  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Franklin Lakes",
      "address_1": "1 Becton Dr",
      "reason_for_recall": "BD Interlink Threaded Lock Cannula REF# 303369 and lot #3021392 is being recalled due to the potential that a small percentage of the lot may exhibit open seals which adversely impacts unit package integrity and product sterility.",
      "address_2": "",
      "product_quantity": "268,500 products",
      "code_info": "Marketing Status: Class II, 510K #  K011858  Device Listing Number:  D043644    Lot number 3021392",
      "center_classification_date": "20140115",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Canada and Japan.",
      "state": "NJ",
      "product_description": "BD Interlink threaded lock cannula  This is a locking blunt plastic cannula specifically designed for IV connection to an Interlink¿ injection site. It is screwed down over the Interlink¿ injection site and a blunt plastic cannula inside the housing enters the Interlink¿ injection site to provide a pathway for IV fluid administration.",
      "report_date": "20140122",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Company",
      "recall_number": "Z-0772-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "67098",
      "termination_date": "20160506",
      "more_code_info": "",
      "recall_initiation_date": "20131126",
      "postal_code": "07417-1815",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}