  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79132",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic and Foreign",
      "recall_number": "Z-0776-2018",
      "product_description": "Intellivue iX Information Center Software - All PIIC iX Surveillance stations including:  866023 IntelliVue Info Center iX A.0  866117 PIIC Classic Upgrade  866389 IntelliVue Info Center iX B.0, C.0  867141 IntelliVue Info Center iX B.0  866025 IntelliVue Server IX A.0  866118 Database Server Upgrade A.0",
      "product_quantity": "20357",
      "reason_for_recall": "Once a surveillance station is restarted on January 1, 2018 or later, the station will be unable to perform patient discharge and transfer operations. Any subsequent attempt to perform these operations will cause the station to restart, resulting in a short period of loss of monitoring at the Surveillance station during such restart. Until this issue can be corrected, users should avoid intentionally restarting their Surveillance stations in 2018.",
      "recall_initiation_date": "20180103",
      "center_classification_date": "20180228",
      "termination_date": "20191114",
      "report_date": "20180307",
      "code_info": "All PIIC iX Surveillance stations including:  866023 IntelliVue Info Center iX A.0  866117 PIIC Classic Upgrade  866389 IntelliVue Info Center iX B.0, C.0  867141 IntelliVue Info Center iX B.0  866025 IntelliVue Server IX A.0  866118 Database Server Upgrade A.0"
    }
  ]
}