      "country": "United States",
      "city": "Waukesha",
      "address_1": "3000 N Grandview Blvd",
      "reason_for_recall": "On the Revolution CT systems equipped with the SmartStep Option, the Z location displayed on image annotation and on the gantry display may not be the actual location of the table and result in the system scanning at a location that is not what was confirmed by the operator",
      "address_2": "",
      "product_quantity": "131 units",
      "code_info": "SmartStep Option",
      "center_classification_date": "20190206",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide Distribution) and countries of: Australia, Belgium, Canada, Estonia, Finland, France, Germany, Hong Kong, Ireland, Japan, Korea Republic of, Malaysia, Morocco, New Zealand, Norway, Saudi Arabia, Singapore, Slovakia, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "state": "WI",
      "product_description": "Revolution CT systems with the SmartStep Option    The system is intended for head, whole body, cardiac and vascular X-ray Computed Tomography applications.",
      "report_date": "20190213",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "GE Healthcare, LLC",
      "recall_number": "Z-0776-2019",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "81922",
      "more_code_info": "",
      "recall_initiation_date": "20180712",
      "postal_code": "53188-1615",
      "voluntary_mandated": "FDA Mandated",
      "status": "Ongoing"
    }
  ]
}