  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Arden",
      "state": "NC",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "84435",
      "recalling_firm": "Medical Action Industries, Inc. 306",
      "address_1": "25 Heywood Road",
      "address_2": "N/A",
      "postal_code": "28704-9302",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "MA, RI, PA, MD, VA, NC, SC, GA, FL, KY, OH, IN, MI, IL, MO, NE, LA, TX, AZ, CA, WA",
      "recall_number": "Z-0779-2020",
      "product_description": "Medical convenience trays and kits, labeled as:    Medical Action Industries, Inc.:  ***CVC*** Clear Sequence Change Kit, REF 77070B  ***DIALYSIS*** Clear Sequence Change Kit, REF 77068B  ***PEDS CVC*** Clear Sequence Change Kit, REF 77190B  ***PICC*** Clear Sequence Change Kit, REF 77069B  Adult Central Line Dressing Change Kit, REF 74636C  Central Line Dressing Kit, VCU, REF 57647B  Central Line Removal Kit, REF 78631B  Chest Tube Cleaning Kit, REF 73000  CVC Dressing Change Kit, REF 80643D  Dressing Change Kit, REF 71538  ED Suture Set, REF 59299C  Laceration Kit, REF 59806B  Laceration Tray, REF 56957  Laceration Tray, REF 57888B  Laceration Tray, REF 59387  Laceration Tray, REF 59387B  Laceration Tray, REF 69477B  Laceration Tray, REF 71582  LACERATION/SUTURE REMOVAL, REF 72347  LACERATION/SUTURE REMOVAL, REF 72347B  Pediatric Central Line Dressing Kit, REF 74635C  PICC Central Line Dressing Kit VCU, REF 59589C  PICC/CVC/Central /Dialysis Clear Sequence Dressing Kit, REF 78197  Suture Kit, REF 59836  Trach Care Kit, Kit de Traqueostomia, REF LH_68162B  Trach Care Tray, REF 56791B  Trach Care Tray, REF 75378  Tracheostomy Care Tray, REF 4129  Tracheostomy Tray, REF 80209  Wound Care Kit, REF 68330  Wound Care Tray, REF 68358    MEDICHOICE:   Tracheostomy Kit, Reorder: TRAC1002    THERAPY SUPPLY HOUSE:   VAD DRESSING CHANGE KIT, Reorder # 76512B",
      "product_quantity": "9924 cases",
      "reason_for_recall": "The kits contain a bottle of saline solution which may exceed the USP specified pH requirements of 4.5-7.0.",
      "recall_initiation_date": "20191202",
      "center_classification_date": "20200106",
      "termination_date": "20230124",
      "report_date": "20200115",
      "code_info": "***CVC*** Clear Sequence Change Kit, REF 77070B  Lot Numbers:  0000250140  0000250814  0000251345  0000252860  0000253786  0000254661  0000255539  0000256372  0000256478  0000256896  0000257916  0000259710  0000260807  0000261950  0000262411  0000262853  0000263255  0000248930  0000248004  0000245391    ***DIALYSIS*** Clear Sequence Change Kit, REF 77068B  Lot Numbers:  0000255949  0000256795  0000258076  0000260410  0000261183  0000262656  0000257187    ***PEDS CVC*** Clear Sequence Change Kit, REF 77190B  Lot Numbers:  0000261186  0000261949  0000263402  0000245684    ***PICC*** Clear Sequence Change Kit, REF 77069B  Lot Numbers:  0000250126  0000250752  0000251646  0000252449  0000255455  0000256371  0000256796  0000256897  0000257625  0000258163  0000260908  0000262404  0000262852  0000249190  0000248773    Adult Central Line Dressing Change Kit, REF 74636C  Lot Numbers:  0000250172  0000251684  0000251775  0000252730  0000255062  0000255669  0000256289  0000257970  0000259210  0000259867  0000260734  0000263279  0000264415  0000248260    Central Line Dressing Kit, VCU, REF 57647B  Lot Numbers:  0000249224  0000248261    Central Line Removal Kit, REF 78631B  Lot Numbers:  0000258405  0000264395    Chest Tube Cleaning Kit, REF 73000  Lot Numbers:  0000254875  0000256642  0000257028  0000261394  0000248107  0000247565  0000246033  0000244487    CVC Dressing Change Kit, REF 80643D  Lot Numbers:  0000263867    Dressing Change Kit, REF 71538  Lot Numbers:  256205  0000256888    ED Suture Set, REF 59299C  Lot Numbers:  0000250820  0000252658  0000254734  0000255276  0000257072  0000260553  0000262731    Laceration Kit, REF 59806B  Lot Numbers:  0000250817  0000252940  0000253601    Laceration Tray, REF 56957  Lot Numbers:  0000255242  0000257303  0000258844  0000260451  0000263359    Laceration Tray, REF 57888B  Lot Numbers:  0000255947  0000257545  0000257931  0000259064  0000260150  0000261007  0000262982  0000263363    Laceration Tray, REF 59387  Lot Numbers:  0000248176  0000247839    Laceration Tray, REF 59387B  Lot Numbers:  0000255606  0000256080  0000256505  0000259051  0000259659  0000260338  0000261192  0000262714  0000263075    Laceration Tray, REF 69477B  Lot Numbers:  0000256510  0000261832  0000263335    Laceration Tray, REF 71582  Lot Numbers:  0000250813  0000252624  0000248502    LACERATION/SUTURE REMOVAL, REF 72347  Lot Numbers:  0000250772  0000248306    LACERATION/SUTURE REMOVAL, REF 72347B  Lot Numbers:  0000256124  0000258422  0000259257  0000260325  0000261864  0000263377  0000263889    Pediatric Central Line Dressing Kit, REF 74635C  Lot Numbers:  0000254205  0000255191  0000258646  0000259884  0000263819  0000249687  0000248262    PICC Central Line Dressing Kit VCU, REF 59589C  Lot Numbers:  0000251690  0000252258  0000248259    PICC/CVC/Central /Dialysis Clear Sequence Dressing Kit, REF 78197  Lot Numbers:  0000262582  0000263685    Suture Kit, REF 59836  Lot Numbers:  0000250818  0000252923  0000253808  0000254794  0000258842  0000260452  0000263353    Trach Care Kit, Kit de Traqueostomia, REF LH_68162B  Lot Numbers:  0000256422  0000258614    Trach Care Tray, REF 56791B  Lot Numbers:  0000252599  0000252842  0000253529  0000255449  0000257898  0000258435  0000263562  0000257350    Trach Care Tray, REF 75378  Lot Numbers:  0000250819  0000252601  0000255861  0000257905  0000259371    Tracheostomy Care Tray, REF 4129  Lot Numbers:  0000253196  0000254514  0000255528  0000256340  0000258894  0000261328    Tracheostomy Tray, REF 80209  Lot Numbers:  0000253370  0000256337  0000257128  0000257903  0000258437  0000259368  0000262397    Wound Care Kit, REF 68330  Lot Numbers:  0000254655  0000255948  0000257266  0000257856  0000259029  0000259816  0000261467  0000263582    Wound Care Tray, REF 68358  Lot Numbers:  0000255446  0000257267  0000259030  0000259817  0000264334      MEDICHOICE:   Tracheostomy Kit, Reorder: TRAC1002  Lot Numbers:  0000253474  0000255267  0000256338  0000259025  0000259351  0000262396  0000263581      THERAPY SUPPLY HOUSE:  VAD DRESSING CHANGE KIT, Reorder # 76512B  Lot Numbers:  0000252929  0000254713"
    }
  ]
}