  "results": [
    {
      "country": "United States",
      "city": "Rochester",
      "address_1": "150 Verona St",
      "reason_for_recall": "Carestream received a report of an exposure being initiated without a patient image captured. The patient was imaged 2 times before an image was captured. The field engineer who visited the site reported the system allowed an exposure without the assignment of a bucky or image receptor.",
      "address_2": "",
      "product_quantity": "4 units",
      "code_info": "Serial Numbers  5049, 5073, 5133, 5162",
      "center_classification_date": "20140129",
      "distribution_pattern": "Nationwide Distribution including TX, IL, and GA.",
      "state": "NY",
      "product_description": "Carestream DRX Evolution System (Standard Q)    The DRX-Evolution System (Standard Q) is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions. These products are permanently installed diagnostic x-ray systems composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstandbucky to complete x-ray exposures.",
      "report_date": "20140205",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Carestream Health Inc.",
      "recall_number": "Z-0774-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67060",
      "termination_date": "20160628",
      "more_code_info": "",
      "recall_initiation_date": "20130820",
      "postal_code": "14608-1733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}