  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86972",
      "recalling_firm": "Ortho-Clinical Diagnostics, Inc.",
      "address_1": "100 Indigo Creek Dr",
      "address_2": "",
      "postal_code": "14626-5101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.",
      "recall_number": "Z-0801-2021",
      "product_description": "VITROS Automation Solutions, VAS, Catalog Numbers 6844300 & 6844301 - Product Usage: intended to automate pre-analytical and post-analytical sample processing in the clinical laboratory.",
      "product_quantity": "40 units",
      "reason_for_recall": "A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.",
      "recall_initiation_date": "20201120",
      "center_classification_date": "20210107",
      "termination_date": "20231024",
      "report_date": "20210113",
      "code_info": "enGen Laboratory Automation Systems and VITROS Automation Solutions Using All TCAutomationTM configured with Thermo Scientific Aliquoter & Labeller Module and using    Software Versions 4.6 and below",
      "more_code_info": ""
    }
  ]
}