  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lewisville",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73069",
      "recalling_firm": "Orthofix, Inc",
      "address_1": "3451 Plano Pkwy",
      "address_2": "N/A",
      "postal_code": "75056-9453",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US, Spain, Italy, and Australia.",
      "recall_number": "Z-0797-2016",
      "product_description": "Firebird Spinal Fixation System Torque Limiting Handle (PN 52-1512)    Intended for posterior, non-cervical pedicle, and non-pedicle fixation (T1-S2/llium).",
      "product_quantity": "36 units",
      "reason_for_recall": "There is a possibility that the device may deliver a torque value less than the required setting after initial sterilization.  In worst case, the unexpected drop in torque value may lead to the spinal construct loosening and requiring corrective surgical intervention.",
      "recall_initiation_date": "20160108",
      "center_classification_date": "20160210",
      "termination_date": "20161116",
      "report_date": "20160217",
      "code_info": "BS535418105,  BS535419118,  BS535420030,   BS535419124,  BS562325021,  BS535418158,  BS535418038,  BS535418138,  BS535418012,  BS535419006,  BS535420039,   BS535418127,   BS549024010,  BS535419141,  BS535418081,  BS535418111,  BS535418117,  BS535420041,  BS535418144,  BS535420081,  BS535419151,  BS562325006,  BS546825008,  BS535418002,  BS535418013,  BS535418131,  BS535420006,  BS535420010,  BS535420026,  BS535420047,  BS556219012,  BS556219026,  BS535418109,  BS535420082,  BS557717009, and   BS562325023.",
      "more_code_info": ""
    }
  ]
}