  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "San Jose",
      "address_1": "2350 Qume Dr",
      "reason_for_recall": "A customer site was identified to have three FACSLyric IVD systems with FACSuite clinical v1.1.1 installed on a computer workstation with windows 10 operating system (OS). Windows 10 OS is not a released configuration for clinical use. It is released for research use only.",
      "address_2": "",
      "product_quantity": "3 devices",
      "code_info": "UDI: (01)00382906628787(11)180508(21)Z662878000056  serial number: Z662878000056  Work Station SN 2UA80820ZZ    UDI: (01)00382906628787(11)180301(21)Z662878000029  serial number: Z662878000029   Work Station SN 2UA80820ZW    UDI: (01)00382903389605(11)180315(21)11662382000007  serial number: R662382000007  Work Station SN 2UA8082101",
      "center_classification_date": "20190201",
      "distribution_pattern": "US Distribution in MD.",
      "state": "CA",
      "product_description": "BD FACSLyric 3L 10C with FACSuite Clinical Software v1.1.1, catalog number 662878    BD FACSLyric is a high-performance flow cytometer designed to support both routine clinical analysis and research for the identification, quantification, and characterization of cells in support of cell analysis. BD FACSuite software is loaded onto an adjacent PC workstation and is used to operate the instrument, acquire samples, and analyze data. Quality control performance, tracking, and reporting are streamlined and automated. Routine tasks such as daily cleaning and shutdown can be programmed to occur automatically. The BD FACSuite clinical software allows users to run BD IVD assays.",
      "report_date": "20190213",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Becton, Dickinson and Company, BD Biosciences",
      "recall_number": "Z-0792-2019",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "81915",
      "termination_date": "20210507",
      "more_code_info": "",
      "recall_initiation_date": "20190102",
      "postal_code": "95131-1812",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}