  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63531",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.",
      "recall_number": "Z-0772-2013",
      "product_description": "Trilogy Acetabular Shell Spiked.    The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.",
      "product_quantity": "275,952 units",
      "reason_for_recall": "Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.",
      "recall_initiation_date": "20120914",
      "center_classification_date": "20130204",
      "termination_date": "20140620",
      "report_date": "20130213",
      "code_info": "00369642  00369648  00370032  00369656  61642334  00369644  61832805  61817455  61269936  61412631  61630564  00369407  61642337  61324694  61630563  61269937  61825837  61548181  61846490  61941029  61952408  61329095  61421263  61548182  61642336  60983133  61237717  61421264  00369660  61734272  62001310  61873606  61548185  61237808  61548184  61642338  60919315  61421260  61955590  61548186  61825841  61215919  61321081  61846488  61421258  61762005  61469142  61803767  61212922  61324695  61630550  61269940  61269941  61431735  61469143  61548190  61642339  61630567  61246475  61548191  61630565  61269943  61734270  60983134  61803771  61235835  61324696  61235836  61955591  61321084  61517446  61842609  61869695  00369695  61469149  61215920  61642341  61630551  61548201  61809017  61421267  61548199  61251222  61421266  61783461  61803769  61947736  61952415  61199836  61199839  61283453  61283454  61921683  00369661  61469151  61548205  61952422  61630552  61809012  61642342  61548206  61237809  61923871  61925091  61783462  61403530  61283455  61191067  00370048  00369663  61469152  61215929  61548212  61630569  61923872  61630553  00369638  61469154  00369651  00370752  61975586  00369664  11001720  61490760  62023765  61742082  61222252  61271664  61511137  61629985  61667049  61769838  61844002  61955479  61979981  61227814  61676615  61236278  61549718  61721127  61550409  62159365  61714399  11002260  62052850  07892033  61843998  61384952  61442061  61499614  61532992  61629982  61667062  61754419  61980115  61549720  61461917  61784734  62054500  61511139  61676596  61742071  61857680  61962840  62136223  61203311  62126090  61384953  61979983  07892046  61416489  61737672  61354833  61330560  61456504  62023767  61442060  61354834  07892052  61375761  61450049  62030472  61531423  61581705",
      "more_code_info": ""
    }
  ]
}