{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rhondda Cynon Taff",
      "state": "",
      "country": "United Kingdom",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84367",
      "recalling_firm": "Flexicare Medical Ltd.",
      "address_1": "Cynon Valley Business Park, Mountain Ash",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US: Nationwide  OUS: Australia, Canada, China,  Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK",
      "recall_number": "Z-0804-2020",
      "product_description": "BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Sterile EO, Qty 10, UDI: (01)15055788724053(17)230901(10)181002320",
      "product_quantity": "1,024,890 devices",
      "reason_for_recall": "The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device.  If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.",
      "recall_initiation_date": "20191127",
      "center_classification_date": "20200113",
      "termination_date": "20210427",
      "report_date": "20200122",
      "code_info": "Part Number/Lot Number:  040-342U 180802596 I OGH  040-342U 181002320 I OGH  040-342U 181100085 I OGH  040-342U 181200025 I OGH",
      "more_code_info": ""
    }
  ]
}