  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Rd",
      "reason_for_recall": "The recalling firm determined that visual inspections of the fork joints behind the detector that are conducted during six month preventive maintenance (PM) may not detect small cracks or deficiencies in the fork weldment.",
      "address_2": "",
      "product_quantity": "255 units",
      "code_info": "ADAC ARGUS: Model #882060; Serial #: 1023, 1178, 106785, 136560, 1100005, 9051005, 9304030, 9309011, 9501007, 9509005, 9510027, 9510028, 9603042, 9603043, 9604044, 9604047, 9605011, 9607060, 9609065, 9610072, 9704081, 9709090, 9710097, 9711103, 9712105, 9712106, 9801107, 9801109, 9802112, 9803001, 9803116, 9804126, 9804129, 9806136, 9807138, 9807422,9810527, 9901156, 9901160, 9901162, 9908167, 9909172, 9910176, 00709B, 0101003A, 01310B, 01525B, 01920B, 01921B, 01922B, 03041B,  03094B, 03338B, 03368B, 2145-30074A, 7718B, A0103015, A0103213, A0105217, A02100230, A02100231, A02110248, A02110249, A02110250, A02110251, A02120254, A02120256, A03010262, A03010264, A03010265, A03010266, A03010270, A03020274, A03020275, A03030282, A03070283, A03120294, B0003196, B0007200, B0309014, ID3935B, M@10639, M@10645, M@10774, M@10777, M@10877, M@11239, M@11568, M@11640, M@11684, M@11685, M@11729, M@11769, M@11813, M@11947, M@12378-1, M@12462, M@12765, M@13725, M@24470, M@4750028851, M@4750028853, M@5348B, M@5807B, M@5874B, M@5894B, M@5932B, M@5941B, M@5963B, M@6041B, M@6062B, M@6229B, M@6354B, M@6477B, M@6485B, M@6546B, M@6555B, M@6692B, M@6754B, M@6809B, M@6815B, M@6948B, M@7007B, M@7023B, M@7093B, M@7194B, M@7198B, M@7207B, M@7214B, M@7280B, M@7285B, M@7286B, M@7408B, M@7540B, M@7596B, M@7675B, M@7704B, M@8156B, M@8324B, M@8344B, M@8429B, M@8523B, M@8578B, M@8588B, M@8637B, M@8639RB, M@8648B, M@8808B, M@8813B, M@8898B, M@8977B, M@9050B, M@9261B, M@9276B, M@9355RB, M@9475B, M@9544B, M@9647B, M@9656B, M@9733B & M@9796B.    ADAC ARGUS: Model #88295; Serial #: 9211014, 9301005, 9304019, 9304024, 9304028, 9401028, 9401031, 9401037, 9505014, 9910033, 93-10678- 1292, 9910028P, M@10225AMT, M@3732B, M@3903B, M@3954B, M@3956B, M@3960B, M@4249B, M@4415B, M@4646B, M@4868B, M@5172B, M@5321B, M@7341B & M@AMT1154.    ADAC CIRRUS: Model #88294; 34, 718558, 9206009, 9207003, 9207004, 9207005, 9207007, 9211006, 9300113, 9304013, 9309012, 9404022, 9407028, 9509042, 9604098, 9910030, 93-10735-0393, 9621RQ, M@10110AMT, M@10324AMT, M@10402AMT, M@3599G-08, M@3905Q, M@4029Q, M@4129Q, M@4142Q, M@4668Q, M@4743Q, M@5003Q, M@5668Q, M@5895Q, M@AMT126R, M@AMT1282 & M@MVS3227A.",
      "center_classification_date": "20140117",
      "distribution_pattern": "Worldwide Distribution: US (nationwide) including states of:  AL, AR, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TN, TX, VA, VT, WA, WI & WV; and Internationally to: Austria, Belgium, Canada, Denmark, Germany, Great Britain, Italy, Netherlands, Norway, Pakistan, Russia, Singapore & South Korea.",
      "state": "OH",
      "product_description": "ADAC ARGUS Gamma Camera Systems & ADAC CIRRUS Gamma Camera Systems.  Distributed by Philips Medical Systems, Cleveland, OH",
      "report_date": "20140129",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-0785-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67130",
      "termination_date": "20151106",
      "more_code_info": "",
      "recall_initiation_date": "20131204",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}