      "country": "United States",
      "city": "Ann Arbor",
      "address_1": "6200 Jackson Road",
      "reason_for_recall": "Inaccurate readings.  Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.",
      "address_2": "",
      "product_quantity": "622 units",
      "code_info": "Catalog number: 101 and Serial Numbers: 0503, 1000, 1017, 1070, 1071, 1073, 1074, 1076, 1077, 1080, 1082, 1083, 1107-1110, 1115-1117,1124,1126-1134, 1136, 1137, 1139-1146, 1148-1152, 1154-1156, 1159-1162, 1164-1166, 1168-1177, 1182, 1185-1190, 1192,1193, 1195, 1197, 1198, 1225, 1228, 1229, 1237, 1239-1241,1243-1245,1247, 1249-1326, 1332-1339, 1344-1347,  1349, 1350-1363, 1365-1409, 1411-1436,1438, 1440-1447, 1449-1453, 1455-1463,1465-1482,1484-1492,1494-1552,1555-1565, 1567-1600,1602,1605-1607,1609-1611,1613-1623,1625-1629,1632-1651,1653, 1655-1660, 1662-1664, 1666, 1667,1669-1679,1681-1684,1686,1689-1691, 1693-1710,1712-1715,1717-1750,1752-1757,1759-1771,1773-1785,1787,1788,1790-1809, 1811-1831,1833,1838,1839,1855-1858,1866, and TK001-TK011.",
      "center_classification_date": "20130212",
      "distribution_pattern": "Worldwide Distribution.",
      "state": "MI",
      "product_description": "Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.",
      "report_date": "20130220",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "recall_number": "Z-0802-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "60695",
      "termination_date": "20131108",
      "more_code_info": "",
      "recall_initiation_date": "20101101",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}