  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mountain View",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64132",
      "recalling_firm": "Hitachi Chemical Diagnostics Inc",
      "address_1": "630 Clyde Ct",
      "address_2": "N/A",
      "postal_code": "94043-2239",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - France only",
      "recall_number": "Z-0793-2013",
      "product_description": "OPTIGEN Food 30; Catalog number: 85035. For allergen testing.      Manufactured by Hitachi Chemical Diagnostics, Mountain View, CA.  Distributed by Theridiag, Cedex, France",
      "product_quantity": "20 kits",
      "reason_for_recall": "There is no evidence that a visual inspection was performed on one tray of coverslips. Possibility of a low fill volume which could lead to a false negative results.",
      "recall_initiation_date": "20121221",
      "center_classification_date": "20130207",
      "termination_date": "20130516",
      "report_date": "20130213",
      "code_info": "Catalog number: 85035.  Lot number 42K283626, expiration May 31, 2013.",
      "more_code_info": ""
    }
  ]
}