  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "72974",
      "recalling_firm": "GE Medical Systems, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- US, including Washington D.C. Puerto Rico, Virgin Islands, and Guam; and the countries of Afghanistan, Albania, Anguilla, Antigua and Barbuda, Armenia, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Rep., Ecuador, Egypt, Ethiopia, Finland, France, Germany, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kuwait, Lebanon, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Papua New Guinea, Paraguay, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Taiwan, Thailand, Trinidad , Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.",
      "recall_number": "Z-0769-2016",
      "product_description": "GE Healthcare, Brivo XR285amx Mobile Digital Ready Radiographic System    Designed to generate diagnostic radiographic images.",
      "product_quantity": "126",
      "reason_for_recall": "Potential safety latch mechanism failure.  The safety latch, within the column that supports the horizontal arm, is a backup in the unlikely event the primary safety cable breaks. A fall of the arm assembly could result in bodily harm to a person.",
      "recall_initiation_date": "20151223",
      "center_classification_date": "20160209",
      "termination_date": "20160907",
      "report_date": "20160217",
      "code_info": "Lot or serial numbers include but not limited to: 00000086177HL3, 00001028564WK1, 115226HL3, 00000083538HL9, 00000092940HL6, 00000084909HL1, 00000081798HL1, 120634HL1, 00000084729HL3, 00000084161HL9, 00000082463HL1, 00000080868HL3, 120832HL1, 00000089868HL4, 00000082736HL0, 00000093379HL6, 00000092616HL2, 00000092633HL7, 00000081160HL4, 00000077253HL3, 00000076913HL3, 00000078498HL3, 00000084160HL1, 00000078315HL9, 00000082726HL1, 00000085039HL6, 00000080635HL6, 00000091403HL6, 00000085598HL1, 00000085661HL7, 00000077923HL1, 00000081481HL4, 00001027046WK0, 00000081152HL1, 00000086364HL7, 00000092622HL0, 00000085000HL8, 00000084999HL2, 00000084725HL1, 00000083029HL9, 00000078313HL4, 00000084162HL7, 00000083020HL8, 00000077471HL1, 00000083019HL0, 00000082738HL6, 00000083598HL3, 00000092618HL8, 00000084180HL9, 00000085594HL0, 00000084917HL4, 00000076790HL5, 00000091413HL5, 00001028931WK2, 00000102856WK4, 00001028927WK0, 00001028924WK7, 00000084159HL3, 00000084422HL5, 00001029525WK1, 00000077482HL8, 00000082250HL2, 00000075962HL1, 00000079632HL6, 00000082240HL3, 00000081484HL8, 00000084732HL7, 00000077924HL9, 00000079918HL9, 00000082247HL8, 00000082245HL2, 00000082243HL7, 00000077255HL8, 00000086373HL8, 00000087093HL1, 00000102539HL4, 00000085004HL0, 00000085599HL9, 00000079559HL1, 00000079282HL0, 00000077782HL1, 00000082480HL5, 00000079116HL0, 00001028894WK2, 00000078317HL5, 00000079280HL4, 00000129740WG7, 00000079278HL8, 00000078492HL6, 00000075947HL2, 00000084908KL3, 00000084412HL6, 00000084169HL2, 00001027692WK1, 00000085036HL2, 0000000012C559, 00000084178HL3, 00000093852HL2, 00000092058HL7, 00000087079HL0, 00001029020WK3, 00001028888WK4, 00001028659WK9, 00001028662WK3, 00001028661WK5, 00000086380HL3, 00001028660WK7, 00000084172HL6, 00001028595WK5, 00000086365HL4, 00000084411HL8, 00000084402HL7, 00000084906HL7, 00001028866WK0, 00001028925WK4, and 00000085034HL7.",
      "more_code_info": ""
    }
  ]
}