  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Innsbruck",
      "state": "",
      "country": "Austria",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69930",
      "recalling_firm": "Leonhard Lang Gmbh",
      "address_1": "Archenweg 56",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of OH and TX.",
      "recall_number": "Z-0802-2015",
      "product_description": "GE Healthcare Pediatric Direct-Wired Disposable Defibrillation/Pacing Electrodes.  REF 2059144-001 and 2059144-005.  .    Product Usage:  Usage:  Skintact Pediatric Multifunction Electrodes are for use on pediatric patients, less than 8 years of age, or weighing less than 25 kg (55 lbs), for external defibrillation, pacing, monitoring and cardioversion. The device is non-sterile and for single use only.",
      "product_quantity": "1080 pieces",
      "reason_for_recall": "The firm discovered that during the use of these defibrillation electrodes a possibility for arcing and a resulting malfunction exists.  This could lead to a situation, in which a patient who is in a life threatening situation requiring a defibrillation can receive such therapy only with a delay or not at all.",
      "recall_initiation_date": "20141203",
      "center_classification_date": "20141219",
      "termination_date": "20150813",
      "report_date": "20141231",
      "code_info": "Lots:  30618-0771, 30702-0778, 40217-0776, 40407-0777, 40424-0772, 40704-0775",
      "more_code_info": ""
    }
  ]
}