  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84433",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of  AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, and WI.",
      "recall_number": "Z-0782-2020",
      "product_description": "Cios Alpha, Model Number 10308191;  mobile X-Ray system designed to provide X-ray imaging of the anatomical structures of patient during clinical application",
      "product_quantity": "190  (updated total of 202 as of 1/20/20)",
      "reason_for_recall": "During a routine inspection of a system, a crack on the C-arm holder was discovered.  In the event the C-arm holder has a crack, the stability of the C-arm may be compromised and could lead to deformations of the C-arm. In very unlikely cases, an extreme impact to the C-arm such as heavy collision with a wall or table, may lead to a detachment of the C-arm and could potentially result in injury to patients and staff.",
      "recall_initiation_date": "20191125",
      "center_classification_date": "20200107",
      "termination_date": "20200506",
      "report_date": "20200115",
      "code_info": "Serial Numbers: 11294  11202  12009  11141  11163  11032  11311  11260  11258  12530  12334  11376  12426  11336  11132  11295  11044  11005  11333  11386  12597  10076  12677  10115  11291  12523  11056  12319  12623  12512  12574  12647  12014  12452  12306  12317  11119  12576  12678  12513  12335  12490  12657  11200  11281  11283  12425  12023  12396  11285  11205  11103  11105  11257  11074  12379  11013  10021  11330  11066  10057  11143  11077  11319  11240  11349  11098  12339  11153  11372  10037  11328  12529  12308  10063  11106  12373  12584  11067  10082  12026  12027  11168  12492  10019  11317  10035  11198  12040  12307  12440  11165  12611  11094  11226  11370  12377  11348  11009  11018  10058  12321  12004  11378  11167  11248  11306  11229  12028  11144  12313  11287  12654  12043  11038  12320  12514  11352  12612  11316  11243  11314  12626  11374  11162  11292  12394  12498  12549  11120  11134  12494  12526  12528  11219  11220  11346  12318  12562  12607  12644  12685  12025  12046  11363  11288  11356  12620  11124  11251  12322  12323  12617  12533  12504  11015  11016  12486  12515  12500  11104  11327  10096  12391  11315  11075  12622  12652  11290  12416  12472  12496  12552  10116  12578  12642  12599  11280  11035  12003  12655  12030  11359  11107  12314  12618  11338  11064  12457  12460    Serial Numbers added as of 1/20/20:  10078  10089  11082  11080  11246  11380  11379  12006  12557  12718  12735  12732"
    }
  ]
}