  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Palm Harbor",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75785",
      "recalling_firm": "Oscor, Inc.",
      "address_1": "3816 Desoto Blvd",
      "address_2": "N/A",
      "postal_code": "34683-1618",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US including CA, NJ, MD, TX, FL, NC, MO, and Puerto Rico.  Internationally to AUSTRIA, BOLIVIA, CANADA, CHILE, CYPRUS, NETHERLANS, EGYPT, GERMANY, HONG KONG, ISRAEL, ITALY, SWITZERLAND, and VIETNAM",
      "recall_number": "Z-0790-2017",
      "product_description": "OSCOR ADELANTE BREEZEWAY 8F C55¿ S61CM D66CM  Catalog # AB081040  Is designed to facilitate the introduction of catheters to any of the heart chambers, including the left atrium via a transseptal puncture.",
      "product_quantity": "156",
      "reason_for_recall": "Complaint  that during insertion of the dilator through the sheath a fragment of inner liner came out of the sheath prior to use.",
      "recall_initiation_date": "20161005",
      "center_classification_date": "20161219",
      "termination_date": "20190726",
      "report_date": "20161228",
      "code_info": "Lot #'s: C1-09714 Exp. February 11, 2018, OR-04657 Exp. April 20, 2018, C1-10900 Exp. July 19, 2018, C1-10899 Exp. July 19, 2018, C1-11048 Exp. August 3, 2018,  C1-11729 Exp. October 4, 2018, C1-11729 Exp. October 4, 2018,   OR-04729 Exp. October 29, 2018, C1-12233 Exp. December 21, 2018, C1-12234 Exp. December 22, 2018, C1-12234 Exp. December 22, 2018, C1-09062 Exp. June 24, 2017, C1-09062 Exp. June 24, 2017, C1-11047 Exp. August 3, 2018 and C1-12234 Exp. December 22, 2018 2"
    }
  ]
}