  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86951",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide",
      "recall_number": "Z-0803-2021",
      "product_description": "CoaguChek XS Prof (Professional) Meters- IVD professional healthcare providers for quantitative prothrombin time testing for the monitoring of warfarin therapy.   Catalog: 04837975001    Units could consist of any of the following strip configurations: 6 count, 24 count, 2x 24 count (i.e. 48 strips)*    *CoaguChek XS 3x48 promo LTC HH  Catalog: 06298176001  CoaguChek XS 6x24 promo LTC HH  Catalog:08468699001  CoaguChek XS 4x48 promo                Catalog: 08468745001  CoaguChek XS 6x48 promo                Catalog: 08468753001",
      "product_quantity": "212,093 units",
      "reason_for_recall": "Possible Misinterpretation of Displayed Measuring Units from INR to %Quick or Seconds; providing  instructions to users for ensuring the meter is displaying INR results",
      "recall_initiation_date": "20201202",
      "center_classification_date": "20210107",
      "termination_date": "20220630",
      "report_date": "20210113",
      "code_info": "All serial numbers",
      "more_code_info": ""
    }
  ]
}