  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "69838",
      "recalling_firm": "Stryker Endoscopy",
      "address_1": "5900 Optical Ct",
      "address_2": "N/A",
      "postal_code": "95138-1400",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-0770-2015",
      "product_description": "PKG, 3MM INSERT, BIOPSY FORCEPS, 20CM, P/N 0250282012  Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.",
      "product_quantity": "180,573 units total",
      "reason_for_recall": "The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods",
      "recall_initiation_date": "20141117",
      "center_classification_date": "20141217",
      "termination_date": "20150717",
      "report_date": "20141224",
      "code_info": "643871  941266  941267  941268  941603  1144641  1145940  1346220"
    }
  ]
}