  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Brooklyn Park",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81702",
      "recalling_firm": "Medtronic Perfusion Systems",
      "address_1": "7611 Northland Dr N",
      "address_2": "N/A",
      "postal_code": "55428-1088",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) and countries of: Australia, Canada, France, Germany, Japan, and Sweden.",
      "recall_number": "Z-0785-2019",
      "product_description": "RESERVOIR \"Y\" ADAPTER,   (a) BT920  (b) BT926    Suction and Anticoagulation assemblies are indicated for use with an autotransfusion system between the suction tip and reservoir.",
      "product_quantity": "1100 units",
      "reason_for_recall": "Medtronic discovered that certain Y-connectors included in the suction lines have exhibited tearing and cracking. These defects could cause the Y-connectors to leak when in use, which could result in potential blood loss and a breach in the circuit.",
      "recall_initiation_date": "20181127",
      "center_classification_date": "20190129",
      "report_date": "20190206",
      "code_info": "BT920:  GTIN/UDI                Lot/Serial Numbers  10673978522460 216044420  10673978522460 216062719  10673978522460 216438180    BT926:  GTIN/UDI                Lot/Serial Numbers  10673978522477 215782209  10673978522477 215903433  10673978522477 216034158  10673978522477 216107790  10673978522477 216372963  10673978522477 216438301"
    }
  ]
}