  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Houston",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84350",
      "recalling_firm": "Trilliant Surgical, LLC",
      "address_1": "727 N Shepherd Dr Ste 100",
      "address_2": "",
      "postal_code": "77007-1320",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CO, CT, IA, TX and WI.    OUS: None",
      "recall_number": "Z-0799-2020",
      "product_description": "Gridlock Ankle Screw Driver Bit REF 320-35-003 - Product Usage: A screwdriver facilitates implant insertion and removal.",
      "product_quantity": "20 units",
      "reason_for_recall": "A screwdriver used to facilitate implant insertion and removal was found to be out of specification for the minor diameter of the drive feature causing the driver to be unable to retain screws as intended.",
      "recall_initiation_date": "20190314",
      "center_classification_date": "20200110",
      "termination_date": "20220316",
      "report_date": "20200122",
      "code_info": "Part Number: 320-35-003  Lot Number: TSL006617  UDI: (01)00812926026893(10)TSL006617",
      "more_code_info": ""
    }
  ]
}