      "status": "Terminated",
      "city": "Batesville",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "89562",
      "recalling_firm": "Hill-Rom, Inc.",
      "address_1": "1069 State Road 46 E",
      "address_2": "N/A",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AT, CA, CZ, DE, DK, ES, FI,  FR, GB, HK, IE, IL, IT, JP, NL, NO, NZ, RU, SE, SG, SK.",
      "recall_number": "Z-0783-2022",
      "product_description": "(1)Traverse Rail Carrier E-System, wide\t31017XX\t component to Liko overhead lift,  (2)Traverse Rail Carrier\t31025XX\t  component to Liko overhead lift.   Intended for to make safe lifting of a patient possible.    Affected Devices:  3101704 TRAV RAIL CARR E-SYST STD   3102514 TR CARRIER SLIM/RAISED, LR  3101705 TRAV RAIL CARR ESYST WIDE  3102517 TR CARR. NORMAL/LOWER. LR  3102506 TR CARRIER WIDE/RAISED ML   3102519 TR CARR. LOW.LR 30MM/PAIR  3102511 TR CARRIER NORMAL, LR   3102531 TRAVERSE RL CARRIER STD/RAISED  3102512 TR CARRIER WIDE, LR   3102532 TRAVERSE RL CARRIER WD/RAISED  3102513 TR CARRIER SLIM, LR   3102562 TR CARRIER, PARALLEL/WIDE",
      "product_quantity": "Traverse Rail Carrier E-System, wide\t31017XX\t5  Traverse Rail Carrier\t31025XX\t1798",
      "reason_for_recall": "Traverse rail carriage delivered with non-conforming screws are too short to ensure appropriate installation. There is potential that the screws are unable to secure the traverse rail appropriately and  could loosen over time causing the rail system to fall, potentially causing injury to caregivers and/or patients",
      "recall_initiation_date": "20220114",
      "center_classification_date": "20220311",
      "termination_date": "20241107",
      "report_date": "20220323",
      "code_info": "Distributed between 04JAN2021 and 24MAR2021. P/N  Product name  3130001  Multirall 200  3301030  LikoGuard L 3301040  LikoGuard XL  3121001  Likorall 200 3122005  Likorall 242 ES natural 3122005NA Likorall 242 ES natural 3122006  Likorall 242 ES white 3122006NA  Likorall 242 ES white 3122009  Likorall 242 S natural 3122009NA  Likorall 242 S natural 3122010 Likorall 242 S white 3122010NA Likorall 242 S white 3122007 Likorall 242 ES R2R, natural 3122007NA Likorall 242 ES R2R, natural 3122008 Likorall 242 ES R2R, white 3122008NA Likorall 242 ES R2R, white 3122011 Likorall 242 S R2R, natural 3122011NA Likorall 242 S R2R, natural 3122012 Likorall 242 S R2R, white 3122012NA Likorall 242 S R2R, white 3123001 Likorall 243 ES, natural 3123001NA Likorall 243 ES, natural 3123002 Likorall 243 ES, white 3123002NA Likorall 243 ES, white 3122501 Likorall 250 ES, natural 3122501NA Likorall 250 ES, natural 3122502 Likorall 250 ES, white 3122502NA Likorall 250 ES, white 3124050 Likorall 250 S, IRC, natural"
    }
  ]
}