  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84511",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.",
      "recall_number": "Z-0794-2020",
      "product_description": "ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids.",
      "product_quantity": "43,495,400 units",
      "reason_for_recall": "Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.",
      "recall_initiation_date": "20191212",
      "center_classification_date": "20200109",
      "report_date": "20200115",
      "code_info": "PRODUCT CODE UDI  7N8300 85412091839  7N8301 85412475387  7N8301K 85412499635  7N8310 85412476728  7N8330K 85412613499  7N8332K 85412599625  7N8334K 85412613482  7N8370K 85412613505  7N8371 85412091822  7N8375K 85412613475  7N8376K 85412599601  7N8377 85412091808  7N8378 85412091792  7N8390 85412475417  7N8391 85412475400  7N8399 85412091785  7N8399K 85412478227      All Lots"
    }
  ]
}