  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Martinez",
      "address_1": "757 Arnold Dr Ste A",
      "reason_for_recall": "Update to correct the possibility for patients or operators to have their fingers trapped or pinched between the table top and the bottom frame of the 550 TxT Treatment Table. Combined with this update instruction is a second complaint issue related to an incident in which a cable inside of the TxT\"  Treatment Table has been damaged due to contacting a moving part.  The resulting short-circuit cau",
      "address_2": "",
      "product_quantity": "305 active devices.",
      "code_info": "550 TxT table - part number 07346534.",
      "center_classification_date": "20140117",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and internationally to: Canada, Italy, P.R. China, Argentina, Pakistan, South Africa, Egypt, Iran, Vietnam, Indonesia, Brazil, Hungary, Bulgaria, India, Spain, Switzerland, Trinidad, Tobago, Italy, Algeria, Croatia, Turkey, Japan, Germany, Portugal, Poland, France, Sri Lanka, Russian Federation, France, Lebanon, Slovakia, Ukraine, Pakistan, and Korean Republic.",
      "state": "CA",
      "product_description": "PRIMUS, ONCOR, ARTISTE Linac systems with component 550 TxT Treatment Table.  TT-A, TT-D, TT-M or TT-S Table tops.    Manufactured by:  SIEMENS AG, MEDICAL SOLUTIONS   Roentgenstrasse 19-21   Kemnath,  GERMANY  95478     The intended use of the SIEMENS branded, ARTISTE\", ONCOR\" and PRIMUS\" family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.",
      "report_date": "20140129",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA,  Inc",
      "recall_number": "Z-0787-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67126",
      "termination_date": "20150130",
      "more_code_info": "",
      "recall_initiation_date": "20131205",
      "postal_code": "94553-3615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}