585GA  SA308449526GA  SA308459869GA  SA308459845GA  SA309032113GA  SA308459863GA  SA308449581GA  SA309042123GA  SA309032106GA  SA309042116GA  SA308459865GA  SA308459643GA  SA308449591GA  SA308449517GA  SA309042169GA  SA309032110GA  SA308449590GA  SA308459691GA  SA309032108GA  SA309042118GA  SA309032105GA  SA308449532GA  SA310382615GA  SA310068824GA  SA308459693GA  SA308449523GA  SA308449574GA  SA308408938GA  SA308439429GA  SA308439452GA  SA308449597GA  SA308449501GA  SA308449490GA  SA308449507GA  SA308409201GA  SA308449524GA  SA307382632GA  SA308084702GA  SA310361644GA  SA307392961GA  SA310507084GA  SA309517651GA  SA310424000GA  SA308196538GA  SA310393281GA  SA308358002GA  SA308266890GA  SA308429373GA  SA307261814GA  SA308419302GA  SA309477004GA  SA310259870GA  SA310259861GA  SA310259860GA  SA310259869GA  SA308449636GA  SA310260365GA  SA309122689GA  SA307382637GA  SA309365073GA  SA307241515GA  SA307241701GA  SA310028223GA  SA308166067GA  SA310290753GA  SA307291907GA  SA307251797GA  SA307251716GA  SA307291914GA  SA310028182GA  SA310028186GA  SA308155868GA  SA307200788GA  SA307251737GA  SA308155846GA  SA308094808GA  SA307221107GA  SA307231397GA  SA307251785GA  SA307231404GA  SA307291935GA  SA307251791GA  SA310028213GA  SA307241740GA  SA307220914GA  SA307241512GA  SA308074604GA  SA307241695GA  SA307291913GA  SA307251795GA  SA310028275GA  SA310028155GA  SA307251800GA  SA307251794GA  SA310028214GA  SA310028160GA  SA309507358GA  SA309527912GA  SA307231244GA  SA308388518GA  SA308186382GA  SA308166083GA  SA308166107GA  SA308347763GA  SA307291911GA  SA308155874GA  SA307291918GA  SA307231212GA  SA307220996GA  SA307261812GA  SA308166074GA  SA307251796GA  SA310028199GA  SA307241578GA  SA307231431GA  SA307251802GA  SA310028265GA  SA307251798GA  SA307251712GA  SA310028266GA  SA308166066GA  SA307231424GA  SA307231406GA  SA307291899GA  SA307291897GA  SA307291910GA  SA308327483GA  SA308155854GA  SA308084751GA  SA308166151GA  SA310290659GA  SA307231310GA  SA310260242GA  SA309517454GA  SA308388365GA  SA307221078GA  SA308337695GA  SA308206604GA  SA307382645GA  SA308501416GA  SA308368133GA  SA310382798GA  SA309517471GA  SA309375379GA  SA309517491GA  SA308481038GA  SA308491136GA  SA308501432GA  SA308489973GA  SA308501430GA  SA308491169GA  SA308501433GA  SA309122659GA  SA309122669GA  SA309122648GA  SA309122649GA  SA309112593GA  SA309122655GA  SA309122665GA  SA309122650GA  SA309122653GA  SA309122647GA  SA309122656GA  SA309122652GA  SA309122671GA  SA309122645GA  SA309122657GA  SA309122646GA  SA309122670GA  SA309122663GA  SA309122661GA  SA309122658GA  SA308186316GA  SA310260396GA  SA310392935GA.DEL  SA308266898GA  SA310169502GA  SA308186357GA  SA308317162GA  SA307504177GA  SA310518690GA  SA307493974GA  SA309062292GA  SA312049302GA  SA310423583GA  SA312080817GA  SA308511676GA  SA308388449GA  SA309253900GA  SA309436305GA  SA308155872GA  SA307322133GA  SA311184477GA  SA308155865GA  SA308327348GA  SA309112574GA  SA307291887GA  SA309477117GA  SA307382563GA  SA307332245GA  SA308489911GA  SA312294779GA  RMF03460161PA  SA309405859GA  SA312059500GA  SA308449502GA  SA312457968GA  SA310219805GA  SA308317199GA  SA310506746GA  SA309517677GA  SA309324519GA  SA308317067GA  SA309375434GA  SA311060245GA  SA311428128GA  SA311287016GA  SA308501319GA  SA309072365GA  SA308501388GA  SA312069674GA.DEL  SA310139218GA  SA310382671GA  SA309203394GA  SA313198220GA  SA309365189GA  SA312049387GA  SA312090928GA.DEL  SA311428087GA  SA310393294GA  SA308166080GA  SA310485810GA  SA308317076GA  SA310300771GA  SA312070057GA  SA312069679GA.DEL  SA310424019GA  SA312091227GA.DEL  SA308419305GA  SA308511495GA  SA311153288GA  SA310518423GR  SA310382724GA  SA312376217GA  SA307382504GA  SA308511696GA  SA310219813GA  SA309446480GA  SA308145686GA  SA308398875GA  SA309092515GA  SA308266819GA  SA311287083GA  SA308166011GA  SA311194965GA  SA309122779GA  SA312152373GA  SA309021915GA  SA311215701GR  SA308186375GA  SA308155989GA  SA309203428GA  SA307423178GA  SA312468961GA  SA307322146GR  SA310058700GA  SA308155965GA  SA307231210GR  SA309011863GA  SA308317246GA  SA312049453GA  SA310454939GA  SA311215851GA  SA312263628GA  SA307493912GA  SA308074579GR  SA311327882GR  SA307332243GA  SA308166149GA  SA310444499GA  SA310382343GA  SA308186388GA  SA307443320GA  SA309011849GA  SA309436300GA  SA312091018GA  SA311132331GA  SA307493925GA  SA308166042GA  SA308166098GA  SA308459748GA  SA308327481GA  SA309203396GA  SA309456674GA  SA308166095GA  SA312059535GA  SA311060023GA  SA312091045GA  SA311287033GA  SA311307511GA  SA307221096GR  SA308337611GA  SA311246639GA  SA307504174GA  SA311458222GA  SA307312108GR  SA311428146GA  SA311153321GA  SA312059516GA  SA308166002GA  SA311488484GA  SA310475517GA  SA310517774GA  SA309344834GA  SA308166010GA  SA312039211GA  SA310506739GA  SA308317093GA  SA307514336GA  SA312080796GA  SA307453454GA  SA310507539GA  SA308155864GA  SA311307578GA  SA308398893GA  SA312122048GA.DEL  SA311163904GA  SA308186350GA  SA308409082GA  SA312111577GA  SA311236323GA  SA310393283GA  SA307423141GA  SA311163598GA  SA308327358GA  SA308337572GA  SA308155840GA  SA307231337GA  SA312039149GA  SA312386385GA  SA312070226GA  SA312091197GA.DEL  SA307382641GA  SA307332224GA  SA311287090GA  SA308327437GA  SA307020391GR  SA312263854GA  SA309436355GA  SA311195011GA  SA308166007GA  SA310444842GA  SA308317238GA  SA308317311GA  SA308135494GA  SA310392943GA  SA310382718GA  SA307382558GA  SA308166062GA  SA310518687GA  SA308166154GA  SA308479900GA  SA308317261GA  SA311111612GA  SA310058708GA  SA309122619GA  SA309477058GA  SA309477000GA  SA312091041GA  SA311122048GA  SA309142962GA  SA308166091GA  SA307251741GR  SA311307579GA  SA310444496GA  SA307332247GA  SA307231410GA  SA310361821GA  SA310199572GA  SA308166108GA  SA309253846GA  SA308166040GA  SA308266868GA  SA308368054GA  SA311142910GA  SA308491104GA  SA308409210GA  SA309183205GA  SA311246650GA  SA312294506GA  SA310423814GA  SA311518703GA  SA308155931GA  SA308166096GA  SA312479909GA  SA307251718GA  SA310517738GA.DE  SA309334823GA  SA308368146GA  SA307291936GA  SA311428151GA  SA308501356GA  SA307231437GA  SA312049315GA  SA311153112GA  SA308501384GA  SA307231435GA  SA310529220GA  SA311195012GA  SA310199606GA  SA307413105GA  SA311194906GA  SA311121899GA  SA308481055GA  SA307443221GA  SA308186405GA  SA310506700GA  SA312080645GA  SA308368210GA  SA308145726GA  SA311205480GA  SA309436334GA  SA307302043GA  SA309426084GA  SA308501273GA  SA311297256GA  SA308347814GA  SA308105082GA  SA312080882GA  SA311317703GA  SA311205614GA  SA308266885GA  SA309203421GA  SA313105700GA  SA312091030GA  SA308166013GA  SA312122082GA  SA308419294GA  SA308266825GA  SA308317145GA  SA308166024GA  SA308196575GA  SA310382743GA  SA312263852GA  SA308317107GA  SA308337689GA  SA310058740GA  SA309436271GA  SA308266877GA  SA307423138GA  SA307453447GA  SA307514352GA  SA308165991GA  SA309021981GA  SA309132829GA  SA307392758GA  SA308317160GA  SA308521783GA  SA308094869GA  SA308196545GA  SA311163903GA  SA312522384GA  SA311236338GA  SA312049318GA  SA311246670GA  SA308327469GA  SA307342329GA  SA312091031GA  SA310290739GA  SA307332216GA  SA307312093GA  SA308327353GA  SA307443204GA  SA309253743GA  SA308327367GA  SA312294776GA  SA311132549GA  SA309142940GA  SA307413116GA  SA308176277GA  SA307453543GA  SA312049436GA  SA311256856GA  SA308166081GA  SA312039105GA  SA312080555GA  SA308155979GA  SA308459837GA  SA312018898GA  SA312263849GA  SA312039228GA  SA309446444GA  SA312152404GA  SA307180707GA  SA307332221GA  SA308155969GA  SA312029055GA  TF0307  SA309011888GA  SA308166009GA  SA308145687GA  SA311184494GA  SA307312096GA  SA312059617GA  SA311153176GA  SA311236319GA  SH611340025SA  SA308368244GA  SA311142972GA  SA311163992GA  SA309243674GA  SA307322126GA  SA312059613GA  SA309253834GA  SA307392833GA  SA311153225GA  SA312080711GA  SA308166118GA  SA310351472GA  SA309436310GA  SA309032002GA  SA309365214GA  SA312049396GA  SA309112557GA  SA308388604GA  SA310048676GA  SA308317118GA  SA312069734GA  SA312458168GA  SA311184419GA  SA310351475GA  SA311174221GA  SA311184759GA  SA309365295GA  SA308491105GA  SA309517535GA  SA311195017GA  SA311122016GA  SA311225893GA  TF0269  SA311195280GA  SA307332241GA  SA308459803GA  SA312080884GA  SA310270582GA  SA307413117GA  SA308409142GA  SA309476868GA  SA308166103GA  SA312039231GA  SA308398749GA  SA310270584GA  SA312049394GA  SA310506675GA  SA308125327GA  SA308368155GA  SA312080753GA  SA308166001GA  SA310058724GA  SA309354887GA  SA311184694GA  SA308459732GA  SA311488454GA  SA309446387GA  SA310300788GA  SA311142913GA  SA308317121GA  SA308521792GA  SA308491126GA  SA309243671GA  SA310507497GA  SA309365301GA  SA312091044GA  SA307231259GA  SA312386396GA  SA309042189GA  SA307443305GA  SA312029025GA  SA312049307GA  SA308501363GA  SA309365291GA  SA311508573GA  SA307392839GA  SA312080683GA  SA308186390GA  SA311142942GA  SA311184366GA  SA313084681GA  SA310361592GA  SA308317071GA  SA311287054GA  SA308317276GA  SA310310868GA  SA307231378GA  SA312091033GA  SA312091239GA  SA311428075GA  SA309233563GA  SA311287167GA  SA310260308GR  SA309092482GA  SA308501302GA  SA311122167GA  SA313229635GA  SA307413106GR  SA308155863GA  SA311174084GA  SA308317078GA  SA308155959GA  SA313033716GA  SA308286907GA  SA311266968GR  SA310518258GA  SA311142963GA  SA311142901GA  SA309405860GA  SA309253765GA  SA308398897GA  SA308166183GA  SA309294085GA  SA312111563GA  SA309011896GA  SA309253899GA  SA307483775GA  SA308166137GA  SA307302064GA  SA310139266GA  SA308489976GA  SA307392838GA  SA310506754GA  SA310310867GA  SA311205410GA  SA308135533GA  SA308166170GA  SA307332246GA  SA311337977GA  SA307030412GR  SA311468394GA  SA313044066GA  SA308368148GA  SA309092518GA  SA312263850GA  SA312263858GA  SA308449620GA  SA308166022GA  SA308388608GA  SA312386372GA  SA307332215GA  SA310300766GA  SA307443340GA  SA309021931GA  SA309426156GA  SA309334582GA  SA312091266GA.DEL  SA310393166GA  SA307514332GA  SA309021984GA  SA309032066GA  SA311153358GA  SA308357889GA  SA312111581GA  SA308266851GA  SA307453479GA  SA312080878GA  SA310169481GA  SA307080523GA  SA308378295GA  SA308459768GA  SA311215730GA  SA307382624GA  SA308115291GA  SA311256831GA  SA311029364GA  SA308459846GR  SA312192675GA  SA313208789GA  SA310068806GA  SA310382504GA.  SA307403136GA  SA307413122GA  SA311153240GA  SA307332250GA  SA311163576GA  SA312121740GA  SA307504159GA  SA312091034GA  SA307220900GR  SA312253153GA  SA307332227GA  SA311194990GA  SA309365270GA  SA30751",
      "more_code_info": "4342GA  SA307332209GA  SA309142942GA  SA309142939GA  SA312263836GA  SA309365351GA  SA312091221GA.DEL  SA312121753GA  SA312049259GA  SA309334681GA  SA308511542GA  SA310382627GA  SA312469392GA  SA307362420GA  SA312080886GA  SA311142982GA  SA311153192GA  SA311132498GA  SA307504199GA  SA310506731GR  SA312111554GA  SA312091204GA  SA308388486GA  SA311236326GA  SA308501267GA  SA309426070GA  SA310199605GA  SA312080686GA  SA312070085GA  SA312091048GA  SA309365179GA  SA309011876GA  SA312069663GA  SA309183244GA  SA310507491GA  SA307362393GA  SA312069693GA.DEL  SA311142764GA  SA309354937GA  SA312101428GA  SA312152326GA  SA309173175GA  SA312152436GA  SA308266826GA  SA307392989GR  SA311287056GA  SA312232845GA  SA308317239GA  SA311236292GA  SA312152536GA  SA312091242GA  SA311215834GA  SA308459712GA  SA308317075GA  SA308166100GA  SA310475501GA.DEL  SA311428092GA  SA307241591GA  SA310159437GA  SA308489964GA  SA311327906GA  SA311142954GA  SA309253908GA  SA309142904GA  SA313105888GA  SA309365065GA  SA311256834GA  SA308166082GA  SA312121863GA  SA311132651GA  SA308378298GA  SA307332229GA  SA311317734GA  SA312091113GA  SA308094860GA  SA311297400GA  SA310300763GA  SA308368132GA  SA310517568GA  SA310392914GA  SA311205299GA  SA310393028GA  SA307241674GR  SA308084710GA  SA308368149GA  SA308459754GA  SA309528011GA  SA311205301GA  SA308459766GA  SA308317278GA  SA308317231GA  SA307392761GA  SA308368145GA  SA309142911GA  SA312091319GA  SA308368157GA  SA311256847GA  SA311287045GA  SA311287077GA  SA311287012GA  SA310485828GA  SA310413544GA  SA310361707GA  SA310475649GA  SA310423958GA  SA310372294GA  SA310475730GA  SA310485859GA  SA310485824GA  SA310485819GA  SA310475663GA  SA310475709GA  SA310475719GA  SA310485843GA  SA310485853GA  SA312080732GA  SA310393088GA  SA308317266GA  SA308317182GA  SA309365343GA  SA311195016GA  SA309142875GA  SA309365359GA  SA307261817GA  SA309365331GA  SA312069707GA  SA309092538GA  SA311163606GA  SA311163446GA  SA312080671GA  SA309456685GA  SA307261827GA  SA307231501GA  SA308409099GA  SA311122159GA  SA309092510GA  SA311195113GA  SA311184355GA  SA310393196GA  SA309011861GA  SA310518710GA  SA308388438GA  SA307231316GA  SA309142873GA  SA308317260GA  SA308419297GA  SA309365324GA  SA310361809GA  SA308074559GR  SA309092536GA  SA309375375GA  SA311256806GA  SA312522575GA  SA308489952GA  SA311122248GA  SA311205428GA  SA311205570GA  SA309223497GA  SA311142794GA  SA311184476GA  SA311121854GA  SA311121894GA  SA311163597GA  SA311468412GA  SA312091341GA  SA311132326GA  SA311163531GA  SA311205595GA  SA311194962GA  SA311142753GA  SA311184799GA  SA311132598GA  SA311174134GA  SA312479842GA  SA311205596GA  SA311205328GA  SA311122052GA  SA311195172GA  SA307524479GA  SA311132398GA  SA308378297GA  SA311184753GA  SA311195049GA  SA311195170GA  SA311122229GA  SA311121850GA  SA311184471GA  SA311184501GA  SA311163805GA  SA311184816GA  SA311184473GA  SA311488467GA  SA311174045GA  SA312101394GA  SA309032013GA  SA311184511GA  SA311122275GA  SA311184719GA  SA311194987GA  SA311184792GA  SA311184464GA  SA312101367GA  SA312101391GA  SA311205419GA  SA311163854GA  SA311163821GA  SA311163699GA  SA311153307GA  SA311174058GA  SA310423834GA  SA311317650GA.DEL  SA312101384GA  SA311142771GA  SA311122031GA  SA311153033GA  SA311184470GA  SA311142778GA  SA312080786GA  SA311184763GA  SA312080887GA  SA311184735GA  SA307231407GA  SA311184791GA  SA311184869GA  SA311132322GA  SA311256736GA  SA311122230GA  SA312049424GA  SA311184788GA  SA310372100GA  SA311195043GA  SA311205339GA  SA309223473GA  SA311246532GA  SA311132314GA  SA311184729GA  SA308105164GA  SA309203399GA  SA311184457GA  SA311184490GA  SA310496649GA  SA311246508GA  SA311163550GA  SA311184488GA  SA310393297GA  SA308115240GA  SA311121870GA  SA311132313GA  SA311163566GA  SA311205620GA  SA311122047GA  SA311195171GA  SA311132593GA  SA311195007GA  SA311184770GA  SA311184486GA  SA311205413GA  SA311163792GA  SA311195051GA  SA311142777GA  SA311184510GA  SA310496159GA  SA311184512GA  SA311205317GA  SA311142750GA  SA310496326GA  SA311184825GA  SA311488466GA  SA311122270GA  SA311195052GA  SA311205421GA  SA311205433GA  SA311184739GA  SA311132556GA  SA308398712GA  SA310517718GA  SA311205594GA  SA311205586GA  SA312101387GA  SA311205403GA  SA311142774GA  SA311122018GA  SA311153258GA  SA311132546GA  SA311122028GA  SA311174047GA  SA309203413GA  SA311184508GA  SA310517748GA  SA312049357GA  SA311184379GA  SA311132588GA  SA311142755GA  SA311163833GA  SA311122251GA  SA311184732GA.DEL  SA308155976GA  SA312101393GA  SA312101392GA  SA311184482GA  SA311256824GA  SA311184521GA  SA311184721GA  SA311205612GA  SA312101373GA  SA311184764GA  SA311163535GA  SA310496143GA  SA311184737GA  SA311246536GA  SA311153077GA  SA311307452GA  SA311163588GA  SA310517816GA  SA311184725GA  SA311225897GA  SA311142761GA  SA311184769GA  SA311195048GA  SA308095053GA  SA311122046GA  SA310496519GA  SA311184798GA  SA311205600GA  SA311163809GA  SA311184730GA  SA311163568GA  SA311122054GA  SA312079949GA  SA311121737GA  SA312253363GA  SA311184690GA  SA312345029GA  SA311246557GA  SA312101388GA.DEL  SA307514264GA  SA309153034GA  SA311205313GA  SA311195047GA  SA311184776GA  SA311195042GA  SA312101368GA  SA311163438GA  SA311205598GA  SA311205603GA  SA309142918GA"
    }
  ]
}