  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wayne",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73137",
      "recalling_firm": "Maquet Cardiovascular Us Sales, Llc",
      "address_1": "45 Barbour Pond Drive",
      "address_2": "N/A",
      "postal_code": "07470",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide in the states of CA, NY, and OR",
      "recall_number": "Z-0804-2016",
      "product_description": "Flow-i Anesthesia System    Product Usage:  The indication for the Flow-i Anesthesia System is administering inhalation anesthesia while controlling the entire ventilation of patients with no ability to breath, as well as in supporting patients with a limited ability to breath. The system is intended for use on neonatal to adult patient populations. The system is intended for use in hospital environments, except MRI environment by healthcare professionals trained in inhalation anesthesia administration.",
      "product_quantity": "7 units within US",
      "reason_for_recall": "Electrical issues in the control, monitoring or panel sub-systems, caused by the printed circuit boards (PCBs) that control the sub-systems. The problems manifested as multiple re-starts of the sub-systems and, in a few instances, disabling of the system has occurred. Replacement PCBs are now available to correct the issues.    Only the Model C20 is affected in the US.",
      "recall_initiation_date": "20151210",
      "center_classification_date": "20160222",
      "termination_date": "20160706",
      "report_date": "20160302",
      "code_info": "Model numbers: C20, C30 and C40 (Only Model # C20 for US)    Part numbers:  66 77 200 (US Only), 66 77 300 and 66 77 400    Serial numbers (2743, 2852, 2854, 2855, 2856, 2858",
      "more_code_info": ""
    }
  ]
}