  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73056",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including DC and PR including the countries of Australia, Germany, China, Chile, Dominican Republic, Panama, Israel, Brazil, Canada",
      "recall_number": "Z-0799-2016",
      "product_description": "Single Lumen Central Venous Catheter Sets and Trays    The subject sets and trays are comprised of a single-lumen uncoated or heparin-coated catheter. Product catalog numbers with heparin coating are designated by the suffix BH.",
      "product_quantity": "1341",
      "reason_for_recall": "A manufacturing process may lead to catheter tip fracture and or separation.  Potential adverse events may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. These catheters are used in venous or arterial pressure monitoring, blood sampling, and administration of drugs and fluids.",
      "recall_initiation_date": "20160106",
      "center_classification_date": "20160224",
      "termination_date": "20170823",
      "report_date": "20160302",
      "code_info": "RPN/Catalog number:  C-PMS-250-PED  C-PMS-251J-PED  C-PMS-300-CHILDRENS-032285  C-PMS-301J-PED  C-PMS-301J-PED-BH  C-PMS-301-PED  C-PMSY-250  C-PMSY-251  C-PMSY-251J  C-PMSY-300  C-PMSY-300J  C-PMSY-300J-UT-A  C-PMSY-300-PED  C-PMSY-301J  C-PMSY-301J-BH  C-PMSY-301J-PED  C-PMSY-301J-PED-OTP  C-PMSY-400-HURLEY-022388  C-PUM-300J  C-PUM-301  C-PUM-301J  C-PUMY-301J    Lots:  NS6168054  NS6178118  NS6063352  NS5855441  5824486  5898058  5967405  6079119  NS5932737  NS6028554  NS6042799  NS5922707  NS5932811  NS6145002  5987014  NS6178099  NS6028540  NS5830758  NS5866960  NS5918174  NS5974263  NS6042577  NS6047450  NS6090278  NS6105922  NS6124990  NS5881337  NS5892116  NS6031991  5782006  5855241  NS5881336  NS5892113  NS5945282  6062729  NS5970950  NS6108255  NS6153173  5793198  5925744  6010406  6148800  NS5796483  NS5928602  NS6047451  NS5970942  NS5954223  5945120  NS5954224  5973354",
      "more_code_info": ""
    }
  ]
}