      "status": "Terminated",
      "city": "Union City",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72937",
      "recalling_firm": "Mizuho OSI",
      "address_1": "30031 Ahern Ave",
      "address_2": "N/A",
      "postal_code": "94587-1234",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US, South Africa, Japan and United Arab Emirates.",
      "recall_number": "Z-0792-2016",
      "product_description": "Mizuho OSI Trios, Table Base;  Voltage: 100-240V ~ 50/60 Hz  Amps: 5,0-5.0A;  Model 7803 and 7803AP    Orthopaedic:  Operating room tablebase - an electronically powered mobile operating table is created, designed for temporary (<24 hours) support and positioning of a patient in a prone, supine, or lateral position depending on the specific table top utilized.",
      "product_quantity": "33",
      "reason_for_recall": "Table Bases may have been assembled incorrectly, missing a retaining ring from the head end assembly.  Over time, the parts could go out of alignment, making clutch engagement or disengagement difficult.  If the clutch cannot disengage, the user would be unable to rotate the patient from supine to prone.",
      "recall_initiation_date": "20151214",
      "center_classification_date": "20160209",
      "termination_date": "20160210",
      "report_date": "20160217",
      "code_info": "Model numbers 7803, 7803AP.    Serial numbers 228-261.",