  "results": [
    {
      "status": "Terminated",
      "city": "Braintree",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79031",
      "recalling_firm": "Haemonetics Corporation",
      "address_1": "400 Wood Rd",
      "address_2": "N/A",
      "postal_code": "02184-2412",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution to states of: AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, LA, MA, MD, ME, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, and WI.",
      "recall_number": "Z-0778-2019",
      "product_description": "Cell Saver 5/5+ Bowl Set, 70ml, Product ID: 0291E-00, 0291A-00    Indications for Use: The Haemonetics Cell Saver 5/5+ System and its related accessory components are intended to be used for the recovery and processing of blood shed during or subsequent to an operation or as a result of trauma.",
      "product_quantity": "11,208 units",
      "reason_for_recall": "Haemonetics has received isolated reports that in some cases the 70ml centrifuge bowl may leak during cell salvage processing.",
      "recall_initiation_date": "20170802",
      "center_classification_date": "20190128",
      "termination_date": "20200806",
      "report_date": "20190206",
      "code_info": "Catalog number: 0291A-00, 0291E-00"
    }
  ]
}