  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "800 MacArthur Blvd.",
      "reason_for_recall": "Mindray has identified a software anomaly that may where, in the presence of an asystole, the Telepack ECG waveform may produce a unique step pattern (instead of a flat line) if Lead 1 or V is being utilized for detecting cardiac arrhythmia.",
      "address_2": "",
      "product_quantity": "51 units",
      "code_info": "Part No Serial No  300BF-PA00002 EH-17000037  300BF-PA00002 EH-17000032  300BF-PA00002 EH-17000033  300BF-PA00002 EH-25000099  300BF-PA00002 EH-18000043  300BF-PA00003 EH-1A000072  300BF-PA00002 EH-22000088  300BF-PA00003 EH-1A000073  300BF-PA00002 EH-12000060  300BF-PA00002 EH-17000038  300BF-PA00003 EH-21000077  300BF-PA00002 EH-18000058  300BF-PA00002 EH-23000096  300BF-PA00003 EH-19000059  300BF-PA00003 EH-19000060  300BF-PA00002 EH-18000051  300BF-PA00002 EH-12000054  300BF-PA00002 EH-16000029  300BF-PA00002 EH-11000052  300BF-PA00002 EH-17000039  300BF-PA00003 EH-1A000064  300BF-PA00002 EH-18000050  300BF-PA00003 EH-13000078  300BF-PA00003 EH-13000085  300BF-PA00002 EH-18000053  300BF-PA00002 EH-17000036  300BF-PA00002 EH-17000031  300BF-PA00003 EH-17000040  300BF-PA00002 EH-18000054  300BF-PA00002 EH-1A000069  300BF-PA00002 EH-12000062  300BF-PA00002 EH-21000076  115-006908-04 Software Kit  300BF-PA00003 EH-13000083  300BF-PA00003 EH-13000084  300BF-PA00003 EH-22000087  300BF-PA00002 EH-18000056  300BF-PA00002 EH-17000035  300BF-PA00002 EH-18000048  300BF-PA00002 EH-18000052  300BF-PA00003 EH-21000081  300BF-PA00002 EH-17000030  115-006908-03 Software Kit  115-006908-03 Software Kit  115-006908-03 Software Kit  300BF-PA00003 EH-19000062  300BF-PA00002 EH-18000057  300BF-PA00003 EH-22000083  300BF-PA00002 EH-11000051  300BF-PA00002 EH-18000041  300BF-PA00002 EH-21000082.",
      "center_classification_date": "20130211",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "NJ",
      "product_description": "Mindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, and 300BF-PA00003 (DPM Central Monitoring System) 115-006908-01, -03, -04 (DPM software upgrade kit). The DPM Central Monitoring System transfers information between DPM Central station and other networked devices.",
      "report_date": "20130220",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Mindray DS USA, Inc. d.b.a. Mindray North America",
      "recall_number": "Z-0798-2013",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64114",
      "termination_date": "20130815",
      "more_code_info": "",
      "recall_initiation_date": "20121128",
      "postal_code": "07430",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}