{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Center Valley",
      "address_1": "3773 Corporate Pkwy",
      "reason_for_recall": "The deflector may not move freely on the ventricular catheter.  The deflector comes pre-positioned at the 5 cm mark on the catheter.  In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle.  There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter",
      "address_2": "",
      "product_quantity": "11",
      "code_info": "Lots 4503489878, 4503792749, 4503682927, 4503508385, 4503789879, 4503800248 and 4503682935.",
      "center_classification_date": "20130212",
      "distribution_pattern": "Nationwide Distribution including MI, PA and TX.",
      "state": "PA",
      "product_description": "Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric Cathet, w/Stylet and RT-ANG Guide    Central Nervous System Shunt and Components intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.",
      "report_date": "20130220",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Aesculap, Inc.",
      "recall_number": "Z-0805-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63244",
      "termination_date": "20140108",
      "more_code_info": "",
      "recall_initiation_date": "20120604",
      "postal_code": "18034-8217",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}