  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mounds View",
      "state": "MN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81945",
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "address_1": "8200 Coral Sea St Ne",
      "address_2": "N/A",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Wordlwide",
      "recall_number": "Z-0806-2019",
      "product_description": "Medtronic implantable pulse generator:  ADAPTA, Dual chamber pacemaker (DDD):   (a) Model Number ADD01  (b) Model Number ADDR01  (c) Model Number ADDR03  (d) Model Number ADDR06  (e) Model Number ADDRL1  (f) Model Number ADDRS1  (g) Model Number ADVDD01",
      "product_quantity": "41802 units",
      "reason_for_recall": "A  subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality",
      "recall_initiation_date": "20190117",
      "center_classification_date": "20190214",
      "report_date": "20190220",
      "code_info": "ADAPTA, Dual chamber pacemaker (DDD):   (a) Model Number ADD01, GTIN:  00643169708754, All Serial Numbers    (b) Model Number ADDR01, GTIN:  00643169708587, 00643169708594, 00643169708600, 0643169708624, 00643169969711, All Serial Numbers     (c) Model Number ADDR03, GTIN:  00643169708631, 00643169708648, All Serial Numbers    (d) Model Number ADDR06, GTIN:  00643169708655, 00643169708662, All Serial Numbers    (e) Model Number ADDRL1, GTIN:  00643169518247, 00643169708679, 00643169708686, 00643169708709, 00643169969803, All Serial Numbers    (f) Model Number ADDRS1, GTIN: 00643169708716, 00643169708723, 00643169708730, 00643169708747, 00643169969810, All Serial Numbers     (g) Model Number ADVDD01, GTIN:  00643169708761, 00643169708778, 00643169969735, All Serial Numbers"
    }
  ]
}