  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73056",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada",
      "recall_number": "Z-0800-2016",
      "product_description": "Single Lumen Pressure Monitoring Sets and Trays    The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.",
      "product_quantity": "2280",
      "reason_for_recall": "A manufacturing process may lead to catheter tip fracture and or separation.  Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.",
      "recall_initiation_date": "20160106",
      "center_classification_date": "20160224",
      "termination_date": "20170823",
      "report_date": "20160302",
      "code_info": "RPN/Catalog number:  C-PMS-250  C-PMS-2502-15-3.5  C-PMS-2502-15-5.0  C-PMS-250-MONTES-A-092084  C-PMS-251  C-PMS-251J  C-PMS-300  C-PMS-300-CHILDRENS-A-032988  C-PMS-300J  C-PMS-301  C-PMS-301J  C-PMS-301J-PKG  C-PMS-400J    Lots:  6028502  6034979  6083984  6194848  NS6014456  NS6014457  6148788  6254047  5906417  6160213  NS6064797  NS6014483  NS6175149  NS6184593  NS6187516  NS6014458  5889463  5898283  5922849  5941909  5964565  5976506  6021072  6028497  6143834  6160238  NS5933957  5828147  5941919  NS5932835  NS5932836  NS6050233  NS6072523  NS6108227  NS6130497  5803601  6034985  6075213  NS5833882  NS5892109  NS5892110  NS5919920  NS6021136  NS6054843  NS6072524  NS6075557  NS6082375  NS6108228  NS6148658  6067668",
      "more_code_info": ""
    }
  ]
}