      "country": "United States",
      "city": "Barrington",
      "address_1": "540 W Northwest Hwy",
      "reason_for_recall": "In Imagecast PACS versions prior to 3.6.51.01, the estimated magnification factor (0018,1114) was not accounted for. These versions of Imagecast PACS look for values to be present in the following tags to calculate a measurement from pixels to millimeters (mm): (0020,0030) Pixel Spacing and, (0018,1164) Imager Pixel Spacing.     Because the estimated magnification factor is not used, distance meas",
      "address_2": "",
      "product_quantity": "43 units",
      "code_info": "Imagecast PACS versions prior to 3.6.51.01",
      "center_classification_date": "20140115",
      "distribution_pattern": "Nationwide Distribution: US including states of: AL, AZ, CT, FL, GA, IA, IN, LA, MA, MI, MN, MO, NC, NH, NJ, NY, OH, PA, TX, VA, WI, WV + District of Columbia.",
      "state": "IL",
      "product_description": "Imagecast PACS is an image management system intended to be used by trained professionals, including but not limited to physicians, nurses and medical technicians. The system is used with general purpose computing hardware to acquire, transmit, process and store images and data throughout a clinical environment. Data and images are acquired through DICOM compliant imaging devices and modalities. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image and interpretations. Mammongraphic may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by the FDA.",
      "report_date": "20140122",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Ge Healthcare It",
      "recall_number": "Z-0773-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "64740",
      "termination_date": "20160516",
      "more_code_info": "",
      "recall_initiation_date": "20110304",
      "postal_code": "60010-3051",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}