      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63531",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.",
      "recall_number": "Z-0773-2013",
      "product_description": "Trilogy Acetabular Shell with Cluster Holes.    The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.",
      "product_quantity": "275,952 units",
      "reason_for_recall": "Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.",
      "recall_initiation_date": "20120914",
      "center_classification_date": "20130204",
      "termination_date": "20140620",
      "report_date": "20130213",
      "code_info": "00369733  00369629  61216911  61750773  61770610  61734420  61880355  61941709  61687733  61076227  61489318  61716316  61825784  62063923  61215851  61216909  61263156  61483182  61655830  61709193  61825783  61855188  61957104  61580848  61943340  62065253  62076915  61215855  61655829  61655831  61783469  61873616  61883392  61883393  61964702  61978754  61439036  61911394  62052806  61957102  61215850  61263155  61374949  61511679  61536418  61665446  61716314  61734423  61768199  61783470  61801116  61860445  61864074  61864075  61916323  61938318  61991357  61435495  62010623  61967998  62063924  61216912  61256547  61310909  61483181  61528587  61580847  61580856  61655832  61665445  61673515  61687736  61687737  61687738  61709191  61709195  61744559  61750775  61783468  61794947  61817467  61825786  61846496  61936095  61941382  61999744  61999746  62065252  61237816  61846498  61911393  62052805  61066740  61237813  61237818  61237819  61246512  61246513  61281211  61292688  61350523  61360120  61407892  61417319  61417320  61439040  61487484  61489317  61497389  61511678  61511682  61528591  61542704  61542705  61542706  61665437  61675891  61687735  61709194  61716311  61716312  61734422  61779045  61801115  61817466  61825785  61846495  61846497  61867244  61867245  61881178  61884422  61926356  61934256  61949820  61953119  61999743  61999745  61360117  61439035  61873615  61941383  61978755  62010624  61880358  61263157  61263159  61281209  61283578  61283579  61360119  61360121  61397916  61397917  61407891  61417318  61429477  61439033  61454837  61454838  61454839  61483184  61487486  61489321  61511681  61521915  61528589  61528593  61536415  61536416  61564183  61564184  61580857  61655827  61665436  61665438  61687739  61709190  61716310  61716313  61750774  61809034  61880356  61880357  61934260  61952442  61957101  62036848  62036849  62076914  61292412  61397920  61412058  61528590  61817468  61916324  61939317  61271328  61489316  61884423  61201906  61215853  61216910  61237814  61237815  61237817  61246484  61256546  61260613  61260614  61263154  61263158  61271327  61271329  61281213  61283580  61292686  61292687  61292689  61310908  61324697  61324698  61324699  61324701  61329096  61329097  61350521  61350522  61360118  61374948  61374952  61381054  61381055  61397914  61397918  61397919  61403531  61403532  61412057  61417321  61429472  61429474  61435494  61435496  61435508  61435509  61439034  61439037  61439038  61439039  61454840  61471001  61483185  61483186  61487485  61487487  61489319  61489320  61497388  61511680  61521916  61528586  61528588  61528592  61536414  61536417  61542702  61542703  61580850  61580851  61580852  61580854  61655828  61675892  61687732  61687740  61709192  61716315  61744560  61770611  61783466  61783467  61801118  61803781  61809035  61809036  61825788  61846499  61855186  61855187  61860446  61880354  61881177  61881179  61901545  61930394  61934255  61934257  61936094  62027430  62036847  62050953  62059665  62059667  62065251  61374950  61374951  61397915  00369735  61706776  61706801  00369701  61598312  62043450  62063927  62022704  61739605  61744561  61749010  61860450  61952439  62050956  62052803  61911395  61367259  61783472  62017506  61926357  61342479  61612735  61842621  61895517  62059672  62027431  61310917  61500691  61728668  61758226  61758229  61762009  61783475  61788126  61896730  61896731  61911397  61237837  61310920  61564197  61598304  61598310  61612734  61627447  61717988  61724212  61734259  61744563  61744564  61790425  61794943  61880350  61895519  61901551  61923933  61964705  62010619  62027432  62059670  62076913  61536696  62059668  61237846  61310913  61519978  61564186  61578389  61578390  61598311  61612736  61642354  61665439  61675901  61688594  61694141  61706798  61749011  61764669  61770607  61801109  61801111  61842620  61867247  61880352  61916325  61939318  61941384  61952441  61956939  61976106  61999748  62017507  62052802  62076912  61724213  62052804  61429560  62063925  61235847  61237840  61269945  61271334  61310918  61397928  61417323  61439102  61439104  61439110  61542712  61564195  61578387  61578391  61612733  61617238  61617239  61627443  61665442  61694148  61706797  61717987  61728669  61744562  61750768  61764672  61775743  61783473  61783474  61794944  61794945  61809041  61822498  61825794  61837773  61860448  61873612  61873613  61895520  61895522  61906148  61911396  61916326  61926571  61926572  61936099  61949821  61978760  61991359  62006529  62010620  62043447  61500687  61717981  61976111  61237830  61263550  61263551  61271332  61283586  61361697  61367263  61397923  61439105  61439111  61487492  61489322  61489324  61500686  61500692  61500694  61519984  61519985  61524175  61524176  61536422  61542708  61542710  61564161  61598306  61598308  61598314  61611850  61617236  61642349  61642351  61661044  61673522  61675893  61688590  61688592  61688593  61694139  61694142  61694145  61717984  61724214  61734260  61734261  61749013  61750772  61788124  61803776  61860447  61862662  61867248  61880348  61896734  61901550  61901553  61923932  61930395  61951209  61957106  61991358  62006532  62010618  62010622  62017510  61260617  61471006  61564192  61788125  61951208  61968000  62065240  62065241  61235850  61237825  61237839  61237844  61256548  61263545  61263547  61283582  61292413  61324705  61340485  61342480  61342481  61350525  61350527  61361696  61397922  61412063  61429572  61439092  61439101  61439103  61455902  61455906  61455910  61471005  61471007  61482148  61482151  61489335  61500689  61511468  61519980  61519982  61521920  61524178  61524179  61536424  61536430  61536695  61542711  61564190  61564191  61564196  61617235  61617237  61627442  61642353  61660734  61661040  61661043  61665441  61665443  61665444  61675894  61706799  61717980  61717983  61717986  61734263  61739606  61750769  61750771  61758231  61764670  61770608  61770613  61783476  61790424  61801112  61809038  61809040  61817469  61817470  61825795  61837774  61846506  61864077  61867246  61867249  61880347  61885850  61895518  61896732  61896733  61906146  61916327  61926569  61938320  61953121  61956938  61976107  61978756  61999747  62010621  61673523  61758227  61880351  62006530  61271331  61880349  61439097  61500688  61500690  61578388  61926568  62006528  61215845  61237822  61237824  61237826  61237828  61237831  61237833  61237836  61237841  61237842  61256550  61256552  61263541  61263542  61263543  61263546  61271330  61271333  61271336  61281220  61281224  61281225  61283585  61283587  61292692  61292693  61292697  61310910  61310911  61310921  61324702  61324708  61327758  61340391  61340487  61342482  61361693  61361694  61361695  61381056  61383436  61383439  61397927  61403533  61403536  61403539  61407894  61407897  61417329  61439093  61439094  61439106  61439112  61439114  61439115  61439116  61439119  61439120  61455907  61455909  61455911  61487489  61487493  61487494  61489323  61497391  61500685  61511471  61519981  61519983  61521917  61521919  61524174  61524177  61536419  61536421  61536423  61536428  61536697  61542709  61564185  61564187  61564188  61564189  61564198  61564199  61627444  61627445  61627446  61642348  61642350  61642355  61660735  61661039  61661041  61673520  61673521  61694144  61694146  61717982  61717985  61758230  61764671  61764673  61775742  61775744  61801114  61817471  61817473  61822499  61822501  61822502  61825792  61825796  61832868  61837772  61862660  61862661  61864076  61864078  61867250  61873611  61880353  61895523  61901546  61901548  61901549  61906145  61906147  61921684  61921685  61923935  61926570  61936097  61936098  61945044  61953120  61964704  61978757  62006527  62065242  61263544  61271335  61281226  61327757  61374954  61381061  61417327  61455912  61497393  61521921  61675903  61783477  61803779  61957105  61976104  62063926  61367264  61542707  61976108  61235848  61235851  61235852  61235853  61237821  61237823  61237832  61237834  61237835  61237838  61237843  61237845  61246514  61246515  61256549  61256551  61260592  61260616  61260619  61262079  61263548  61263549  61263552  61281218  61281219  61281221  61281222  61281223  61283581  61283583  61283584  61283588  61292414  61292415  61292690  61292691  61292694  61292695  61292696  61292698  61310912  61310914  61310915  61310916  61310919  61324703  61324704  61324706  61324707  61324709  61327755  61327756  61327759  61340390  61342483  61350524  61350526  61361692  61367261  61367262  61374953  61374955  61374957  61374958  61374959  61374960  61374961  61381057  61381058  61381059  61381060  61383437  61383438  61383440  61383441  61397921  61397924  61397925  61397926  61403534  61403538  61407895  61407896  61407898  61412060  61412061  61412062  61417322  61417324  61417325  61417326  61417328  61417330  61429563  61429564  61429565  61429567  61429570  61439090  61439091  61439095  61439096  61439098  61439107  61439108  61439109  61439113  61439117  61439118  61439121  61439122  61455901  61455903  61455904  61455905  61455913  61455914  61482149  61482150  61487488  61487490  61487491  61487495  61489325  61489328  61489329  61489332  61489333  61489334  61497390  61497392  61500683  61500684  61500693  61511469  61511470  61511472  61519977  61519979  61521918  61524173  61536420  61536425  61536426  61536427  61536429  61564160  61564193  61564194  61598302  61598307  61598313  61611848  61642347  61642352  61661042  61665440  61675902  61694143  61706796  61717989  61750770  61758228  61770606  61770612  61775741  61801110  61803777  61803778  61803780  61809039  61809042  61817472  61817474  61822505  61825791  61832866  61832867  61832870  61832871  61846504  61860449  61862663  61864079  61873610  61873614  61896735  61901547  61901552  61906456  61906457  61921686  61939319  61941710  61943341  61945043  61952437  61957107  61976109  62006531  62022703  61260618  61374956  61429562  61439099  61825793  61846505  61923934  61952438  62079969  62086033  62079970  62086034  61855194  61855192  62079967  62079968  61855193  61495880  00369659  61483187  61911407  61911406  61945046  61923936  61945045  61564163  62060659  61612737  61483189  61567741  61688597  62065237  62079960  61632195  61246516  62065238  61263557  61367268  61612748  61891253  61548258  61611855  61612747  61688598  61694152  62010609  62065234  62079958  61324714  61455874  61536711  61612742  61632196  62010610  61407903  61324712  61500697  61511480  61536708  61536709  61542720  61675916  61675924  61350529  61407901  61439074  61439081  61500701  61630570  61642363  61744565  61749015  61762013  61779049  61957125  61986608  62065236  61717993  62073496  61511477  61483191  61237848  61260623  61342484  61350530  61429548  61439067  61483190  61536703  61536706  61536707  61542714  61542716  61612739  61612740  61642360  61642364  61675904  61758237  61801120  61839045  61855189  61864083  61899041  61993147  62060661  61417335  61809045  61855191  61201936  61235855  61235856  61237861  61256555  61342485  61374962  61439070  61500699  61511684  61542719  61542721  61548255  61548257  61567739  61598317  61630571  61661048  61665449  61688596  61694149  61728666  61750779  61798058  61880360  61882580  61890574  61911401  62006534  62026419  61739607  61884426  62060665  61536704  61212913  61235858  61237850  61237854  61237859  61260593  61292701  61292702  61292706  61350528  61393299  61393304  61407902  61417331  61439068  61439078  61459855  61459856  61489337  61500696  61500702  61500703  61511473  61511475  61511478  61511479  61511481  61542717  61567740  61598319  61612738  61612746  61632193  61632194  61642359  61642361  61642362  61661047  61675923  61706785  61706787  61717849  61764679  61788133  61794949  61798057  61825801  61839044  61846502  61884427  61938323  61957123  61986609  62010611  62065235  61271337  61455877  61764681  61820792  61869705  61511683  61195550  61201934  61235857  61235859  61237849  61237853  61237857  61237862  61246517  61260621  61263560  61281229  61283592  61324710  61324713  61327761  61374964  61374965  61397932  61417332  61429545  61439069  61455872  61459857  61500695  61500698  61511476  61536705  61536710  61542722  61548256  61567742  61612745  61665447  61717994  61717995  61724215  61749014  61758235  61758236  61764682  61788135  61788136  61794950  61798059  61809043  61817477  61817478  61825802  61842618  61855190  61864080  61864081  61880359  61911399  61916328  61926359  61930396  61957145  62026417  62065239  61256553  61511474  61542718  61869707  61906149  62036850  62079961  61429537  61548259  61846503  61911398  61195552  61212917  61215841  61235860  61237856  61237860  61260622  61263553  61263555  61263559  61271338  61271339  61271341  61281228  61292416  61292699  61292704  61292705  61310925  61324711  61324715  61327760  61361698  61367269  61374968  61383442  61393307  61397933  61412064  61412065  61439072  61439073  61439075  61459859  61459860  61500704  61500706  61564162  61567738  61598318  61612741  61612744  61642358  61661046  61675910  61675922  61717991  61744566  61744567  61744568  61749016  61750781  61764678  61764680  61788134  61803782  61809044  61817476  61825799  61839043  61840613  61851839  61873620  61873622  61878443  61890575  61899043  61901554  61916330  61938321  61957138  61957141  61964706  61976112  61976123  62026415  62036857  61342486  61350531  61779046  61851836  61890573  61916331  62026416  61393298  61955599  61237852  61350532  61374967  61417334  61459858  61869706  61891252  61941712  61957142  61191631  61201935  61212914  61212915  61215840  61237855  61237858  61237863  61246518  61263554  61263556  61263558  61271340  61281227  61283589  61283590  61283591  61283593  61292417  61292703  61310922  61310923  61310924  61310926  61310927  61310928  61327762  61367265  61367267  61374963  61374966  61383443  61383444  61393302  61393305  61397929  61397930  61397931  61397934  61429546  61429549  61429550  61429551  61429552  61429553  61429554  61439066  61439071  61439076  61439077  61439079  61455870  61455876  61459854  61489336  61500705  61511685  61536702  61542713  61542715  61548254  61567737  61567743  61661045  61665448  61706777  61717990  61717992  61728665  61739608  61750778  61750780  61758238  61779047  61779048  61794948  61817475  61825798  61825804  61840611  61840612  61842619  61846501  61851837  61864082  61869708  61873621  61889270  61889273  61890576  61899042  61916329  61947564  61957126  61986610  62006533  62026418  62036854  62060660  62079959  61256554  61412066  61612743  61955597  61955598  00369633  61945048  61706806  61783485  61783487  61783494  61783499  61839048  61945047  61660746  61864084  62026412  61489345  62065247  61846509  61263568  61724217  61660738  61739609  61864085  61986613  62036860  61310929  61536714  61598327  61661051  61706807  62065244  61564202  62026413  62052795  61846510  61412069  61439126  61517442  61660744  61665451  61706803  61724216  61957147  61976136  61417336  61536717  61598323  61675929  61717998  61762012  61764677  61794942  62036859  61728672  61986612  62010608  61292418  61367272  61489338  61598321  61598328  61598329  61660743  61688603  61694155  61739610  61764674  61779040  61842612  61890578  61899045  61957146  61957155  61851841  61930397  61235872  61237870  61246520  61263561  61271345  61324716  61342487  61407905  61407906  61439123  61439124  61489339  61612751  61612752  61688601  61718001  61744570  61744572  61749018  61750776  61764676  61788127  61790427  61809033  61820789  61966576  61976135  62065245  62079963  61439125  61817482  61916334  62006535  62026410  61934274  61899044  61237871  61237880  61263563  61263567  61271342  61281231  61281234  61292709  61316055  61374970  61381062  61412067  61455919  61473719  61500712  61542728  61548280  61548282  61564203  61598322  61598324  61665450  61717999  61744571  61749021  61809031  61817481  61832844  61873627  61890577  61934271  61957158  61986611  62073490  62065246  62065249  61880362  61489341  61235870  61237865  61237877  61256556  61283599  61292708  61310933  61316054  61393311  61417338  61439128  61439132  61439146  61455921  61473718  61487498  61487499  61489342  61489346  61500711  61511686  61511688  61517445  61536718  61542725  61548269  61548276  61564204  61564205  61564208  61564209  61612749  61612750  61660741  61661050  61675930  61706804  61717996  61762010  61764675  61779041  61779042  61788129  61809032  61851844  61864086  61869715  61891255  62065250  62073489  61316053  61350534  61393314  61439133  61455922  61790428  62073495  62091845  61439143  62076917  61260594  61417339  61825807  61916333  61235873  61237874  61237878  61246521  61263562  61263564  61263565  61263566  61271343  61271346  61281233  61281235  61281236  61283594  61310932  61342488  61367271  61374971  61381063  61381064  61383446  61397938  61439130  61439135  61439141  61439144  61439145  61455920  61455924  61482152  61482153  61489343  61500709  61500714  61511687  61536713  61536715  61548278  61564164  61564207  61598325  61598326  61660745  61661049  61675931  61688604  61744569  61749019  61750777  61790429  61798049  61798050  61817479  61820790  61825808  61832843  61842613  61864087  61869712  61869713  61869714  61873625  61873626  61891254  61911403  61916332  61916335  61932008  61934270  61936108  62006536  62021104  62021107  62026411  62036858  62043452  62052796  62091843  61237867  61310931  61374973  61397940  61542731  61660739  61825805  62021106  61487496  61941714  61235871  61237866  61237868  61237869  61237872  61237873  61237875  61237876  61237879  61246519  61256557  61260624  61260625  61260626  61271344  61281230  61281232  61283595  61283596  61283597  61283598  61292707  61292710  61292711  61310930  61310934  61324717  61327763  61327764  61327765  61342489  61342490  61350533  61350535  61361699  61361701  61367270  61374972  61383445  61383447  61393309  61393312  61397935  61397936  61397937  61397939  61403567  61407904  61439127  61439129  61439131  61439134  61439136  61439137  61439138  61439139  61439140  61439142  61439147  61470956  61470958  61473715  61473716  61473717  61473720  61482154  61487497  61489340  61489344  61500707  61500708  61500710  61500713  61536712  61536716  61542723  61542724  61542726  61542727  61542729  61542730  61548272  61548274  61564165  61564206  61665452  61675928  61688602  61694158  61717997  61728671  61749020  61758232  61758233  61758234  61788128  61798051  61801119  61817480  61825806  61832845  61832848  61842614  61851840  61855195  61880361  61880363  61884428  61884429  61890579  61899046  61901555  61906150  61906151  61911402  61934276  61936110  61957150  61957152  61957157  61966577  61968002  62010606  62010607  62021105  61374969  61412068  61417337  61455923  61564201  61246487  00369737  00369606  61292714  61491768  61711437  61292712  61711438  61809048  61551142  61292419  61743371  61743372  61292420  61292713  61711435  61809047  61212909  61307498  61307483  61307479  61307492  61307474  61307482  61307484  61307496  61417340  61906458  61542732  62052793  61665454  61934265  62010615  61237888  61779043  61817485  61825810  61367273  61612753  61675933  61934263  61237883  61237887  61407908  61718002  62026421  62026422  61237889  61389730  61439148  61491774  61548290  61564214  61675934  61706810  61764683  62052797  61269950  61402332  61612754  61706809  61706812  61764684  61851847  61891251  61909567  61957162  61500715  61758241  61968003  61176482  61201941  61237892  61310937  61356428  61356431  61389726  61403581  61407907  61417341  61470893  61491770  61500716  61542736  61548288  61694160  61706815  61758240  61790431  61798053  61842615  61851845  61890581  61246523  61439150  62043453  61215830  61374978  61491765  61564210  61564211  61661053  61688607  61718003  61749022  61775745  61864088  61890580  61957163  62060657  61976137  61976138  61271347  61491767  61832860  61195571  61235877  61256560  61283600  61316057  61381066  61429459  61455918  61470892  61487501  61491776  61500717  61500718  61511689  61511690  61542733  61548289  61548292  61660752  61675932  61706814  61728670  61749024  61749025  61764685  61764686  61788132  61842616  61863575  61899048  61957161  62073491  61470888  61762014  61864089  61937610  61986614  61536724  62091847  61201939  61235875  61237882  61237885  61237890  61246489  61256558  61256559  61256561  61260629  61260630  61269947  61374975  61374977  61389731  61402331  61455916  61487504  61542735  61564213  61564216  61567577  61567581  61567583  61660756  61694159  61734266  61734267  61758239  61758242  61788131  61817484  61825811  61825813  61851849  61863576  61884430  61926911  61986616  62010616  61246488  61412071  61491771  61491775  61548291  61567580  61832863  61869711  62052798  61909568  61842617  61176484  61212912  61235874  61235876  61235879  61237886  61260628  61269946  61271348  61283602  61310935  61310936  61310938  61316056  61324719  61356432  61361702  61361706  61389728  61403579  61429462  61429463  61429464  61439158  61455915  61455917  61470894  61487505  61491772  61491773  61511691  61536721  61548293  61564212  61564215  61660751  61660755  61675935  61694161  61734264  61798054  61798055  61820791  61825812  61832864  61855196  61863573  61869709  61873617  61884431  61890582  61899047  61906459  61930398  61934261  61941716  61957159  62010617  62026420  62052794  62060658  62076918  61271350  61381068  61429461  61936102  62036862  62060656  62076916  61235878  61439149  61869710  61945049  61993148  61176485  61176486  61176487  61201940  61212910  61215828  61237881  61237884  61237893  61246522  61269948  61271349  61271351  61281237  61281238  61283601  61310939  61324718  61356430  61356438  61361703  61361704  61361705  61367274  61374974  61374976  61381069  61389727  61389729  61389732  61412070  61429460  61439151  61439152  61439153  61439154  61439155  61439156  61439157  61439159  61470889  61470890  61470891  61470910  61470911  61470912  61487502  61487503  61491766  61500719  61536719  61536720  61536722  61536723  61542734  61564217  61567578  61567579  61567582  61660748  61660754  61661052  61661054  61665453  61688605  61688606  61749023  61775746  61779044  61788130  61790432  61817486  61817487  61873618  61873619  61891250  61899049  61906460  61926354  61936101  61937611  61957160  61986615  61994885  61381067  61417342  61491769  61964707  00369650  61682015  61711440  61724383  61402333  61711448  61612757  61532298  61612759  61235881  61660759  61660763  61705656  61899052  61987190  61611314  61633908  61758222  61855197  61612762  61758223  61957268  61794940  62043454  61682016  61212904  61535283  61611318  61612764  61633909  61660772  61665455  61665458  61665459  61682017  61688611  61705651  61711443  61711444  61762008  61798047  61863579  61891256  61612756  61779034  61909572  61937609  61890586  61976144  61987188  62043455  61361707  61402334  61439058  61459869  61480099  61526515  61535287  61611317  61611330  61682021  61682024  61688610  61743369  61768197  61794941  61817490  61832858  61839046  61839047  61860451  61860452  61885855  61885856  61909571  61987186  61987189  62065232  61860453  61066788  61439052  61455897  61459862  61483196  61487508  61500726  61511693  61511697  61511705  61526520  61532304  61535288  61535290  61612761  61612767  61660768  61660770  61661036  61661038  61682019  61682023  61688615  61705655  61711441  61711445  61711449  61734254  61734256  61739604  61750762  61779033  61809024  61832856  61882578  61891257  61916336  61964709  61066791  61532301  61899051  62065233  61945050  61271352  61281240  61283603  61389734  61455880  61487512  61500723  61511695  61511702  61532302  61532303  61611319  61611320  61611322  61612758  61612760  61612766  61633907  61660766  61660771  61682022  61682025  61734255  61739603  61743370  61758224  61758225  61768198  61775739  61775740  61788121  61809025  61817491  61825828  61832853  61869697  61937608  61957266  61983683  61292422  61439051  61526518  61873624  62036864  61066643  61195589  61201947  61215824  61237896  61256562  61260633  61271354  61350541  61389738  61389740  61402335  61407909  61417346  61429516  61439055  61439056  61439059  61455884  61455888  61487509  61487510  61487513  61500727  61500728  61511701  61526516  61535282  61535284  61535286  61612763  61612765  61612768  61612769  61633906  61682020  61688612  61688614  61711442  61711446  61724381  61734257  61750761  61758221  61779032  61801108  61809026  61809027  61817492  61825822  61846508  61855198  61864090  61885857  61890585  61909569  61957262  61957265  61994887  61316058  61429496  61439057  61521923  61535289  61724380  61899050  61916337  61930399  61936111  61964708  61976141  62065231  61937607  61941720  61186223  61201943  61201944  61212902  61212903  61212905  61212907  61237894  61237897  61237901  61237902  61237906  61237909  61237912  61269954  61269955  61269956  61269959  61271356  61292719  61292720  61316060  61324720  61350537  61350543  61374979  61374981  61374985  61389742  61403585  61412072  61412074  61412075  61417343  61429489  61429501  61429517  61439060  61455893  61459863  61470998  61487507  61500724  61511696  61511703  61511704  61526519  61532299  61532305  61661037  61665456  61665457  61688613  61705653  61711450  61724382  61750760  61750763  61775737  61779031  61788122  61788123  61794939  61798046  61798048  61803773  61803774  61809029  61809030  61817489  61820788  61825814  61832849  61832852  61855199  61862664  61863577  61863578  61864091  61869698  61869701  61873623  61906463  61916338  61957267  61966575  61968004  61978761  61996714  62065230  61256564  61269951  61283606  61350545  61403584  61407911  61455886  61612770  61775738  61832850  61934278  61976146  61994886  62036865  62043456  61374984  61407910  61487506  61511706  61176495  61195580  61195592  61201945  61201946  61201948  61201949  61201950  61212906  61215825  61215826  61235880  61235882  61237895  61237898  61237899  61237903  61237904  61237905  61237907  61237908  61237910  61237911  61246490  61246491  61246524  61246525  61246526  61256563  61256565  61260595  61260632  61260634  61260635  61269952  61269953  61269958  61269960  61271353  61271355  61271357  61281239  61283604  61283605  61283607  61283608  61283609  61292421  61292715  61292716  61292717  61292718  61306515  61306516  61310940  61310941  61310942  61310943  61310944  61310945  61316059  61324721  61324722  61342491  61342492  61342493  61350536  61350539  61350540  61350542  61350544  61368470  61368471  61368472  61374980  61374982  61374983  61380817  61380818  61389735  61389736  61389737  61389739  61389741  61389743  61389744  61389745  61402336  61403583  61412073  61417344  61417345  61417347  61429485  61429513  61429518  61429519  61439048  61439049  61439050  61439053  61439061  61439062  61439063  61439064  61439065  61455882  61455891  61455894  61459864  61459865  61459866  61459867  61459868  61470996  61470997  61483197  61483198  61483199  61487511  61500720  61500721  61500722  61500725  61500729  61500730  61511692  61511694  61511698  61511699  61521922  61526517  61526521  61526522  61535285  61661035  61682018  61711447  61817488  61825816  61825819  61825827  61846507  61869699  61882577  61890583  61890584  61895516  61896737  61899053  61906461  61906462  61909570  61926912  61928506  61936113  61938328  61947565  61957264  61976143  61978762  61983682  61987187  61996715  61269957  61439054  61342494  62079971  62079973  62060662  62079972  61878444  00369636  61688616  61511707  61978763  61705658  61532310  61526532  61612773  61983686  62021110  61825830  61978764  61983684  61660778  61846493  61987192  61779038  61532309  61901556  61936096  61952447  61987191  62052799  61201955  61235883  61316061  61535292  61567750  61611876  61660773  61711452  61832834  61878447  61987193  61237919  61310947  61660777  61682027  61711451  61750767  61794936  61548327  61724387  61237915  61429534  61483202  61548328  61567749  61705659  61750765  61779035  61801106  61803775  61936093  61532308  61794937  61235884  61495291  61535294  61567747  61682031  61711453  61724385  61749028  61794938  61846494  61952445  61435499  61801105  62010614  61215834  61256566  61495296  61495297  61535291  61535293  61612771  61809021  61932005  62010612  62021108  62079966  61237913  61705661  61711454  61932006  62026414  61186225  61201953  61271360  61350547  61407912  61429532  61439041  61495298  61567751  61612772  61688617  61705660  61724386  61790426  61851834  61878446  61928507  62021109  61412077  61417348  61682030  61724388  61825833  61891258  61911405  61079424  61186226  61215837  61246527  61260637  61271361  61292721  61310946  61389754  61389756  61435512  61439047  61495295  61511710  61532307  61548329  61567745  61567748  61711455  61743368  61750766  61770605  61809023  61832833  61832837  61869702  61869704  61901558  61926352  61947566  61996716  62059675  62065255  61271358  61381072  61429526  61682033  61724384  61926913  62073494  61926353  61186224  61215835  61215838  61235886  61237917  61237921  61246492  61269965  61283613  61292723  61350546  61361709  61368473  61374987  61374988  61381071  61389749  61402337  61412076  61429523  61439042  61439043  61495301  61511708  61526530  61567752  61612774  61660775  61665460  61688618  61779036  61825831  61846492  61851833  61878445  61906464  61952446  61996717  61168297  61256568  61324723  61340489  61407913  61435498  61439045  61455860  61471002  61471004  61801107  61855185  61863204  61901557  61911404  62073493  61455857  61201952  61201954  61215836  61237914  61237916  61237918  61237920  61237922  61246528  61256567  61260596  61269961  61269962  61269963  61269964  61269966  61271359  61281241  61283610  61283611  61283612  61292722  61292724  61306519  61310948  61316062  61316063  61329098  61329099  61340490  61340491  61350548  61361708  61361710  61374986  61374989  61374990  61389750  61389751  61389752  61389753  61389755  61402338  61417349  61429530  61435497  61435501  61435513  61439044  61439046  61455855  61455859  61470960  61471003  61487514  61487515  61487516  61487517  61495292  61495293  61495294  61495299  61495302  61495303  61511709  61526531  61526533  61612775  61660774  61743367  61749029  61779039  61825832  61832838  61869703  61891259  61926351  61945052  62010613  62052801  62079965  61235885  61237923  61260638  61381070  61429533  00370052  00369804  61901559  61976148  61271363  61976147  61711458  61201957  61642369  61452203  61809049  61916339  61216914  61310949  61642367  61271362  61642371  61896736  62059679  61532311  61882579  61642368  61260640  61281242  61389760  61470999  61483205  61487519  61497394  61497395  61237925  61389759  61389761  61497396  61642370  61711459  61945053  61201956  61237926  61269967  61269968  61329100  61374991  61452202  61452204  61497398  61749031  61860454  61926914  61976149  61216913  61246494  61246529  61412078  61417350  61452200  61483204  61749035  61779050  61783480  61837776  61883395  61938325  61215856  61215857  61237924  61246493  61256569  61283614  61292423  61292725  61306521  61316064  61324724  61350549  61368474  61374992  61381073  61389758  61407914  61417351  61452199  61452201  61455862  61455866  61455868  61459861  61470962  61487518  61497397  61509779  61511711  61526534  61526535  61526536  61548339  61548340  61548341  61564218  61678484  61688619  61688620  61711457  61783483  61825834  61825836  61837775  61906465  61916340  61928508  62059677  61389757  61728664  00370198  00370051  00369742  00370609  00369607  61938326  61934268  61389764  61497401  61511712  61195607  61381077  61483207  61439160  61203781  61237927  61788137  61526541  61162516  61260641  61281243  61329101  61417352  61439162  61455899  61497399  61526540  61817493  61216915  61246495  61269969  61292726  61324725  61324726  61389762  61389763  61439161  61470964  61642372  62081597  00370049  00370753  00369744  00370451  61452205  00369707  61452206  61452207  61389765  61324727  61260642  61292727  00370197  00370196  00370452  00370755  00369746  00369462  61324728  00369641",
      "more_code_info": "61439087  61321090  61511713  00370050  00370754  00369715  61452741  61412643  61388518  00369609  61388517  00114228  00115821  11002997  61456496  62185199  07892010  61371929  61488238  61490889  62159364  61375819  61730090  62093694  62127070  62143476  61330592  61596102  61346645  61514348  61639549  61667063  61838680  07892011  61227810  61227850  61231139  61236231  61236232  61236233  61236234  61269011  61278062  61330643  61384948  61403328  61419064  61450019  61476213  61480725  61483448  61543110  61550407  61585973  61629976  61648310  61667056  61684947  61721145  61742081  61754428  61772408  61791916  61819007  61821404  61857676  61865184  61935788  61950842  61962826  61980105  61998847  62011495  62030494  62049514  62056750  62073333  61469643  00114222  00114227  00115825  11001308  61461912  61456489  07892018  07892019  61375820  62143480  62078805  62093696  62127071  62127072  62143478  61236243  61333101  61366111  61610339  61656317  07892020  61236236  61236241  61271656  61330644  61574185  61603364  61847030  61998849  61217903  61346650  61490881  61721137  62016971  07891295  07891670  07892021  61203328  61209111  61217901  61217902  61227851  61231140  61231141  61231142  61236235  61236237  61236238  61236239  61236240  61236242  61278063  61278064  61278065  61330594  61330645  61330646  61338639  61346649  61346651  61360706  61360754  61366112  61384949  61398113  61403329  61410342  61416482  61419065  61429365  61429366  61442071  61442088  61442089  61456497  61476214  61480726  61483450  61483451  61488239  61488240  61490882  61499621  61514349  61514351  61543111  61550408  61553785  61576463  61576471  61585974  61585975  61603365  61622962  61629977  61633333  61639550  61648311  61667053  61667057  61667058  61670781  61693492  61693499  61709301  61714396  61726609  61726620  61737665  61742070  61742073  61748055  61754422  61769839  61769845  61772410  61784836  61784843  61791910  61796126  61807720  61813382  61819009  61819014  61821399  61831581  61838681  61847027  61857677  61860179  61865186  61946159  61946327  61950843  61955481  61962829  61962830  61980113  61986393  61998850  62023777  62030495  62045776  62049516  62049517  62052892  62056739  62056758  62067320  62073334  61461922  61271655  61384926  61469644  61456490  61456498  00113313  00114223  11000919  11000634  11000638  11003865  61498726  61693503  62152866  62093698  62127074  07891703  61278068  61549744  62078806  62101922  62109741  62109742  62127073  61231143  61269013  61398114  61403330  61585976  61615301  61615303  61813383  07892032  61217908  61231145  61236244  61236247  61271657  61271658  61278069  61338640  61346653  61381471  61384950  61450020  61576464  61633334  61633335  61769841  61851624  61950845  61986395  07892027  07892028  07892029  07892030  07892031  61203331  61206432  61217905  61217906  61217907  61217909  61217910  61217911  61227811  61231144  61231146  61236245  61236246  61236248  61236249  61236250  61236251  61236252  61269012  61278066  61278067  61278070  61330647  61330648  61333102  61346652  61346654  61360755  61371930  61371931  61375759  61381470  61403331  61410413  61410414  61419066  61419067  61422950  61422951  61422952  61422953  61429367  61456491  61476215  61480727  61480728  61488241  61488242  61488243  61488244  61488245  61511157  61511158  61511159  61514352  61514353  61566840  61574186  61574187  61581971  61596103  61603366  61639551  61645216  61645217  61656318  61658798  61658824  61667059  61670778  61670782  61680960  61684948  61709303  61714398  61721139  61726614  61726622  61737668  61737669  61742074  61748058  61754424  61761504  61769847  61772412  61780751  61784844  61791911  61796128  61796129  61807721  61807722  61813377  61819011  61819018  61821401  61838676  61838682  61843997  61851618  61857678  61860180  61879041  61926651  61930221  61935789  61946330  61955483  61962832  61962833  61962834  61979982  61980114  61992775  61998852  62011470  62011498  62016973  62042215  62045782  62049521  62049522  62052904  62056759  62067322  62073335  61384927  61476216  61461913  61461923  61469645  00108987  07892037  61461914  61511164  61514364  61992754  62093301  62093302  62084418  62115271  62143481  62078808  62093304  62093305  62115270  62127075  62127076  62152910  61498735  61217912  61217913  61231147  61231148  61236255  61236260  61271660  61278071  61278072  61381472  61429368  61658799  61992755  62016974  07892039  61278073  07892038  07892040  61206433  61206435  61209113  61209114  61227812  61231149  61236253  61236254  61236256  61236257  61236258  61236259  61236261  61271659  61278074  61330650  61330651  61333103  61338641  61360707  61360756  61360766  61371932  61371933  61381463  61381473  61384928  61398115  61398141  61403332  61403333  61416483  61429369  61450021  61450022  61450023  61456492  61456499  61476217  61480729  61483452  61483453  61488246  61488247  61488248  61498727  61499625  61511160  61511161  61549740  61549745  61553786  61576465  61576472  61585977  61585978  61589928  61603367  61603368  61629978  61633337  61645207  61645209  61648312  61648313  61658825  61670783  61680961  61680962  61684949  61709234  61709305  61714402  61721140  61726615  61730086  61730093  61737671  61742076  61748062  61761516  61769842  61772396  61772399  61784837  61784838  61791912  61796123  61807714  61813379  61813384  61819012  61819013  61819019  61831588  61838678  61843999  61851625  61857679  61860183  61865182  61930130  61930132  61946160  61946331  61955484  61962837  61962838  61962839  61973148  61998853  62005864  62011474  62011499  62036267  62052909  62054499  62073336  61461924  61469646  61469647  62084420  62136224  62136225  62101924  61217914  61217915  61381474  62127077  61813391  61231150  61857681  62011417  07892045  61236267  61236268  61381464  61384929  61450024  61633338  61726618  62016975  61371935  07892043  07892044  61209119  61227813  61231151  61231152  61236262  61236263  61236264  61236265  61236266  61269014  61269015  61271661  61271662  61278075  61278076  61278077  61333104  61338642  61346655  61354844  61371934  61398117  61403334  61410343  61410344  61416484  61419072  61429370  61429371  61450025  61456493  61476218  61480730  61480731  61483454  61483455  61488249  61499626  61499627  61511162  61543112  61549741  61576466  61585979  61589929  61596104  61603369  61603370  61615305  61622963  61629979  61633299  61648314  61658801  61658826  61680963  61684950  61714403  61721141  61730094  61742077  61748063  61761517  61772398  61780746  61784839  61787602  61796124  61800864  61801291  61801294  61813253  61813380  61819015  61819021  61838673  61838679  61844001  61860185  61946332  61955485  61962842  61973150  61973152  61992756  61992777  61998854  62011500  62036268  62042218  62049524  62054501  62056761  62073337  61461925  61456501  61461915  00113304  00113586  61585980  62165609  61566841  61658819  61857682  62136226  62067324  61801292  62084421  62127079  62143483  62152915  61381465  62011501  61236270  61442064  61231153  61236269  61346656  61398118  61403335  61442062  61483456  61499628  61633341  62073338  07891612  07892048  61206439  61222248  61222249  61231154  61231155  61236271  61236272  61269016  61271663  61278078  61278079  61333105  61346657  61366102  61371936  61375760  61381475  61398119  61416485  61419068  61442063  61456502  61476219  61480719  61483457  61483458  61488250  61543113  61549742  61574188  61576467  61596105  61615307  61629980  61645210  61656319  61670784  61684951  61714404  61730095  61742078  61748064  61761518  61769844  61780749  61791915  61807715  61819006  61819017  61838674  61838683  61847029  61913064  61930131  61962843  61973151  61992778  62005857  62030471  62030499  62042219  62049526  61384930  61469648  00114054  00114056  61639552  61633347  61670785  61450026  61422949  62143486  62143487  62049527  62067325  61381466  61416486  61278080  61346658  61819022  61831578  07891613  07892050  07892051  61206440  61222250  61231156  61231157  61236273  61236274  61278081  61346659  61366103  61403336  61450027  61498728  61499629  61543114  61553787  61574189  61581972  61622948  61656320  61680964  61721143  61730087  61742079  61772402  61800869  61818048  61851620  61962844  62005858  62023778  00112875  61384931  61456494  61469649  00112853  11000926  00114057  61633348  62185201  61422948  62109743  62049528  07891614  61450028  61346660  61236275  61416487  61574190  07892053  61222251  61231158  61278082  61730096  61784741  61821396  61950847  62005859  61384932  61456503  61784840  61847031  61721144  07892056  07892055  61236276  61278083  61410345  61450029  61581948  61456495  61461916  61576477  07892058  00115860  61469650"
    }
  ]
}