  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "73076",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US Nationwide, Armenia, Australia, Austria, Belgium, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Netherlands, New Zealand, Norway, P R China, Portugal, Poland, Russia, Singapore, Slovenia, South Africa, Spain, Sweden, South Korea, Switzerland, Thailand, Turkey, UAE, UK, and Brazil.",
      "recall_number": "Z-0787-2016",
      "product_description": "SIGMA HP MBT Non-Keel Punch Knee Instrument.  Designed to be used as an option in stabilizing the tibial trial during trial reduction.",
      "product_quantity": "351",
      "reason_for_recall": "Increased punch height, causing a 0.063 in (1.6 mm) gap between the trial and the instrument. This may cause the surgeon to select an insert that is too thin.  Surgical delay may result if the punch dislodges from the inserter and needs to be removed from the joint space.  Insert Dislocation or Spin Out, or Poor Joint Mechanics may result, if not noted during surgery. May require revision surgery.",
      "recall_initiation_date": "20160108",
      "center_classification_date": "20160209",
      "termination_date": "20170324",
      "report_date": "20160217",
      "code_info": "Cat. No.  GTIN  Lot  Description    950502016  10603295227366  ABB92402  SIGMA HP MBT Cemented Punch size 1-1.5    950502017  10603295227373  ABB84491  SIGMA HP MBT Cemented Punch size SZ 2-3    950502018  10603295227380  ABB92358  SIGMA HP MBT Cemented Punch size SZ 4-7    950502020  10603295227403  ABC33125  SIGMA HP MBT Noncemented Punch size 2-3    950502020  10603295227403  ABB81733  SIGMA HP MBT Noncemented Punch size SZ 2-3    950502020  10603295227403  ABB81732  SIGMA HP MBT Noncemented Punch size SZ 2-3    950502021  10603295227410  ABB39952  SIGMA HP MBT Noncemented Punch size SZ 4-7    950502021  10603295227410  ABC33123  SIGMA HP MBT Noncemented Punch size SZ 4-7",
      "more_code_info": ""
    }
  ]
}