  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Deerfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84511",
      "recalling_firm": "Baxter Healthcare Corporation",
      "address_1": "1 Baxter Pkwy",
      "address_2": "N/A",
      "postal_code": "60015-4625",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.",
      "recall_number": "Z-0793-2020",
      "product_description": "Clearlink System Continu-Flo Administration Sets, Product Codes:  2C8519, 2C8537, and 2C8541 - Product Usage:For the administration of fluids from a container into the patients vascular system through a vascular access device.",
      "product_quantity": "271,014,984 units",
      "reason_for_recall": "Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.",
      "recall_initiation_date": "20191212",
      "center_classification_date": "20200109",
      "report_date": "20200115",
      "code_info": "Product Codes:  2C8519, UDI 00085412048994; 2C8537, UDI 00085412048970; and 2C8541, UDI 00085412071107  All Lots"
    }
  ]
}