{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marlborough",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "86909",
      "recalling_firm": "Boston Scientific Corporation",
      "address_1": "300 Boston Scientific Way",
      "address_2": "",
      "postal_code": "01752-1291",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution - Foreign: TBD.",
      "recall_number": "Z-0773-2021",
      "product_description": "AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x10mm UPN: M00553640 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.",
      "product_quantity": "13 units",
      "reason_for_recall": "The rotating luer fitting at the distal end of the  delivery system handle may become detached from the nose of the delivery  during device preparation, stent delivery or use, resulting in delay or procedure or  risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity",
      "recall_initiation_date": "20201201",
      "center_classification_date": "20210106",
      "termination_date": "20230714",
      "report_date": "20210113",
      "code_info": "Lot Number: 26134666  Exp. Date: 10/6/2022  GTIN:08714729904588",
      "more_code_info": ""
    }
  ]
}