      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73056",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada",
      "recall_number": "Z-0801-2016",
      "product_description": "Femoral Artery Pressure Monitoring Catheter Sets and Trays    The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.",
      "product_quantity": "1515",
      "reason_for_recall": "A manufacturing process may lead to catheter tip fracture and or separation.  Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.",
      "recall_initiation_date": "20160106",
      "center_classification_date": "20160224",
      "termination_date": "20170823",
      "report_date": "20160302",
      "code_info": "RPN/Catalog number:  C-NPMS-501J-15  C-NPMSY-501J-15  C-PMS-300-FA  C-PMS-300J-FA  C-PMS-301-FA  C-PMS-301J-FA  C-PMS-400-FA  C-PMS-401-FA  C-PMS-401J-FA  C-PMSY-300-FA  C-PMSY-300J-FA  C-PMSY-400J-FA  C-PMSY-401-FA    Lots:  6054830  6054831  6072436  6078670  6081720  6083977  NS6054841  NS6054842  NS6075552  NS6082372  NS6082373  NS5833881  NS5881267  NS5881268  NS5892106  NS5892107  NS5892108  NS5914766  NS5922706  NS5928596  NS5932820  NS5932821  NS5967447  NS5967448  NS5967449  NS6014474  NS6014475  NS6014484  NS6014485  NS6042778  NS6042795  NS6042796  NS6050046  NS6063330  NS6063354  NS6075555  NS6100599  NS6100600  NS6108225  NS6124985  NS5888542  NS5928597  NS6028553  NS6042797  NS6090273  NS6090274  NS6100601  NS6130149  5800570  5860387  5891906  5980254  6003160  6062725  6143835  NS5830771  NS5866956  NS5866957  NS5881269  NS6014460  NS6042798  NS6050047  NS6075534  NS6075556  5895598  6129058  6148613  NS5932822  NS5932837  NS5994907  NS6130498  6067782  6139280  NS6042805  NS6130500  6049951  6137353  6139281  6034986  6049957  6067784  NS5833887  NS5866959  NS5881272  NS5881334  NS5881335  NS5919925  NS6047449  NS6054847  NS6054848  NS6072526  NS6072555  NS6082379  NS6090277  NS6105921  NS6124989  NS6130501  NS6148639  NS5866961  NS5866962  NS5892115  NS5922709  NS5932844  NS5974256  NS6031988  NS6031990  NS6042578  NS6050052  NS6050053  NS6031992  NS6031993  NS6130504  NS6137424  NS6054849",
      "more_code_info": ""
    }
  ]
}