      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63531",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.",
      "recall_number": "Z-0775-2013",
      "product_description": "Trilogy Acetabular Shells with Holes.    The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.",
      "product_quantity": "275,952 units",
      "reason_for_recall": "Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.",
      "recall_initiation_date": "20120914",
      "center_classification_date": "20130204",
      "termination_date": "20140620",
      "report_date": "20130213",
      "code_info": "61567730  61558081  61611792  61559843  61567731  61611793  61215883  00369728  00369696  61930401  61459870  61611794  61564135  61201896  61292658  61783434  61915194  00369729  00369680  61921675  61638695  61915195  61906180  61808994  61425796  61580824  61471012  61263144  61201897  00369730  00370028  00370201  00370605  00370449  00369628  61996704  61779026  61533558  61758219  61455810  61452211  61911378  61596869  61734419  61783435  61794928  61201845  61783436  62055820  61580825  61825846  62022693  61580826  61638696  61889255  61263146  61281167  61321048  61329084  61471011  61638699  61896724  61921677  61283615  61309774  61215878  61216899  61638698  61855181  61906445  62052789  61374925  61638697  61199815  61199814  61403503  61412019  61486653  61521911  61533559  61580827  61633747  61817461  61921676  61957085  61263145  61412020  61470969  61486654  61809006  61292660  61237662  61374926  61452210  61528582  61837767  61873598  61930403  61945055  62006515  62052788  00370200  00369694  00370606  61732766  62063917  61374929  61454830  61360114  61500667  61808989  61996705  61734247  61635717  61640790  61920504  61716307  62022695  61215901  61949819  61215902  61397892  61486655  61580830  61635718  61779021  61882566  61638701  61709187  61884425  61201851  61271311  61374927  61454832  61483179  61533563  61761996  61783437  61825847  61832823  61957087  61957088  62038296  61803761  61846473  61873601  61999739  61425778  61439219  61455776  61528584  61533562  61580828  61638702  61640789  61750751  61846472  61878440  61528585  61783438  61945056  62055823  61281168  61283616  61292661  61368469  61439218  61454831  61486656  61533561  61542694  61542695  61580832  61580833  61640786  61640787  61820780  61846474  61860441  61895514  61941700  62063918  61397895  61454829  61709189  61842601  61860440  61867233  61911379  61216902  61263148  61263149  61269930  61271310  61283617  61309777  61374928  61397896  61412021  61497356  61533560  61564136  61580831  61611812  61638703  61716305  61761995  61803760  61808988  61867232  61906446  61921678  61947557  62022696  61825848  61939314  61216901  61260601  61292662  61309776  61329085  61340386  61342477  61374930  61397894  61403504  61455777  61455778  61455781  61489305  61497357  61521912  61580829  61580834  61640788  61709188  61732765  61744554  61750750  61779020  61842602  62036845  62055824  62063916  61412022  00369731  00370027  00368634  00369658  61454834  61867235  61640791  61716308  61957090  61580837  61716309  61867234  61542693  61928497  61734250  61215868  61215869  61246506  61564167  61350515  61260602  61263150  61271313  61528579  61770602  61779024  61967996  61808993  61820783  61895506  61283618  61500670  61536405  61580838  61638704  61770601  61794933  61794934  62017509  61360115  61321052  61901543  61216904  61237664  61246444  61360116  61439201  61476884  61822463  61860442  61911382  61957092  62022697  62055830  61216903  61263151  61580839  61873599  61397901  61237666  61309778  61397899  61403505  61425783  61439196  61439198  61439199  61476882  61528580  61542692  61580835  61638705  61734249  61750752  61768195  61837768  61882567  61911381  61916321  61943334  61947558  61199819  61237667  61271312  61292397  61292663  61309779  61324687  61329086  61397898  61412023  61412024  61439197  61454833  61470971  61471010  61476883  61476885  61476886  61500669  61528581  61536406  61536407  61580836  61638706  61762003  61846485  61939315  61941696  61945058  62055825  62055827  00369700  00369734  61215879  61237671  61580844  61901544  61732769  61957095  61980772  61761997  61397907  61263532  61749006  61809001  61911383  61926573  61996707  62022701  61911384  61627439  61790422  61867239  62036843  61926350  62027429  61237672  61403506  61500672  61524168  61627441  61717971  61749005  61761998  61896725  61957096  61978758  61980770  61980771  61638708  61916322  61215880  61237668  61417354  61511483  61511485  61542697  61542698  61542701  61580841  61638709  61678476  61717972  61717974  61750749  61790421  61864066  61896726  61896727  61383431  61867240  61327750  61839042  61957094  61627440  61309781  61367250  61439187  61439188  61439190  61482140  61524169  61524171  61536409  61536412  61580840  61717973  61732771  61788116  61794930  61825850  61842605  61842606  61867236  61867241  61882568  61949818  61999741  62017508  61435510  61580842  61996706  61999740  62022698  61256572  61292398  61292667  61350517  61374931  61407863  61439185  61439186  61470974  61482145  61500671  61524170  61536408  61542696  61542699  61564168  61638707  61638710  61678475  61744555  61775726  61775728  61779022  61822464  61825849  61846476  61846477  61855182  61882569  61926574  61926575  61930400  61939316  61999742  62022699  61327751  61350516  61425784  61822467  61867238  61941703  61215881  61237669  61237670  61246457  61246507  61246508  61260603  61263533  61263534  61271314  61271315  61271316  61281169  61281170  61283619  61283620  61292664  61292666  61309780  61321054  61361689  61374932  61374935  61374936  61383432  61397903  61397904  61397905  61397906  61407862  61412027  61417353  61425786  61435506  61435511  61439184  61439189  61439191  61439192  61470972  61482142  61482143  61482144  61482146  61486657  61486658  61486659  61486660  61489306  61497387  61500673  61511484  61536410  61536411  61580843  61724211  61770597  61775727  61846478  61895508  61906137  61945072  62079955  00369631  61822470  62043431  61790423  61952342  62043429  61612717  61955585  61980774  61980773  62060654  61825852  61567732  61761999  61486665  61564139  61632191  61783440  62006516  61500675  61564171  61976097  61486661  61744556  61783439  61987181  62010598  61885837  61941704  61486666  61536683  61536699  61564169  61612716  61675890  61739596  61775732  61921679  62006517  61256573  61367252  61439183  61486664  61511486  61536701  61638711  61717977  61739597  61749007  61846486  61869686  61911386  61638712  61281171  61374933  61429452  61435514  61486667  61511487  61536698  61678478  61717976  61770599  61809009  61895510  61397909  61407867  61439165  61536682  61762001  61825851  61191659  61260605  61263535  61292668  61292669  61309782  61342478  61374934  61397910  61397911  61412028  61435505  61439182  61500674  61564170  61860443  61873602  61891241  61911385  61951204  61327752  61439166  61598279  61920508  61263536  61435504  61186232  61191658  61201882  61201883  61211984  61211985  61237674  61271317  61271318  61283621  61292399  61309783  61383433  61417355  61439163  61439164  61486663  61536684  61564137  61564138  61564140  61717978  61820785  61832829  61851831  61869685  62038298  61439181  00369873  61896728  61739601  61640795  61536687  61237678  61517429  61633897  61640799  61744557  61891243  61500676  61976098  62060655  61598282  61633898  61966562  61564141  61598283  61640793  61640800  61775733  61820782  62006519  62052786  61306602  61439180  61598281  61846481  62010601  61869689  62006518  61899039  61813944  61383435  61076208  61271319  61374938  61486672  61500677  61511465  61564142  61564144  61640797  61779028  61783443  61846479  61928498  61941705  62021103  61783442  61246509  61256574  61281172  61292401  61292670  61383434  61403508  61407869  61412029  61417356  61425791  61439178  61486670  61511489  61536685  61536686  61564172  61739602  61775734  61803762  61851829  61885839  61980775  61439173  61439174  61511488  61564173  61640801  61846480  61951205  61955586  61237675  61237676  61237677  61260606  61263537  61263538  61283622  61309784  61309785  61321058  61327753  61329087  61350518  61374937  61403507  61407870  61439172  61439175  61439176  61439177  61439179  61473714  61482147  61486671  61489307  61511490  61564143  61564174  61640794  61640796  61640798  61762004  61839041  61869690  61891242  61911387  61920505  62052787  00369604  00369766  61734252  62036841  61292671  61292672  61307473  61307471  61439167  61980776  61996709  61941685  61271320  61976099  61511491  61640807  61895507  61350514  61417362  61598284  61640805  62094938  61439195  61564146  61564175  61567573  61775731  61820781  61839040  61846484  61891244  61920502  61367253  61638713  61536688  61536690  61640806  61739599  61882570  61808996  61885840  61911388  61237682  61199824  61439171  61473722  61536691  61779023  62076896  61794929  61869688  61920503  61439193  61407871  61212919  61215862  61283623  61389712  61402340  61412030  61476887  61476889  61476891  61567574  61717979  61941708  61350513  61439170  61201886  61318327  61374942  61491778  61536689  61564176  61638714  61640803  61739598  61808995  61996708  62052784  61237683  61832827  61191662  61212918  61237680  61237681  61246510  61260607  61269931  61281173  61281174  61318325  61321061  61374939  61374940  61374941  61389711  61425779  61439168  61439169  61439194  61473723  61489308  61491777  61564145  61567575  61638715  61640804  61762002  61775730  61851830  61869687  00369765  00369634  61642166  61642159  61642158  61891247  61952348  61612720  62060651  61851828  61526490  61964698  61611261  61612719  61996710  61803764  61511673  61526489  61532283  61611255  61612722  61642163  61808990  61487476  61820784  61983677  61292402  61374944  61724375  61532285  61237688  61439209  61489311  61532284  61642156  61711432  61885841  61389713  61837769  61906447  61271321  61567733  61611258  61612721  61633902  61638717  61642160  61837770  61891246  61869691  62021100  61271322  61374943  61439211  61486683  61489309  61511674  61521913  61535278  61567734  61638716  61642161  61642162  61732768  61734248  61749008  61920499  62036839  62060652  61341071  61439210  61201889  61237689  61260608  61389714  61412031  61417363  61425787  61439212  61486679  61489310  61535279  61564147  61724376  61724377  61732767  61743363  61783452  61846483  61864067  61869692  61911389  61920498  61952345  61964699  61980777  61996711  62060653  62076898  61283624  61403510  61471008  61567735  61899038  60750944  61186234  61201888  61201890  61235816  61235817  61237684  61237685  61237686  61237687  61246511  61256575  61269932  61269933  61281175  61292673  61292674  61292675  61309786  61309787  61321064  61329088  61329089  61341072  61389715  61389716  61407872  61425788  61425789  61425790  61439208  61439213  61455813  61455834  61455835  61480102  61486682  61487474  61487475  61611254  61611257  61612718  61633901  61638718  61743364  61770598  61775729  61783450  61794931  61808991  61837771  61851827  61921680  61953113  61955587  62010600  62021101  62050961  61403512  00369738  00369635  61548179  61928505  61532294  61548180  61642333  61803766  61983681  62036846  61237716  61309792  61403529  61412049  61889258  61256571  61424318  61455846  61476898  00369739  00370054  00369637  61882571  61638719  61983678  61758220  61292678  61439215  61509780  61788118  61932004  61642196  61864068  61895512  62010602  61564180  62021102  61476895  62043436  61611267  61389721  61407874  61425794  61439217  61526504  61564155  61817464  61883391  61920510  61201892  61216905  61237701  61532291  61642197  61199828  61269935  61271324  61321065  61321066  61329091  61389720  61439214  61439216  61476894  61500679  61526505  61564156  61612731  61642195  61794935  61813951  61425795  00370053  00369740  00370607  00369840  61329092  61489312  61983679  61928499  61878442  62010603  61389722  61564181  61612732  61452208  61511675  61564157  61642203  61788119  62027422  61201893  61292679  61407875  61452209  61532292  61642202  61321067  61216906  61425793  00370199  00370026  00370872  00369743  00370608  00369764  61564158  61459872  61186240  61471009  61611268  61329093  61283626  00369747  00370056  00370450  00370430  00369484  00369608  61487480  61511676  61318672  61321068  00370055  00370434  61564159  61459871  00369704  61168262  00369763  00369874  00369748  00369611  61450088  62245444  62155060  61375769  61384935  61461918  07892003  61784841  61442054  61490803  61553788  61622951  61676585  61693494  61730083  61761521  61821403  61384936  61456505  62255381  61498729  61791909  07892004  61236174  61227816  61269036  61278084  61278085  61333092  61375770  61410381  61442055  61480720  61490804  61531428  61549721  61553776  61596121  61622952  61648305  61676597  61684954  61714405  61730082  61737673  61754427  61384963  61416505  61559760  61456506  62143475  61231172  07892006  61236175  61730084  07892005  61222254  61231173  61271665  61403345  61442056  61480721  61490805  61532993  61553789  61596115  61615390  61639523  61648306  61676616  61693495  61748059  61784835  61865183  62073331  61469656  61384938  62136220  61456507  61499630  61761522  07892007  61498730  61857683  61684956  62093691  62126089  61206442  61821397  61231175  61236177  61422944  61615393  61610332  61843996  07891616  07891666  07891667  07892008  07892009  61222255  61231176  61231177  61236176  61236178  61236179  61269037  61271666  61278086  61403346  61419069  61419070  61422943  61422945  61442066  61490806  61490807  61532994  61532998  61550411  61550412  61559754  61581709  61585971  61629971  61629983  61645211  61648317  61658806  61676586  61680965  61693496  61709297  61721126  61730089  61742080  61754420  61772407  61807718  61955478  61962825  61973145  61986391  62011416  62016969  62042213  62054498  62073332  61384937  61384965  61456516  61469657  00113253  00114049  00114229  00115856  07892012  11001309  11001867  00108990  00113250  61680966  61498732  61498731  61714394  62115294  61389597  62078804  62084414  62115293  62115295  62136221  61222257  61456517  61227817  61442059  61831579  61278087  07892013  61222256  61236180  61236183  61236185  61236190  61419071  61490808  61629984  61658807  61676587  61684935  61693497  61865185  61962827  61980112  62067319  61236188  61389603  07891668  07891669  07892014  07892015  07892016  61222258  61222259  61227818  61236181  61236182  61236184  61236187  61236189  61236191  61236192  61269039  61271667  61271668  61278088  61338643  61346662  61346667  61389598  61389601  61389602  61403347  61410382  61416506  61416507  61416508  61422946  61422947  61442057  61442058  61450089  61480722  61480723  61480724  61490809  61490810  61490811  61531429  61532995  61532999  61543118  61550413  61550414  61559749  61559755  61566836  61574191  61574192  61581955  61589933  61596122  61610333  61610334  61615394  61622953  61622965  61639524  61645212  61648307  61656283  61667054  61676598  61693491  61709295  61709299  61714395  61721136  61726608  61730085  61730091  61737664  61742072  61748054  61748060  61754421  61754429  61761523  61772409  61791917  61796125  61813376  61819008  61821398  61831568  61847026  61851621  61860178  61955480  61962828  61973146  61986392  61992774  61998848  62011496  62016970  62042214  62049515  62052887  62056751  62056757  61456508  61389599  61389604  61461919  61469660  61469661  00108984  00114224  61574193  61684958  07892025  62093298  62084415  62084416  62101920  61360738  61227819  07892023  61278091  61680967  61721138  61726612  61333093  61442090  61553777  61847028  07891701  07891702  07892024  61206443  61222260  61236193  61236194  61236195  61236196  61236197  61236198  61236199  61236201  61269040  61271669  61278089  61278090  61333106  61349608  61354835  61354839  61354840  61366104  61375771  61375773  61375774  61410383  61429377  61429378  61442091  61442092  61450090  61476220  61498733  61498734  61543094  61559750  61566837  61581957  61589934  61589935  61610335  61615395  61622966  61639525  61648308  61656284  61667055  61676599  61693501  61714397  61730088  61737667  61748056  61754423  61769840  61769846  61796127  61819010  61831569  61865181  61986394  61998851  62011497  62052894  61469663  61389605  61389606  61456509  61461920  00109003  62215529  61222261  62136222  62185052  62101923  62030496  61236202  61236204  61349593  61375775  61680968  07891704  07892034  07892035  61227820  61236203  61269041  61271670  61278092  61333094  61338644  61354836  61354841  61354842  61360739  61366105  61389617  61403351  61450091  61450092  61476221  61559751  61576445  61589937  61589938  61622955  61639526  61656285  61676588  61709230  61714401  61726623  61742075  61754425  61769848  61796130  61821402  61860181  61962835  61992776  62255388  61389615  61456510  61461921  61354837  62093701  62159366  61366101  61236205  61589939  61622956  61680969  07892041  07892042  61206446  61222262  61236206  61236207  61271671  61333095  61349594  61349609  61375776  61410384  61450093  61553778  61566838  61656286  61726617  61754426  61791913  61819020  61962841  61986396  62030498  62245454  62262493  61456511  61476222  62241415  07892047  61360740  61476223  61222263  61236208  61269042  61330571  61349595  61410385  61450094  61566839  61596123  61656287  61721142  61780747  62262495  61456512  62257718  61800866  61354838  07892049  61236209  61278093  61338645  61403352  61610336  61693493  61456513  62255392  61360686  61813381  61236210  61476224  61403353  61278094  61222264  61236211  61366106  07892054  61456514  07892057  61456515  61203315  61269044  61236213",
      "more_code_info": ""
    }
  ]
}