  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Allendale",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69784",
      "recalling_firm": "Stryker Spine",
      "address_1": "2 Pearl Ct",
      "address_2": "N/A",
      "postal_code": "07401-1611",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed to GA, IN,  MI, NJ, OH, PA, TX & VA.",
      "recall_number": "Z-0803-2015",
      "product_description": "Specialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile.    Used to final tighten to 12 Nm the blocker of the XIA 3, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors.",
      "product_quantity": "15 units",
      "reason_for_recall": "Stryker has received a complaint from customers relating to a fracture during the use of the Specialty Audible Torque Wrench in surgery.",
      "recall_initiation_date": "20141006",
      "center_classification_date": "20141215",
      "termination_date": "20170508",
      "report_date": "20141224",
      "code_info": "Lot number 4279"
    }
  ]
}