  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Blaine",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "75459",
      "recalling_firm": "Nuvectra",
      "address_1": "10675 Naples St NE",
      "address_2": "N/A",
      "postal_code": "55449-5802",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: Nationwide (CA, TX, KY, FL, IL,PA,OH, GA, NJ, NC, KS, AR, CO, NE, NY, VA, AZ, OK).    OUS: Germany",
      "recall_number": "Z-0781-2017",
      "product_description": "Algovita Spinal Cord Stimulation System,  Percutaneous Leads, 8-Electrode Percutaneous Lead Kit;    Algovita Spinal Cord Stimulation System Percutaneous Leads, 12-Electrode Percutaneous Lead Kit;   Algovita Spinal Cord Stimulation System Percutaneous Trial Leads, 8-electrode Lead Kit;  Algovita Spinal Cord Stimulation System Percutaneous Trial Leads, 12-electrode Lead Kit.   Model Number (s)  112x-xx (CT), 108x-xx(T).   These are part of the Algovita Spinal Cord Stimulation System.    The Algovita Spinal Cord Stimulation system consists of a stimulator (EPG or IPG) that is physically and electrically connected to one or more leads inserted into the patient s spinal epidural space. The lead delivers electrical stimulation originating at the stimulator with the purpose of blocking pain signals going to the patient s brain. The stimulators are rechargeable, and the frequency of recharging is dependent on  individual patient s use of the system.",
      "product_quantity": "406 OUS, 985 US",
      "reason_for_recall": "Nuvectra is conducting a recall due to two clinical risks that are  associated with certain percutaneous leads currently being used in the Algovita Spinal Cord Stimulation System.  It has been reported that in some percutaneous leads, the stylet protruded beyond the distal lead tip.",
      "recall_initiation_date": "20161011",
      "center_classification_date": "20161217",
      "termination_date": "20180808",
      "report_date": "20161228",
      "code_info": "Lot Numbers:  W2886402  , W2888100  , W2904692, W2909622  , W2909623  , W2912562  , W2918496, W2918506  , W2930816, W2930816  , W2938384  , W2939156  , W2943651  , W2946424  , W2952157  , W2964008  , W2964009, W2964009  , W2964010, W2964010  , W2964011  , W2972187  , W2978500  , W2981155, W2981155  , W2988382, W2988382  , W3040781  , W3041965, W3041965  , W3043467, W3048908, W3052511  , W3052514, W3052515  , W3057023, W3057024, W3057027, W3057033, W3057033  , W3063066, W3063073, W3063073  , W3088964, W3110760  , W3111249, W3114108, W3119243, W3123219  , W3123220, W3123220  , W3124197, W3127472, W3127472  , W3127473, W3133254, W3133264  , W3135521, W3232789, W3232789  , W3232790, W3232791, W3232792, W3232792  , W3236234, W3236235, W3236237, W3236239, W3236239  , W3321458  , W3321459  , W3321460  , W3321461  , W3321463  , W3321464, W3324818  , W3324820, W3324821  , W3324826  , W3324827, W3324829  , W3324830  , W3335022, W3335022  , W3335027, W3335033  , W3335041  , W3335050, W3335050  , W3335055, W3335055  , W3343777, W3343777  , W3348715  , W3356371, W3356371  , W3357155, W3357155  , W3357157  , W3357159, W3357159  , W3357160  , W3357162, W3357164  , W3357167, W3357169  , W3362186  , W3362187, W3362187  , W3362188, W3362188  , W3362209  , W3374818  , W3374820, W3374821  , W3374823  , W3378359, W3378360, W3389104, W3389105  , W3389112, W3389117  , W3393665  , W3393666  , W3393667, W3393668  , W3396586  , W3396603, W3396608, W3396609  , W3399272  , W3424531, W3424531  , W3424542, W3424565  , W3424585, W3424586, W3424586  , W3424589, W3424591  , W3424593, W3424599, W3424602, W3424616  , W3424623  , W3424627  , W3442616  , W3442617, W3442617  , W3442618, W3442618  , W3442619  , W3448170, W3448171, W3448172  , W3448173  , W3448174  , W3454201  , W3454202  , W3462791  , W3462792  , W3462794, W3462795, W3462795  , W3462796, W3462797, W3462798, W3472638, W3472639, W3472639  , W3472640  , W3472641, W3472641  , W3475288, W3475289, W3475289  , W3475290, W3475290  , W3475291, W3475291  , W3493137, W3498560, W3498562, W3498564, W3504760, W3504762, W3504763, W3504766  , W3506847, W3506848, W3506851, W3536038  , W3536039, W3536039  , W3539484, W3540924."
    }
  ]
}