  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75742",
      "recalling_firm": "Smiths Medical ASD Inc.",
      "address_1": "6000 Nathan Ln N",
      "address_2": "N/A",
      "postal_code": "55442-1690",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Foreign Distribution in the  Italy, France, Switzerland, Germany, Sweden",
      "recall_number": "Z-0789-2017",
      "product_description": "Acuvance¿2 Safety Catheter, Product No. 1720-AI    Product Usage:  The Acuvance¿ 2 Safety Catheter is a Peripheral Intravenous Catheter (PIVC). PIVC\u0019s are designed for single use venipuncture on all patient populations where access to the veins and arteries are required. The catheter has a Safety Needle Assembly (SAN) designed to reduce risk of accidental needle sticks to the operators. The lateral port is equipped with a non-return silicone valve and a polypropylene snap cap with a winged polypropylene hub for fastening.",
      "product_quantity": "26,750",
      "reason_for_recall": "Smiths Medical became aware that three (3) lots of Acuvance2 Safety Catheter, product number 1720-AI, contain devices with missing or incorrectly positioned side port valves.",
      "recall_initiation_date": "20161121",
      "center_classification_date": "20161218",
      "termination_date": "20170801",
      "report_date": "20161228",
      "code_info": "3120882, 3120883, 3220733"
    }
  ]
}