  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73056",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada",
      "recall_number": "Z-0802-2016",
      "product_description": "Radial Artery Pressure Monitoring Catheter Sets and Trays    The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix \u0013BH.",
      "product_quantity": "12691",
      "reason_for_recall": "A manufacturing process may lead to catheter tip fracture and or separation.  Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.",
      "recall_initiation_date": "20160106",
      "center_classification_date": "20160224",
      "termination_date": "20170823",
      "report_date": "20160302",
      "code_info": "RPN/Catalog number:  C-PMS-250-RA  C-PMS-251-RA  C-PMS-300-RA  C-PMS-300-RA-CHOP  C-PMS-301J-RA  C-PMS-301-RA  C-PMS-400-RA  C-PMSY-250-RA  C-PMSY-300-RA  C-PMSY-301-RA    Lots:  6049998  6160195  6160212  NS5898558  NS6063328  NS6178204  NS6178206  5895605  6191185  5793196  5800313  5800443  5800569  5806981  5810905  5810917  5824484  5824546  5824547  5824548  5866984  5866985  5866986  5870319  5881447  5881519  5889496  5891905  5911486  5922847  5922848  5925742  5928632  5939446  5939447  5939448  5976502  6025016  6025094  6033768  6034981  6034983  6050000  6062738  6062742  6067675  6067676  6075206  6075207  6083936  6083954  6083985  6083986  6129056  6148790  6160236  6165591  6174974  6175289  6191362  6269150  NS6063331  NS6108226  NS6170739  NS6170767  5905252  6003109  NS6072553  5793197  5800571  5800572  5824485  5824551  5845063  5864072  5866988  5873125  5881151  5881152  5884641  5884642  5889464  5892063  5892064  5922850  5939431  5942091  5951549  5953768  5954322  5967404  5974281  5994817  5996926  6003161  6003162  6017048  6025095  6025145  6034984  6042699  6050002  6075210  6075211  6075212  6081723  6083988  6100756  6124809  6125111  6148791  6153122  6160239  6160240  6168131  6171368  6187477  6250532  6059759  NS6047396  NS6047397  NS6047398  NS6047399  NS6047440  NS6047442  6124812  5793192  5824526  5884647  5892068  5942090  5967410  6003167  6059765  6134128  6153124  6168137  6254068  5793193  5866995  5892069  5942092  6003168  6025099  6067664  6168138",
      "more_code_info": ""
    }
  ]
}