  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ra'Anana",
      "state": "N/A",
      "country": "Israel",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81877",
      "recalling_firm": "WEST PHARMA. SERVICES IL, LTD",
      "address_1": "4, Ha-Sheizaf",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distributor in MO",
      "recall_number": "Z-0767-2019",
      "product_description": "Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only",
      "product_quantity": "3.9 million units",
      "reason_for_recall": "Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women.  There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections.",
      "recall_initiation_date": "20181226",
      "center_classification_date": "20190128",
      "termination_date": "20220720",
      "report_date": "20190206",
      "code_info": "UDI:   (01)07290108240061(17)211031(10)A839(90)3609812 5(91)6070111    Lot Numbers: 7153, 7154, 7634, 7688, 7752, 7909, 7955, 8152, 8153, 8154, 8232, 8233, 8234, 8235, 9028, 9175, 9029, 9176, 9177, 9178, 9367, 9368, 9369, 9370, 9654, 9656, 9655, 9918, 9710, 9711, 9712, 9713, 9839, 9840, 9983, 9984, 9992, 9993, 9994, 9995, A123, A124, A790, A805, A806, A791, A792, A793, A794, A795, A796, A797, A807, A808, A809, A810, A834, A835, A836, A837, A838 & A839"
    }
  ]
}